Comparative Characterization Study of a LaBr3(Ce) Scintillation Crystal in Two Surface Wrapping Scenarios: Absorptive and Reflective

被引:13
作者
Aldawood, Saad [1 ,2 ]
Castelhano, Ines [1 ,3 ]
Gemhaeuser, Roman [4 ]
Van der Kolff, Hugh [1 ,5 ]
Lang, Christian [1 ]
Liprandi, Silvia [1 ]
Lutter, Rudolf [1 ]
Maier, Ludwig [4 ]
Marinsek, Wm [1 ]
Schaart, Dennis R. [5 ]
Parodi, Katia [1 ]
Thirolf, Peter G. [1 ]
机构
[1] Univ Munich, Fac Phys, Munich, Germany
[2] King Saud Univ, Dept Phys & Astron, Riyadh, Saudi Arabia
[3] Univ Lisbon, Fac Sci, P-1699 Lisbon, Portugal
[4] Tech Univ Munich, Phys Dept E12, Garching, Germany
[5] Delft Univ Technol, Fac Appl Sci Radiat Sci & Technol, Delft, Netherlands
来源
FRONTIERS IN ONCOLOGY | 2015年 / 5卷
关键词
LaBr3:Ce scintillator; gamma spectroscopy; medical imaging; crystal surface coating; Compton camera; ENERGY-RESOLUTION; DETECTOR; LABR3-CE;
D O I
10.3389/fonc.2015.00270
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The properties of a 50 mm x 50 mm x 30 mm monolithic LaBr3:Ce scintillator crystal coupled to a position-sensitive multi-anode photomultiplier (PMT, Hamamatsu H9500), representing the absorbing detector of a Compton camera under study for online ion (proton) beam range verification in hadron therapy, was evaluated in combination with either absorptive or reflective crystal surface coating. This study covered an assessment of the energy and position-dependent energy resolution, exhibiting a factor of 2.5-3.5 improvement for the reflectively wrapped crystal at 662 keV. The spatial dependency was investigated using a collimated Cs-137 source, showing a steep degradation of the energy resolution at the edges and corners of the absorptively wrapped crystal. Furthermore, the time resolution was determined to be 273 ps (FWHM) and 536 ps (FWHM) with reflective and absorptive coating, respectively, using a Co-60 source. In contrast, the light spread function (LSF) of the light amplitude distribution on the PMT segments improved for the absorptively wrapped detector. Both wrapping modalities showed almost no differences in the energy-dependent photopeak detection efficiency.
引用
收藏
页数:9
相关论文
共 20 条
  • [1] Scintillation properties of ⊘ 1x1 inch3 LaBr3:: 5%Ce3+ crystal
    Bizarri, G
    de Haas, JIM
    Dorenbos, P
    van Eijk, CWE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (02) : 615 - 619
  • [2] The timing resolution of scintillation-detector systems: Monte Carlo analysis
    Choong, Woon-Seng
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (21) : 6495 - 6513
  • [3] Characterization of large volume 3.5" x 8" LaBr3:Ce detectors
    Giaz, A.
    Pellegri, L.
    Riboldi, S.
    Camera, F.
    Blasi, N.
    Boiano, C.
    Bracco, A.
    Brambilla, S.
    Ceruti, S.
    Coelli, S.
    Crespi, F. C. L.
    Csatlos, M.
    Frega, S.
    Gulyas, J.
    Krasznahorkay, A.
    Lodetti, S.
    Million, B.
    Owens, A.
    Quarati, F.
    Stuhl, L.
    Wieland, O.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 729 : 910 - 921
  • [4] Crystal growth of large diameter LaBr3:Ce and CeBr3
    Higgins, W. M.
    Churilov, A.
    van Loef, E.
    Glodo, J.
    Squillante, M.
    Shah, K.
    [J]. JOURNAL OF CRYSTAL GROWTH, 2008, 310 (7-9) : 2085 - 2089
  • [5] Development and calibration of the tracking Compton/Pair telescope MEGA
    Kanbach, G
    Andritschke, R
    Zoglauer, A
    Ajello, M
    McConnell, ML
    Macri, JR
    Ryan, JM
    Bloser, P
    Hunter, S
    DiCocco, G
    Kurfess, J
    Reglero, V
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 541 (1-2) : 310 - 322
  • [6] Design of a lanthanum bromide detector for time-of-flight PET
    Kuhn, A
    Surti, S
    Karp, JS
    Raby, PS
    Shah, KS
    Perkins, AE
    Muehllehner, G
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) : 2550 - 2557
  • [7] Sub-millimeter nuclear medical imaging with high sensitivity in positron emission tomography using β+γ coincidences
    Lang, C.
    Habs, D.
    Parodi, K.
    Thirolf, P. G.
    [J]. JOURNAL OF INSTRUMENTATION, 2014, 9
  • [8] Factors Influencing Time Resolution of Scintillators and Ways to Improve Them
    Lecoq, P.
    Auffray, E.
    Brunner, S.
    Jarron, P.
    Knapitsch, A.
    Meyer, T.
    Powolny, F.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (05) : 2411 - 2416
  • [9] Marinsek T, 2015, THESIS
  • [10] Studying the properties and response of a large volume (946 cm3) LaBr3:Ce detector with γ-rays up to 22.5 MeV
    Mazumdar, I.
    Gothe, D. A.
    Kumar, G. Anil
    Yadav, N.
    Chavan, P. B.
    Patel, S. M.