Mammography Quality in Turkey: Auditors' Report on a Nationwide Survey

被引:1
|
作者
Ozsoy, Arzu [1 ]
Aribal, Erkin [2 ]
Araz, Levent [1 ]
Erdogdu, Merter Bora [3 ]
Sari, Aydin [3 ]
Sencan, Irfan [3 ]
Elmas, Nevra [4 ]
机构
[1] Ankara Numune Training & Res Hosp, Dept Radiol, Ankara, Turkey
[2] Marmara Univ, Fac Med, Dept Radiol, Istanbul, Turkey
[3] Minist Hlth Turkey, Ankara, Turkey
[4] Ege Univ Fac Med, Dept Radiol, Izmir, Turkey
关键词
Mammography; Screening; Turkey; IMAGE QUALITY;
D O I
10.5812/iranjradiol.32936
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To evaluate the mammography image and reporting quality in Turkey. Materials and Methods: One hundred sixty five units which provide mammography examination services were included in the audit. Samples of mammographic examinations conducted between July 1st and December 31st, 2012 were evaluated. Auditors completed web-based evaluation forms, including quality parameters and descriptive statistics on the patients and institutions. Results: Nine hundred eighty eight mammography examinations were evaluated; 47.34% of them used mammography machines less than 5 years old. Most of the older machines (10 years old or more) were grouped in private hospitals (22.1%) and university hospitals (21%). Mammographic image quality evaluation showed 56.1% non-compliance with the standard parameters. Private institutions showed the highest failure rates (P < 0.001), and low-quality images mostly (42.2%) resulted from mistakes in positioning. In 32% of the mammography reports, some mandatory information was incomplete. The reports were deemed inadequate in 59.2% of the cases. Conclusion: The audit showed significant deficiencies in terms of quality. We suggest investing more time and effort to train both radiologists and radiologic technicians to implement image and report quality standards in their units. Moreover, our study emphasized the importance of the accreditation of the units.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Image quality of mammography in Croatian nationwide screening program: Comparison between various types of facilities
    Brnic, Zoran
    Blaskovic, Darko
    Klasic, Branimir
    Ramac, Jelena Popic
    Flegaric-Bradic, Mirjana
    Stimac, Damir
    Lubin, Ivan Zvonimir
    Brnic, Vedran
    Faj, Dario
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (04) : E478 - E485
  • [2] Mammography device use in Turkey, and quantity and quality analysis of mammography education
    Voyvoda, Nuray
    Oezdemir, Ayseguel
    Gueltekin, Serap
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2007, 13 (03): : 129 - 133
  • [3] Quality assurance in Mammography: An overview
    Feigin, Kimberly
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 165
  • [4] Impact of using the new American College of Radiology digital mammography phantom on quality survey in modern digital mammography systems: Evidence from nationwide surveys in Taiwan
    Hwang, Yi-Shuan
    Tsai, Hui-Yu
    Lin, Yu-Ying
    Liao, Ying-Lan
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 117 : 9 - 14
  • [5] Mammography Dose Survey Using International Quality Standards
    Boujemaa, Soumaya
    Bosmans, Hilde
    Bentayeb, Farida
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2019, 50 (04) : 529 - 535
  • [6] Assessment of mammography quality in Istanbul
    Gurdemir, Burcu
    Aribal, Erkin
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2012, 18 (05) : 468 - 472
  • [7] Nationwide cross-sectional adherence to mammography screening guidelines: national behavioral risk factor surveillance system survey results
    Narayan, Anand
    Fischer, Alexander
    Zhang, Zihe
    Woods, Ryan
    Morris, Elizabeth
    Harvey, Susan
    BREAST CANCER RESEARCH AND TREATMENT, 2017, 164 (03) : 719 - 725
  • [8] Nationwide cross-sectional adherence to mammography screening guidelines: national behavioral risk factor surveillance system survey results
    Anand Narayan
    Alexander Fischer
    Zihe Zhang
    Ryan Woods
    Elizabeth Morris
    Susan Harvey
    Breast Cancer Research and Treatment, 2017, 164 : 719 - 725
  • [9] Clinical Image Quality and Sensitivity in an Organized Mammography Screening Program
    Theberge, Isabelle
    Guertin, Marie-Helene
    Vandal, Nathalie
    Daigle, Jean-Marc
    Dufresne, Michel-Pierre
    Wadden, Nancy
    Shumak, Rene
    Samson, Caroline
    Langlois, Andre
    Larocque, Isabelle
    Perron, Linda
    Pelletier, Eric
    Brisson, Jacques
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2018, 69 (01): : 16 - 23
  • [10] The Association of Mammography Environment and Image Quality
    Sarquis-Kolber, Shakira
    Schilling, Kathy
    Hanney, William J.
    Kolber, Morey J.
    JOURNAL OF BREAST IMAGING, 2020, 2 (05) : 436 - 442