Small field output factors: Comparison of measurements with various detectors and effects of detector orientation with primary jaw setting
被引:11
作者:
Godson, Henry Finlay
[1
,2
]
Ravikumar, M.
[1
]
Ganesh, K. M.
[1
]
Sathiyan, S.
[1
]
Ponmalar, Y. Retna
[1
,2
]
机构:
[1] Kidwai Mem Inst Oncol, Dept Radiat Phys, Hosur Rd, Bangalore 560029, Karnataka, India
[2] Christian Med Coll & Hosp, Dept Radiotherapy, Vellore, Tamil Nadu, India
关键词:
Detector orientation;
Output factor;
Small field;
PLASTIC SCINTILLATION DETECTORS;
STEREOTACTIC RADIOSURGERY BEAMS;
SIMULATED CORRECTION FACTORS;
SMALL PHOTON FIELDS;
MONTE-CARLO MODEL;
IONIZATION-CHAMBER;
DOSIMETER RESPONSE;
RADIATION-THERAPY;
DIODE DETECTOR;
ION-CHAMBER;
D O I:
10.1016/j.radmeas.2015.12.038
中图分类号:
TL [原子能技术];
O571 [原子核物理学];
学科分类号:
0827 ;
082701 ;
摘要:
Accurate dosimetry in small photon fields used in modern radiotherapy is a challenging task due to electronic disequilibrium, steep dose gradients, source occlusion and size of the sensitive volume of the detector. These challenging effects and the lack of metrological dosimetric reference instigated an investigation on the acquisition of output factor with various detectors in parallel and perpendicular orientations. Small field output factor measurements of tertiary collimators such as BrainLab circular cones, BrainLab mMLC and Millennium MLC were carried out in this study. The data acquired show the differences between output factor results with different detectors for all collimating systems. Good agreement in output values was observed in field sizes greater than similar to 2 x 2 cm(2) for all detectors and all tertiary collimators. For smaller fields when compared to electron field diode (EFD), 0.125 cm(3) ion chamber underestimates the output by up to -11.1% and -20.4% and pinpoint ion chamber underestimates the output by up to -1.5% and -6.1% in their parallel and perpendicular orientation, respectively. In contrast, PTW SRS diode and photon field diode (PFD) overestimate the output factor by up to 2.5% and 6.9% respectively in its parallel orientation. The investigated data for the effect of jaw position (0.25 x 0.25 cm(2), 0.5 x 0.5 cm(2) and 1 x 1 cm(2)) away from the field edge generated by different tertiary collimating systems inferred that the opening of X-Y jaw highly influences the small field output factors. The orientation of the detectors and the position of the jaws could influence the output factors considerably in small fields. (C) 2015 Elsevier Ltd. All rights reserved.