Comparing the clinical utility of single-shot echo-planar imaging and readout-segmented echo-planar imaging in diffusion-weighted imaging of the liver at 3 tesla

被引:5
作者
Xie, Simin [1 ]
Masokano, Ismail Bilal [1 ]
Liu, Wenguang [1 ]
Long, Xueying [1 ]
Li, Guijin [2 ]
Pei, Yigang [1 ,3 ]
Li, Wenzheng [1 ]
机构
[1] Cent South Univ, Xiangya Hosp, Dept Radiol, Changsha 410008, Hunan, Peoples R China
[2] Siemens Healthcare Ltd, Guangzhou Branch, Guangzhou 510620, Peoples R China
[3] Cent South Univ, Xiangya Hosp, Changsha 410008, Hunan, Peoples R China
基金
中国博士后科学基金;
关键词
Hepatic diffusion-weighted imaging; Single-shot echo-planar imaging (EPI); Readout-segmented EPI; NT-DWI; Clinical utility;
D O I
10.1016/j.ejrad.2020.109472
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the clinical utility of single-shot echo-planar imaging (SS-EPI) using different breathing schemes and readout-segmented EPI (RS-EPI) in the repeatability of apparent diffusion coefficient (ADC) measurements, signal-to-noise ratio (SNR) and image quality. Methods: In this institutional review board-approved prospective study, hepatic DWIs (b = 50, 300, 600 s/mm(2)) were performed in 22 volunteers on 3.0 T MRI using SS-EPI with free-breathing diffusion-weighted imaging (FB-DWI), breath-hold (BH-DWI), respiratory-triggered (RT-DWI) and navigator-triggered (NT-DWI), and readout-segmented EPI (RS-DWI). ADC and surrogate SNR (sSNR) were measured in nine anatomic locations in the right lobe, and image quality was assessed on all FB-DWI, BH-DWI, RT-DWI, NT-DWI, and RS-DWI sequences. The sequence with the optimal clinical utility was decided by systematically comparing the ADC repeatability, sSNR and image quality of the above DWIs. Results: In all the five sequences, NT-DWI had the most reliable intra-observer agreement (intraclass correlation coefficient (ICC): 0.900-0.922; all P > 0.05), and a better interobserver agreement (ICC: 0.853 0.960; all p > 0.05) than RS-DWI (ICC:0.881-0.916; some P < 0.05). NT-DWI had the best ADC repeatability in the nine locations (mean ADC absolute differences: 38.47-56.38 x 10(-6) mm(2)/s, limits of agreement (LOA): 17.33-22.52 x 10(-6) mm(2)/s). Also, NT-DWI had the highest sSNR (Reader 1: 50.58 +/- 20.11 (Superior), 74.06 +/- 28.37 (Central), 80.99 +/- 38.11(Inferior)); Reader 2: 48.07 +/- 23.92 (Superior), 68.23 +/- 32.91 (Central), 76.78 +/- 33.07 (Inferior)) in three representative sections except for RS-DWI. Furthermore, NT-DWI had a better image quality than RS-DWI (P < 0.05) and was superior to FB-DWI and BH-DWI in sharpness of the liver (at b = 300 s/mm(2)) (P < 0.05) Conclusion: RS-DWI has the best SNR. However, NT-DWI can provide sufficient SNR, excellent image quality, and the best ADC repeatability on 3.0 T MRI. It is thus the recommended sequence for the clinical application of hepatic DWI.
引用
收藏
页数:10
相关论文
共 26 条
  • [1] Basic principles of diffusion-weighted imaging
    Bammer, R
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2003, 45 (03) : 169 - 184
  • [2] Bickel H, 2015, INVEST RADIOL, V50, P95, DOI 10.1097/RLI.0000000000000104
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Reliability and validity of MR image lung volume measurement in fetuses with congenital diaphragmatic hernia and in vitro lung models
    Buesing, Karen A.
    Kilian, A. Kristina
    Schaible, Thomas
    Debus, Angelika
    Weiss, Christel
    Neff, K. Wolfgang
    [J]. RADIOLOGY, 2008, 246 (02) : 553 - 561
  • [5] Liver Diffusion-weighted MR Imaging: Reproducibility Comparison of ADC Measurements Obtained with Multiple Breath-hold, Free-breathing, Respiratory-triggered, and Navigator-triggered Techniques
    Chen, Xin
    Qin, Lei
    Pan, Dan
    Huang, Yanqi
    Yan, Lifen
    Wang, Guangyi
    Liu, Yubao
    Liang, Changhong
    Liu, Zaiyi
    [J]. RADIOLOGY, 2014, 271 (01) : 113 - 125
  • [6] Comparison of breathhold, navigator-triggered, and free-breathing diffusion-weighted MRI for focal hepatic lesions
    Choi, Ji Soo
    Kim, Myeong-Jin
    Chung, Yong Eun
    Kim, Kyung Ah
    Choi, Jin-Young
    Lim, Joon Seok
    Park, Mi-Suk
    Kim, Ki Whang
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (01) : 109 - 118
  • [7] Comparison of Free-Breathing With Navigator-Controlled Acquisition Regimes in Abdominal Diffusion-Weighted Magnetic Resonance Images: Effect on ADC and IVIM Statistics
    Jerome, Neil P.
    Orton, Matthew R.
    d'Arcy, James A.
    Collins, David J.
    Koh, Dow-Mu
    Leach, Martin O.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (01) : 235 - 240
  • [8] Reproducibility of measurement of apparent diffusion coefficients of malignant hepatic tumors: Effect of DWI techniques and calculation methods
    Kim, So Yeon
    Lee, Seung Soo
    Park, Bumwoo
    Kim, Namkug
    Kim, Jeong Kon
    Park, Seong Ho
    Byun, Jae Ho
    Song, Ki Jun
    Koo, Ja-heung
    Choi, Eun Kyung
    Lee, Moon-Gyu
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (05) : 1131 - 1138
  • [9] Malignant Hepatic Tumors: Short-term Reproducibility of Apparent Diffusion Coefficients with Breath-hold and Respiratory-triggered Diffusion-weighted MR Imaging
    Kim, So Yeon
    Lee, Seung Soo
    Byun, Jae Ho
    Park, Seong Ho
    Kim, Jeong Kon
    Park, Bumwoo
    Kim, Namkug
    Lee, Moon-Gyu
    [J]. RADIOLOGY, 2010, 255 (03) : 815 - 823
  • [10] Diffusion-weighted single-shot echoplanar MR imaging for liver disease
    Kim, T
    Murakami, T
    Takahashi, S
    Hori, M
    Tsuda, K
    Nakamura, H
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (02) : 393 - 398