Development of a fiber-coupled laser-induced breakdown spectroscopy instrument for analysis of underwater debris in a nuclear reactor core

被引:78
|
作者
Saeki, Morihisa [1 ]
Iwanade, Akio [1 ]
Ito, Chikara [2 ]
Wakaida, Ikuo [3 ]
Thornton, Blair [4 ]
Sakka, Tetsuo [5 ]
Ohba, Hironori [1 ]
机构
[1] Japan Atom Energy Agcy, Sect Nucl Sci Res, Quantum Beam Sci Ctr, Tokai, Ibaraki 3191195, Japan
[2] Japan Atom Energy Agcy, Oarai Res & Dev Ctr, Oarai, Ibaraki 3111393, Japan
[3] Japan Atom Energy Agcy, Sect Nucl Sci Res, Nucl Sci & Engn Ctr, Tokai, Ibaraki 3191195, Japan
[4] Univ Tokyo, Inst Ind Sci, Meguro Ku, Tokyo 1538505, Japan
[5] Kyoto Univ, Grad Sch Engn, Nishikyo Ku, Kyoto 6158510, Japan
关键词
inspection; laser; post-irradiation examination; transportable fiber-coupled LIBS instrument; underwater debris; high-radiation field; single-pulse breakdown with gas flow; double-pulse breakdown; ELEMENTAL ANALYSIS; ABSORPTION-BANDS; SYSTEM; PROBE; SEA;
D O I
10.1080/00223131.2014.917996
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
To inspect the post-accident nuclear core reactor of the TEPCO Fukushima Daiichi nuclear power plant (F1-NPP), a transportable fiber-coupled laser-induced breakdown spectroscopy (LIBS) instrument has been developed. The developed LIBS instrument was designed to analyze underwater samples in a highradiation field by single-pulse breakdown with gas flow or double-pulse breakdown. To check the feasibility of the assembled fiber-coupled LIBS instrument for the analysis of debris material (mixture of the fuel core, fuel cladding, construction material and so on) in the F1-NPP, we investigated the influence of the radiation dose on the optical transmittance of the laser delivery fiber, compared data quality among various LIBS techniques for an underwater sample and studied the feasibility of the fiber-coupled LIBS system in an analysis of the underwater sample of the simulated debris in F1-NPP. In a feasible study conducted by using simulated debris, which was a mixture of CeO2 (surrogate of UO2), ZrO2 and Fe, we selected atomic lines suitable for the analysis of materials, and prepared calibration curves for the component elements. The feasible study has guaranteed that the developed fiber-coupled LIBS system is applicable for analyzing the debris materials in the F1-NPP.
引用
收藏
页码:930 / 938
页数:9
相关论文
共 50 条
  • [21] Coal property analysis using laser-induced breakdown spectroscopy
    Yuan, Tingbi
    Wang, Zhe
    Lui, Siu-Lung
    Fu, Yangting
    Li, Zheng
    Liu, Jianming
    Ni, Weidou
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (07) : 1045 - 1053
  • [22] Pressure effects on underwater laser-induced breakdown spectroscopy: an interpretation with self-absorption
    Zhang, Yongquan
    Tian, Ye
    Lu, Yuan
    Guo, Lin
    Li, Ying
    Guo, Jinjia
    Zheng, Ronger
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2021, 36 (03) : 644 - 653
  • [23] Single-Pulse Underwater Laser-Induced Breakdown Spectroscopy with Nongated Detection Scheme
    Matsumoto, Ayumu
    Tamura, Ayaka
    Fukami, Kazuhiro
    Ogata, Yukio H.
    Sakka, Tetsuo
    ANALYTICAL CHEMISTRY, 2013, 85 (08) : 3807 - 3811
  • [24] Elemental analysis of materials in an underwater archeological shipwreck using a novel remote laser-induced breakdown spectroscopy system
    Guirado, Salvador
    Fortes, Francisco J.
    Javier Laserna, J.
    TALANTA, 2015, 137 : 182 - 188
  • [25] Quantitative Analysis of Nitrogen in Compound Fertilizers Using Laser-Induced Breakdown Spectroscopy Coupled with Multivariate Regression
    Lu, C.
    Shi, Ch
    Dai, H.
    Lv, G.
    Zhang, Zh
    Jin, F.
    Hu, B.
    JOURNAL OF APPLIED SPECTROSCOPY, 2022, 89 (04) : 705 - 711
  • [26] Quantitative Analysis of P in Compound Fertilizer by Laser-Induced Breakdown Spectroscopy Coupled with Linear Multivariate Calibration
    Sha Wen
    Li Jiang-tao
    Lu Cui-ping
    Zhen Chun-hou
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39 (06) : 1958 - 1964
  • [27] Elemental Analysis of Black Salt by Laser-Induced Breakdown Spectroscopy and Inductively Coupled Plasma-Optical Emission Spectroscopy
    Qasim, M.
    Anwar-ul-Haq, M.
    Afgan, M. Sher
    Kalyar, M. A.
    Baig, M. A.
    ANALYTICAL LETTERS, 2016, 49 (13) : 2108 - 2118
  • [28] Development of a 3D Laser Scanning System for Laser-induced Breakdown Spectroscopy
    Ikezawa, Satoshi
    Zimin, Yury L'vovich
    Ueda, Toshitsugu
    2014 IEEE SENSORS, 2014,
  • [29] In situ Analysis of Fireworks Using Laser-Induced Breakdown Spectroscopy and Chemometrics
    Awasthi, S.
    Kumar, R.
    Rai, A. K.
    JOURNAL OF APPLIED SPECTROSCOPY, 2017, 84 (05) : 811 - 815
  • [30] The effect of solution temperature on the quantitative analysis of laser-induced breakdown spectroscopy
    Lin, Jingjun
    Yang, Jiangfei
    Gao, Xun
    Huang, Yutao
    Lin, Xiaomei
    APPLIED PHYSICS B-LASERS AND OPTICS, 2022, 128 (07):