Design of a Wireless Intraocular Pressure Monitoring System for a Glaucoma Drainage Implant

被引:0
|
作者
Kakaday, T. [1 ]
Plunkett, M. [2 ]
McInnes, S. [3 ]
Li, J. S. Jimmy [1 ]
Voelcker, N. H. [3 ]
Craig, J. E. [4 ]
机构
[1] Flinders Univ S Australia, Sch Comp Sci Engn & Math, Sturt Rd,Bedford Pk, Adelaide, SA 5042, Australia
[2] Ellex Med Lasers R&D, Adelaide, SA, Australia
[3] Flinders Univ S Australia, Sch Chem Phys & Earth Sci, Adelaide, SA, Australia
[4] Flinders Med Ctr, Dept Ophthalmol, Adelaide, SA, Australia
来源
13TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, VOLS 1-3 | 2009年 / 23卷 / 1-3期
关键词
Glaucoma; intraocular pressure; glaucoma drainage implant; micro-electro-mechanical systems (MEMS); and capacitive pressure sensor; SENSOR;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Glaucoma is a common cause of blindness. Wireless, continuous monitoring of intraocular pressure (IOP) is an important, unsolved goal in managing glaucoma. An IOP monitoring system incorporated into a glaucoma drainage implant (GDI) overcomes the design complexity with incorporating a similar system in a more confined space within the eye. The device consists of a micro-electro-mechanical systems (MEMS) based capacitive pressure sensor combined with an inductor printed directly onto a polyimide printed circuit board (PCB). The device is designed to be placed onto the external plate of a therapeutic GDI. The resonance frequency changes as a function of IOP and is tracked remotely using a spectrum analyzer. A theoretical model for the reader antenna was developed to enable maximal inductive coupling with the IOP sensor implant, including modeling of high frequency effects. Pressure chamber tests indicate that the device has adequate sensitivity in the IOP range with excellent reproducibility over time. Additionally, we show that sensor sensitivity does not change significantly after encapsulation with polydimethylsiloxane (PDMS) to protect the device from the aqueous environment. In vitro experiments showed that the signal measured wirelessly through sheep corneal tissue was adequate indicating potential for using the system in human subjects.
引用
收藏
页码:198 / +
页数:2
相关论文
共 50 条
  • [21] Intraocular pressure in non penetrating deep sclerectomy with SkGel implant in patients with primary angle glaucoma
    Almodin, Juliana
    Santos Picosse, Claudio Rabelo
    Cvintal, Tadeu
    Almodin, Flavia
    REVISTA BRASILEIRA DE OFTALMOLOGIA, 2010, 69 (06) : 383 - 388
  • [22] In vitro and in vivo evaluation of an intraocular implant for glaucoma treatment
    Natu, Madalina V.
    Gaspar, Manuel N.
    Fontes Ribeiro, Carlos A.
    Cabrita, Antonio M.
    de Sousa, Herminio C.
    Gil, M. H.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2011, 415 (1-2) : 73 - 82
  • [23] Impact of Drugs on Glaucoma and Intraocular Pressure
    Schlote, T.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2017, 234 (02) : 179 - 184
  • [24] The Effect of Exercise on Intraocular Pressure and Glaucoma
    Gildea, David
    Doyle, Aoife
    O'Connor, Jeremy
    JOURNAL OF GLAUCOMA, 2024, 33 (06) : 381 - 386
  • [25] Advances in diagnostic applications for monitoring intraocular pressure in Glaucoma: A review
    Sanchez, Irene
    Martin, Raul
    JOURNAL OF OPTOMETRY, 2019, 12 (04) : 211 - 221
  • [26] Cataract surgery and intraocular pressure in glaucoma
    Young, Cara E. Capitena
    Seibold, Leonard K.
    Kahook, Malik Y.
    CURRENT OPINION IN OPHTHALMOLOGY, 2020, 31 (01) : 15 - 22
  • [27] Head Elevation and Intraocular Pressure in Glaucoma
    Park, Ji-Hye
    Nam, Ki Tae
    Yoo, Chungkwon
    Kim, Yong Yeon
    OPTOMETRY AND VISION SCIENCE, 2016, 93 (09) : 1163 - 1170
  • [28] Relationship of systemic blood pressure with ocular perfusion pressure and intraocular pressure of glaucoma patients in telemedical home monitoring
    Juergens, Clemens
    Grossjohann, Rico
    Tost, Frank H. W.
    MEDICAL SCIENCE MONITOR, 2012, 18 (11): : MT85 - MT89
  • [29] Magneto-elastic for wireless intraocular pressure monitoring
    Lin, Yu
    Xiang, Shengxin
    Zou, Wenjun
    Chen, Xiaohua
    Gao, Feixiang
    Liu, Yi
    Sun, Yijie
    Chen, Ping
    Zhang, Quanxi
    Han, Xuelian
    Wang, Baoping
    Yao, Yong
    Shi, Qiongfeng
    Wu, Jun
    CHEMICAL ENGINEERING JOURNAL, 2024, 498
  • [30] Intraocular pressure - All that is raised is not glaucoma
    Tripathi, A
    O'Donnell, NP
    OPHTHALMOLOGICA, 2001, 215 (05) : 376 - 377