Three-dimensional conformal setup (3D-CSU) of patients using the coordinate system provided by three internal fiducial markers and two orthogonal diagnostic X-ray systems in the treatment room

被引:18
作者
Shirato, H [1 ]
Oita, M [1 ]
Fujita, K [1 ]
Shimizu, S [1 ]
Onimaru, R [1 ]
Uegaki, S [1 ]
Watanabe, Y [1 ]
Kato, N [1 ]
Miyasaka, K [1 ]
机构
[1] Hokkaido Univ, Sch Med, Dept Radiol, Kita Ku, Sapporo, Hokkaido 0608638, Japan
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2004年 / 60卷 / 02期
关键词
three-dimensional conformal setup; real-time tumor-tracking radiation therapy; fiducial marker; setup error; rotation correction; rotational setup error; real-time tumor-tracking system; linear accelerator;
D O I
10.1016/j.ijrobp.2004.05.042
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To test the accuracy of a system for correcting for the rotational error of the clinical target volume (CTV) without having to reposition the patient using three fiducial markers and two orthogonal fluoroscopic images. We call this system "three-dimensional conformal setup" (3D-CSU). Methods and Materials: Three 2.0-mm gold markers are inserted into or adjacent to the CTV. On the treatment couch, the actual positions of the three markers are calculated based on two orthogonal fluoroscopies crossing at the isocenter of the linear accelerator. Discrepancy of the actual coordinates of gravity center of three markers from its planned coordinates is calculated. Translational setup error is corrected by adjustment of the treatment couch. The rotation angles (alpha, beta, gamma) of the coordinates of the actual CTV relative to the planned CTV are calculated around the lateral (x), craniocaudal (y), and anteroposterior (z) axes of the planned CTV. The angles of the gantry head, collimator, and treatment couch of the linear accelerator are adjusted according to the rotation of the actual coordinates of the tumor in relation to the planned coordinates. We have measured the accuracy of 3D-CSU using a static cubic phantom. Results: The gravity center of the phantom was corrected within 0.9 +/- 0.3 mm (mean +/- SD), 0.4 +/- 0.2 mm, and 0.6 +/- 0.2 mm for the rotation of the phantom from 0-30 degrees around the x, y, and z axes, respectively, every 5 degrees. Dose distribution was shown to be consistent with the planned dose distribution every 10 degrees of the rotation from 0-30 degrees. The mean rotational error after 3D-CSU was -0.4 +/- 0.4 (mean +/- SD), -0.2 +/- 0.4, and 0.0 +/- 0.5 degrees around the x, y, and z axis, respectively, for the rotation from 0-90 degrees. Conclusions: Phantom studies showed that 3D-CSU is useful for performing rotational correction of the target volume without correcting the position of the patient on the treatment couch. The 3D-CSU will be clinically useful for tumors in structures such as paraspinal diseases and prostate cancers not subject to large internal organ motion. (C) 2004 Elsevier Inc.
引用
收藏
页码:607 / 612
页数:6
相关论文
共 14 条
[1]   Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation [J].
Dehnad, H ;
Nederveen, AJ ;
van der Heide, UA ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) :295-302
[2]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[3]   Detection of organ movement in cervix cancer patients using a fluoroscopic electronic portal imaging device and radiopaque markers [J].
Kaatee, RSJP ;
Olofsen, MJJ ;
Verstraate, MBJ ;
Quint, S ;
Heijmen, BJM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (02) :576-583
[4]   Registration accuracy and possible migration of internal fiducial gold marker implanted in prostate and liver treated with real-time tumor-tracking radiation therapy (RTRT) [J].
Kitamura, K ;
Shirato, H ;
Shimizu, S ;
Shinohara, N ;
Harabayashi, T ;
Shimizu, T ;
Kodama, Y ;
Endo, H ;
Onimaru, R ;
Nishioka, S ;
Aoyama, H ;
Tsuchiya, K ;
Miyasaka, K .
RADIOTHERAPY AND ONCOLOGY, 2002, 62 (03) :275-281
[5]   Tolerance of organs at risk in small-volume, hypofractionated, image-guided radiotherapy for primary and metastatic lung cancers [J].
Onimaru, R ;
Shirato, H ;
Shimizu, S ;
Kitamura, K ;
Xu, B ;
Fukumoto, S ;
Chang, TC ;
Fujita, K ;
Oita, M ;
Miyasaka, K ;
Nishimura, M ;
Dosaka-Akita, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :126-135
[6]   Evaluating the effect of rectal distension and rectal movement on prostate gland position using cine MRI [J].
Padhani, AR ;
Khoo, VS ;
Suckling, J ;
Husband, JE ;
Leach, MO ;
Dearnaley, DP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 44 (03) :525-533
[7]   Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy [J].
Seppenwoolde, Y ;
Shirato, H ;
Kitamura, K ;
Shimizu, S ;
van Herk, M ;
Lebesque, JV ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (04) :822-834
[8]   Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers [J].
Shimizu, S ;
Shirato, H ;
Kitamura, K ;
Shinohara, N ;
Harabayashi, T ;
Tsukamoto, T ;
Koyanagi, T ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (05) :1591-1597
[9]   Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy [J].
Shirato, H ;
Harad, T ;
Harabayashi, T ;
Hida, K ;
Endo, H ;
Kitamura, K ;
Onimaru, R ;
Yamazaki, K ;
Kurauchi, N ;
Shimizu, T ;
Shinohara, N ;
Matsushita, M ;
Dosaka-Akita, H ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :240-247
[10]   Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor [J].
Shirato, H ;
Shimizu, S ;
Kitamura, K ;
Nishioka, T ;
Kagei, K ;
Hashimoto, S ;
Aoyama, H ;
Kunieda, T ;
Shinohara, N ;
Dosaka-Akita, H ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (02) :435-442