Dose-Dependent Suppression of the Electrically Elicited Stapedius Reflex by General Anesthetics in Children Undergoing Cochlear Implant Surgery

被引:28
作者
Crawford, Mark W. [1 ]
White, Michelle C. [1 ]
Propst, Evan J. [2 ]
Zaarour, Christian [1 ]
Cushing, Sharon [2 ]
Pehora, Carolyne [1 ]
James, Adrian L. [2 ]
Gordon, Karen A. [2 ]
Papsin, Blake C. [2 ]
机构
[1] Univ Toronto, Hosp Sick Children, Dept Anesthesia & Pain Med, Toronto, ON M5G 1X8, Canada
[2] Univ Toronto, Hosp Sick Children, Dept Otolaryngol Head & Neck Surg, Toronto, ON M5G 1X8, Canada
关键词
MINIMUM ALVEOLAR CONCENTRATION; STIMULATION LEVELS; EVOKED-POTENTIALS; ISOFLURANE; PROPOFOL; SEVOFLURANE; THRESHOLD; VOLATILE; INFANTS; DOSAGE;
D O I
10.1213/ane.0b013e31819bdfd5
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Cochlear implants stimulate the auditory nerve to enable hearing. Determining appropriate upper and lower limits of stimulation is essential for successful cochlear implantation. The intraoperative evoked stapedius reflex threshold (ESRT) and evoked compound action potential (ECAP) are commonly used to determine the limits of implant stimulation. In this study, we evaluated the dose-related effects of sevoflurane, desflurane, isoflurane, and propofol on the intraoperative ESRT and ECAP. METHODS: Forty-four children aged 6 mo to 17 yr undergoing cochlear implantation were recruited. Each child was randomly assigned to receive sevoflurane, desflurane, isoflurane, or propofol. Evoked responses were measured by a blinded investigator at end-tidal anesthetic concentrations corresponding to 0, 0.75, and 1.5 age-adjusted minimum alveolar concentration administered in random sequence and at targeted blood concentrations of propofol of 0, 1.5, and 3.0 mu g/mL. Data were analyzed using one-way repeated-measures analysis of variance. P < 0.05 was considered statistically significant. RESULTS: The ESRT increased dose dependently with increasing volatile anesthetic concentration (P < 0.01). The stapedius reflex was completely abolished by volatile anesthesia in more than half of children. Propofol minimally affected the ESRT. In contrast, the ECAP was unaffected by anesthesia. CONCLUSIONS: Volatile anesthetics suppress the stapedius reflex in a dose-dependent manner, suggesting that ESRT measurements acquired during volatile anesthesia will overestimate the maximum comfort level, which may cause discomfort postoperatively and adversely affect the child's adaptation to the implant. We advise against the use of volatile anesthetics for measurement of the stapedius reflex threshold during cochlear implant surgery.
引用
收藏
页码:1480 / 1487
页数:8
相关论文
共 30 条
  • [1] Almqvist B, 2000, AUDIOLOGY, V39, P146
  • [2] Pharmacologc and physiologic influences affecting sensory evoked potentials - Implications for perioperative monitoring
    Banoub, M
    Tetzlaff, JE
    Schubert, A
    [J]. ANESTHESIOLOGY, 2003, 99 (03) : 716 - 737
  • [3] Influence of volatile and intravenous anesthetics on the threshold of the acoustically evoked stapedius reflex
    Bissinger, U
    Plinkert, PK
    Sesterhenn, G
    Grimm, A
    Lenz, G
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2000, 257 (07) : 349 - 354
  • [4] NEURONAL ORGANIZATION OF ACOUSTIC MIDDLE-EAR REFLEX - PHYSIOLOGICAL AND ANATOMICAL STUDY
    BORG, E
    [J]. BRAIN RESEARCH, 1973, 49 (01) : 101 - 123
  • [5] EVIDENCE FOR DIRECT ACTIONS OF GENERAL-ANESTHETICS ON AN ION-CHANNEL PROTEIN - A NEW LOOK AT A UNIFIED MECHANISM OF ACTION
    DILGER, JP
    VIDAL, AM
    MODY, HI
    LIU, Y
    [J]. ANESTHESIOLOGY, 1994, 81 (02) : 431 - 442
  • [6] DILGER JP, 1992, MOL PHARMACOL, V41, P127
  • [7] Analysis of the EEG bispectrum, auditory evoked potentials and the EEG power spectrum during repeated transitions from consciousness to unconsciousness
    Gajraj, RJ
    Doi, M
    Mantzaridis, H
    Kenny, GNC
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1998, 80 (01) : 46 - 52
  • [8] THE EFFECT OF GENERAL-ANESTHESIA ON INTRAOPERATIVE ELECTRICALLY-EVOKED STAPEDIUS REFLEX MEASUREMENT
    GNADEBERG, D
    BATTMER, RD
    LULLWITZ, E
    LASZIG, R
    DYBUS, U
    LENARZ, T
    [J]. LARYNGO-RHINO-OTOLOGIE, 1994, 73 (03) : 132 - 135
  • [9] Gordon K, 2004, INT J AUDIOL, V43, pS28
  • [10] Toward a battery of behavioral and objective measures to achieve optimal cochlear implant stimulation levels in children
    Gordon, KA
    Papsin, BC
    Harrison, RV
    [J]. EAR AND HEARING, 2004, 25 (05) : 447 - 463