A technique for spatial resolution improvement in helium-beam radiography

被引:6
作者
Amato, C. [1 ,2 ,3 ]
Martisikova, M. [1 ,2 ]
Gehrke, T. [1 ,2 ,4 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, Heidelberg, Germany
[2] Natl Ctr Radiat Res Oncol NCRO, Heidelberg Inst Radiat Oncol HIRO, Heidelberg, Germany
[3] Univ Pisa, Dept Phys, Pisa, Italy
[4] Heidelberg Univ, Dept Phys & Astron, Heidelberg, Germany
关键词
energy degrader; helium-beam radiography; ion-beam imaging; ion-beam radiotherapy; silicon pixel detector; spatial resolution; MODULATION TRANSFER-FUNCTION; PROTON COMPUTED-TOMOGRAPHY; DETECTOR SYSTEM; ION RADIOGRAPHY; THERAPY; PATH; PARTICLE; DESIGN; TIME;
D O I
10.1002/mp.14051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Ion-beam radiography exhibits a significantly lower spatial resolution (SR) compared to x-ray radiography. This is mostly due the multiple Coulomb scattering (MCS) that the ions undergo in the imaged object. In this work, a novel technique to improve the spatial resolution in helium-beam radiography was developed. Increasing helium-beam energies were exploited in order to decrease the MCS, and therefore increase the SR. Methods The experimental investigation was carried out with a dedicated ion-tracking imaging system fully composed of thin, pixelated silicon detectors (Timepix). Four helium beams with increasing energies (from 168.8 to 220.5 MeV/u) were used to image a homogeneous 160 mm PMMA phantom with a 2 mm air gap at middle depth. An energy degrader (ED) was placed between the rear tracking system and the energy-deposition detector to compensate for the longer range associated with more energetic ions. The SR was measured for each beam energy. To take into account the overall impact on the image quality, the contrast-to-noise ratio (CNR), the single-ion water equivalent thickness (WET) precision and the absorbed dose in the phantom were also evaluated as a function of the initial beam energy. FLUKA Monte Carlo simulations were used to support the conceptual design of the experimental setup and for dose estimation. Results In the investigated energy interval, a total SR increase by around 30% was measured with increasing beam energy, reaching a maximum value of 0.69 lp/mm. For radiographs generated with 350 mu Gy of absorbed dose and 220 mu m pixel size, a CNR decrease of 32% was found as the beam energy increases. For 1 mm pixel size, the CNR decreases only by 22%. The CNR of the images was always above 6. The single-ion WET precision was found to be in a range between 1.2% and 1.5%. Conclusions We have experimentally shown and quantified the possibility of improving SR in helium-beam radiography by using increasing beam energies in combination with an ED. A significant SR increase was measured with an acceptable decrease of CNR. Furthermore, we have shown that an ED can be a valuable tool to exploit increasing beam energies to generate energy-deposition radiographs.
引用
收藏
页码:2212 / 2221
页数:10
相关论文
共 42 条
[1]  
[Anonymous], 2010, NUCL INSTRUM METH B, DOI [DOI 10.1016/j.nimb.2010.02.091, DOI 10.1016/J.NIMB.2010.02.091]
[2]   Novel scintillation detector design and performance for proton radiography and computed tomography [J].
Bashkirov, V. A. ;
Schulte, R. W. ;
Hurley, R. F. ;
Johnson, R. P. ;
Sadrozinski, H. F. -W. ;
Zatserklyaniy, A. ;
Plautz, T. ;
Giacometti, V. .
MEDICAL PHYSICS, 2016, 43 (02) :664-674
[3]   Development of proton computed tomography detectors for applications in hadron therapy [J].
Bashkirov, Vladimir A. ;
Johnson, Robert P. ;
Sadrozinski, Hartmut F. -W. ;
Schulte, Reinhard W. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 809 :120-129
[4]  
Battistoni G, 2007, AIP CONF PROC, V896, P31, DOI 10.1063/1.2720455
[5]   MOLIERE THEORY OF MULTIPLE SCATTERING [J].
BETHE, HA .
PHYSICAL REVIEW, 1953, 89 (06) :1256-1266
[6]   A theoretical framework to predict the most likely ion path in particle imaging [J].
Collins-Fekete, Charles-Antoine ;
Volz, Lennart ;
Portillo, Stephen K. N. ;
Beaulieu, Luc ;
Seco, Joao .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (05) :1777-1790
[7]   A maximum likelihood method for high resolution proton radiography/proton CT [J].
Collins-Fekete, Charles-Antoine ;
Brousmiche, Sebastien ;
Portillo, Stephen K. N. ;
Beaulieu, Luc ;
Seco, Joao .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (23) :8232-8248
[9]   Sensitivity study of proton radiography and comparison with kV and MV x-ray imaging using GEANT4 Monte Carlo simulations [J].
Depauw, Nicolas ;
Seco, Joao .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08) :2407-2421
[10]   Prediction of image noise contributions in proton computed tomography and comparison to measurements [J].
Dickmann, J. ;
Wesp, P. ;
Raedler, M. ;
Rit, S. ;
Pankuch, M. ;
Johnson, R. P. ;
Bashkirov, V ;
Schulte, R. W. ;
Parodi, K. ;
Landry, G. ;
Dedes, G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (14)