A study of parameters affecting the quality of residual stress measurements using XRD

被引:8
作者
Fry, AT [1 ]
Kandil, FA [1 ]
机构
[1] Natl Phys Lab, NPL Mat Ctr, Teddington TW11 0LW, Middx, England
来源
ECRS 6: PROCEEDINGS OF THE 6TH EUROPEAN CONFERENCE ON RESIDUAL STRESSES | 2002年 / 404-7卷
关键词
residual stress; X-ray diffraction; measurement uncertainty; spring steel; aluminium; titanium; repeatability; peak fitting; height offset; step size; psi tilts; counting time;
D O I
10.4028/www.scientific.net/MSF.404-407.579
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Extensive work has been conducted over the last three years as part of a UK national project on the measurement of residual stress in components, in which over 250 individual X-ray diffraction tests were conducted to evaluate the sensitivity of XRD residual stress measurements to changes in the test set-up. These measurements provided information that enabled the uncertainty in the reported results to be evaluated for three different materials. The effect of changes in the test parameters is also shown, as is the variation in results obtained using different peak fitting routines.
引用
收藏
页码:579 / 584
页数:6
相关论文
共 7 条
[1]  
*BIPM IEC IFCC ISO, 1993, GUID EXPR UNC MEAS
[2]  
Cullity BD, 1978, ELEMENTS XRAY DIFFRA
[3]  
FITZPATRICK ME, 2002, XRD GOOD PRACTICE GU
[4]  
JONES AM, 1989, AERER13005
[5]  
Kandil F.A., MATCA04 NPL
[6]  
KANDIL FA, 2000, MANUAL CODES PRACTIC
[7]  
Noyan I. C., 1987, RESIDUAL STRESS MEAS