Urinary catheter capable of repeated on-demand removal of infectious biofilms via active deformation

被引:27
作者
Levering, Vrad [1 ,2 ]
Cao, Changyong [1 ,3 ]
Shivapooja, Phanindhar [1 ,2 ]
Levinson, Howard [4 ]
Zhao, Xuanhe [1 ,3 ,5 ]
Lopez, Gabriel P. [1 ,2 ,3 ]
机构
[1] Duke Univ, Res Triangle Mat Res Sci & Engn Ctr, Durham, NC 27708 USA
[2] Duke Univ, Dept Biomed Engn, Durham, NC 27708 USA
[3] Duke Univ, Dept Mech Engn & Mat Sci, Durham, NC 27708 USA
[4] Duke Univ, Sch Med, Dept Surg, Durham, NC 27710 USA
[5] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
关键词
Urinary catheter; Biofilm removal; Deformation; Infection; Finite element model; PROTEUS-MIRABILIS; ESCHERICHIA-COLI; MEDICAL DEVICES; STAPHYLOCOCCUS-AUREUS; CRYSTALLINE BIOFILMS; URETHRAL CATHETERS; BACTERIAL BIOFILMS; DISEASES SOCIETY; TRACT-INFECTIONS; FOLEY CATHETERS;
D O I
10.1016/j.biomaterials.2015.10.070
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Biofilm removal from biomaterials is of fundamental importance, and is especially relevant when considering the problematic and deleterious impact of biofilm infections on the inner surfaces of urinary catheters. Catheter-associated urinary tract infections are the most common cause of hospital-acquired infections and there are over 30 million Foley urinary catheters used annually in the USA. In this paper, we present the design and optimization of urinary catheter prototypes capable of on-demand removal of biofilms from the inner luminal surface of catheters. The urinary catheters utilize 4 intra-wall inflation lumens that are pressure-actuated to generate region-selective strains in the elastomeric urine lumen, and thereby remove overlying biofilms. A combination of finite-element modeling and prototype fabrication was used to optimize the catheter design to generate greater than 30% strain in the majority of the luminal surface when subjected to pressure. The catheter prototypes are able to remove greater than 80% of a mixed community biofilm of Proteus mirabilis and Escherichia coli on-demand, and furthermore are able to remove the biofilm repeatedly. Additionally, experiments with the prototypes demonstrate that biofilm debonding can be achieved upon application of both tensile and compressive strains in the inner surface of the catheter. The fouling-release catheter offers the potential for a non-biologic, non-antibiotic method to remove biofilms and thereby for impacting the thus far intractable problem of catheter-associated infections. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:77 / 86
页数:10
相关论文
共 38 条
[1]   Biofilm cohesiveness measurement using a novel atomic force microscopy methodology [J].
Ahimou, Francois ;
Semmens, Michael J. ;
Novak, Paige J. ;
Haugstad, Greg .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2007, 73 (09) :2897-2904
[2]   A review of the clinical implications of anti-infective biomaterials and infection-resistant surfaces [J].
Campoccia, Davide ;
Montanaro, Lucio ;
Arciola, Carla Renata .
BIOMATERIALS, 2013, 34 (33) :8018-8029
[3]   Direct measurement of the adhesive strength of biofilms in pipes by micromanipulation [J].
Chen, MJ ;
Zhang, Z ;
Bott, TR .
BIOTECHNOLOGY TECHNIQUES, 1998, 12 (12) :875-880
[4]   Impact of engineered surface microtopography on biofilm formation of Staphylococcus aureus [J].
Chung, Kenneth K. ;
Schumacher, James F. ;
Sampson, Edith M. ;
Burne, Robert A. ;
Antonelli, Patrick J. ;
Brennana, Anthony B. .
BIOINTERPHASES, 2007, 2 (02) :89-94
[5]   Biofilm in implant infections: Its production and regulation [J].
Costerton, JW ;
Montanaro, L ;
Arciola, CR .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (11) :1062-1068
[6]   Systematic review of antimicrobial drug prescribing in hospitals [J].
Davey, P ;
Brown, E ;
Fenelon, L ;
Finch, R ;
Gould, I ;
Holmes, A ;
Ramsay, C ;
Taylor, E ;
Wiffen, P ;
Wilcox, M .
EMERGING INFECTIOUS DISEASES, 2006, 12 (02) :211-216
[7]   Infectious Diseases Society of America and the Society for Healthcare Epidemiology of America guidelines for developing an institutional program to enhance antimicrobial stewardship [J].
Dellit, Timothy H. ;
Owens, Robert C. ;
McGowan, John E., Jr. ;
Gerding, Dale N. ;
Weinstein, Robert A. ;
Burke, John P. ;
Huskins, W. Charles ;
Paterson, David L. ;
Fishman, Neil O. ;
Carpenter, Christopher F. ;
Brennan, P. J. ;
Billeter, Marianne ;
Hooton, Thomas M. .
CLINICAL INFECTIOUS DISEASES, 2007, 44 (02) :159-177
[8]   Biofilm formation: A clinically relevant microbiological process [J].
Donlan, RM .
CLINICAL INFECTIOUS DISEASES, 2001, 33 (08) :1387-1392
[9]   Advances in Microbial Biofilm Prevention on Indwelling Medical Devices with Emphasis on Usage of Acoustic Energy [J].
Dror, Naama ;
Mandel, Mathilda ;
Hazan, Zadik ;
Lavie, Gad .
SENSORS, 2009, 9 (04) :2538-2554
[10]   Control of bacterial biofilm growth on surfaces by nanostructural mechanics and geometry [J].
Epstein, A. K. ;
Hochbaum, A. I. ;
Kim, Philseok ;
Aizenberg, J. .
NANOTECHNOLOGY, 2011, 22 (49)