Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources

被引:16
作者
Ghorbani, Mahdi [1 ]
Salahshour, Fateme [2 ]
Haghparast, Abbas [2 ]
Moghaddas, Toktam Ahmadi [2 ]
Knaup, Courtney [3 ]
机构
[1] Mashhad Univ Med Sci, Fac Med, Med Phys Res Ctr, Mashhad, Iran
[2] Kermanshah Univ Med Sci, Fac Med, Med Phys & Med Engn Dept, Kermanshah, Iran
[3] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
brachytherapy; ICRU; photon emitting sources; relative dose; soft tissue; MONTE-CARLO CHARACTERIZATION; IR-192; BRACHYTHERAPY; DOSIMETRY PARAMETERS; PD-103; YB-169; CO-60; WATER; UPDATE; I-125; LUNG;
D O I
10.5114/jcb.2014.42024
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The aim of this study is to compare the dose in various soft tissues in brachytherapy with photon emitting sources. Material and methods: Pd-103, I-125, Yb-169, Ir-192 brachytherapy sources were simulated with MCNPX Monte Carlo code, and their dose rate constant and radial dose function were compared with the published data. A spherical phantom with 50 cm radius was simulated and the dose at various radial distances in adipose tissue, breast tissue, 4-component soft tissue, brain (grey/white matter), muscle (skeletal), lung tissue, blood (whole), 9-component soft tissue, and water were calculated. The absolute dose and relative dose difference with respect to 9-component soft tissue was obtained for various materials, sources, and distances. Results: There was good agreement between the dosimetric parameters of the sources and the published data. Adipose tissue, breast tissue, 4-component soft tissue, and water showed the greatest difference in dose relative to the dose to the 9-component soft tissue. The other soft tissues showed lower dose differences. The dose difference was also higher for Pd-103 source than for I-125, Yb-169, and Ir-192 sources. Furthermore, greater distances from the source had higher relative dose differences and the effect can be justified due to the change in photon spectrum (softening or hardening) as photons traverse the phantom material. Conclusions: The ignorance of soft tissue characteristics (density, composition, etc.) by treatment planning systems incorporates a significant error in dose delivery to the patient in brachytherapy with photon sources. The error depends on the type of soft tissue, brachytherapy source, as well as the distance from the source.
引用
收藏
页码:54 / 67
页数:14
相关论文
共 34 条
  • [11] A Study on the Dose Distributions in Various Materials from an Ir-192 HDR Brachytherapy Source
    Hsu, Shih-Ming
    Wu, Chin-Hui
    Lee, Jeng-Hung
    Hsieh, Ya-Ju
    Yu, Chun-Yen
    Liao, Yi-Jen
    Kuo, Li-Cheng
    Liang, Ji-An
    Huang, David Y. C.
    [J]. PLOS ONE, 2012, 7 (09):
  • [12] Radiotherapy treatment planning based on Monte Carlo techniques
    Juste, Belen
    Miro, Rafael
    Campayo, Juan M.
    Diez, Sergio
    Verdu, Gumersindo
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) : 252 - 257
  • [13] COMPARISON OF RADIATION-DOSIMETRY IN WATER AND IN SOLID PHANTOM MATERIALS FOR I-125 AND PD-103 BRACHYTHERAPY SOURCES - EGS4 MONTE-CARLO STUDY
    LUXTON, G
    [J]. MEDICAL PHYSICS, 1994, 21 (05) : 631 - 641
  • [14] Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source
    Medich, David C.
    Munro, John J., III
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 1999 - 2006
  • [15] Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources
    Melhus, Christopher S.
    Rivard, Mark J.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1729 - 1737
  • [16] Cancer nanotechnology: application of nanotechnology in cancer therapy
    Misra, Ranjita
    Acharya, Sarbari
    Sahoo, Sanjeeb K.
    [J]. DRUG DISCOVERY TODAY, 2010, 15 (19-20) : 842 - 850
  • [17] Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques
    Palmer, Antony
    Bradley, David
    Nisbet, Andrew
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (02) : 81 - 91
  • [18] Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy
    Palmer, Antony
    Hayman, Orla
    Muscat, Sarah
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (01) : 52 - 59
  • [19] Dosimetry characteristics of the Plus and 12i Gammamed PDR 192Ir sources
    Pérez-Calatayud, J
    Ballester, F
    Serrano-Andrés, MA
    Puchades, V
    Lluch, JL
    Limami, Y
    Casal, F
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2576 - 2585
  • [20] Supplement to the 2004 update of the AAPM task group no. 43 report
    Rivard, Mark J.
    Butler, Wayne M.
    DeWerd, Larry A.
    Huq, M. Saiful
    Ibbott, Geoffrey S.
    Meigooni, Ali S.
    Melhus, Christopher S.
    Mitch, Michael G.
    Nath, Ravinder
    Williamson, Jeffrey F.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2187 - 2205