Alumina Delta-on-Highly Crosslinked-Remelted Polyethylene Bearing in Cementless Total Hip Arthroplasty in Patients Younger than 50 Years

被引:13
作者
Kim, Young-Hoo [1 ]
Park, Jang-Won [2 ]
Kim, Jun-Shik [2 ]
机构
[1] Ewha Womans Univ, SeoNam Hosp, Joint Replacement Ctr, Seoul, South Korea
[2] Ewha Womans Univ, MokDong Hosp, Joint Replacement Ctr, 911-1 MokDong, Seoul 158710, South Korea
关键词
cementless total hip arthroplasty; alumina delta-on-alumina-delta; alumina delta-on-highly crosslinked; polyethylene; young patients; polyethylene wear; osteolysis; RANDOMIZED PROSPECTIVE EVALUATION; MOLECULAR-WEIGHT POLYETHYLENE; CERAMIC FEMORAL-HEAD; RADIOSTEREOMETRIC ANALYSIS; WEAR; FRACTURE; OSTEOLYSIS; OUTCOMES; STEM; REPLACEMENT;
D O I
10.1016/j.arth.2016.05.020
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The purpose of this retrospective study was to determine the clinical and radiographic results, prevalence of polyethylene wear and osteolysis, and fracture of alumina delta ceramic femoral head or highly crosslinked, remelted polyethylene (HXLPE) liner associated with the use of alumina delta ceramic femoral head-on-HXLPE bearing in cementless total hip arthroplasty in patients younger than 50 years. Methods: We reviewed the cases of 119 patients (130 hips) who underwent a cementless total hip arthroplasty using alumina delta ceramic-on-HXLPE bearing when they were 50 years or younger at the time of surgery. The most common diagnoses were osteonecrosis (51%) and osteoarthritis secondary to developmental dysplastic hip (39%). Osteolysis and polyethylene wear rates were evaluated with use of radiography and computed tomography. In addition, prevalence of fracture of alumina delta ceramic head and polyethylene line was documented. The mean follow-up was 8.3 years (range, 7-9 years). Results: The mean Harris hip score, Western Ontario and McMaster Universities Osteoarthritis Index score, University of California, Los Angeles activity score were 94 points, 14 points, and 8.1 points, respectively, at the final follow-up. No patient had thigh pain. All acetabular components and all but one femoral components were well fixed. The mean annual penetration rate of femoral head was 0.022 +/- 0.003 mm/year. No hip had osteolysis or ceramic head or HXLPE liner fracture. Conclusion: Our average 8.3-year results with the use of alumina delta ceramic-on-HXLPE bearing in the patients younger than 50 years suggest that cementless acetabular and femoral components provide a high survival rate without evidence of osteolysis or ceramic head or AXLPE liner fracture. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:2800 / 2804
页数:5
相关论文
共 39 条
  • [1] Revision total hip arthroplasty performed after fracture of a ceramic femoral head - A multicenter survivorship study
    Allain, J
    Roudot-Thoraval, F
    Delecrin, J
    Anract, P
    Migaud, H
    Goutallier, D
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) : 825 - 830
  • [2] [Anonymous], J ARTHROPLASTY
  • [3] BELLAMY N, 1988, J RHEUMATOL, V15, P1833
  • [4] Fatigue crack propagation resistance of highly crosslinked polyethylene
    Bradford, L
    Baker, D
    Ries, MD
    Pruitt, LA
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) : 68 - 72
  • [5] Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36-vs 28-mm femoral heads
    Bragdon, Charles R.
    Greene, Meridith E.
    Freiberg, Andrew A.
    Harris, William H.
    Malchau, Henrik
    [J]. JOURNAL OF ARTHROPLASTY, 2007, 22 (06) : 125 - 129
  • [6] THE UNCEMENTED POROUS-COATED ANATOMIC TOTAL HIP-PROSTHESIS - 2-YEAR RESULTS OF A PROSPECTIVE CONSECUTIVE SERIES
    CALLAGHAN, JJ
    DYSART, SH
    SAVORY, CG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) : 337 - 346
  • [7] Clarke IC, 2000, CLIN ORTHOP RELAT R, P34
  • [8] Second-generation Annealed Highly Cross-linked Polyethylene Exhibits Low Wear
    D'Antonio, James A.
    Capello, William N.
    Ramakrishnan, Rama
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (06) : 1696 - 1704
  • [9] 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets -: Two randomized studies using radiostereometric analysis
    Digas, Georgios
    Karrholm, Johan
    Thanner, Jonas
    Herberts, Peter
    [J]. ACTA ORTHOPAEDICA, 2007, 78 (06) : 746 - 754
  • [10] Dorr L., 1986, TECH ORTHOP, V1, P22, DOI [10.1097/00013611-198610000-00007, DOI 10.1097/00013611-198610000-00007]