Repeatability and comparability of corneal power and corneal astigmatism obtained from a point-source color light-emitting diode topographer, a Placido-based corneal topographer, and a low-coherence reflectometer

被引:28
作者
Ventura, Bruna V. [1 ,2 ,3 ]
Al-Mohtaseb, Zaina [1 ]
Wang, Li [1 ]
Koch, Douglas D. [1 ]
Weikert, Mitchell P. [1 ]
机构
[1] Baylor Coll Med, Dept Ophthalmol, Cullen Eye Inst, Houston, TX 77030 USA
[2] Altino Ventura Fdn, Recife, PE, Brazil
[3] Univ Fed Sao Paulo, Dept Ophthalmol, Sao Paulo, Brazil
关键词
AUTOMATED KERATOMETRY; REFRACTIVE SURGERY; AXIS LOCATION; ERROR; AGREEMENT; REPRODUCIBILITY; CATARACT; BIOMETRY; EYES;
D O I
10.1016/j.jcrs.2015.11.003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the repeatability and agreement of corneal power and astigmatism obtained from the Cassini point-source color light emitting diode (LED) topographer, Humphrey Atlas 9000 Placido-based corneal topographer, and Lenstar LS-900 low-coherence reflectometer in normal eyes. SETTING: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA. DESIGN: Evaluation of diagnostic test or technology. METHODS: Consecutively enrolled patients with normal corneas were enrolled. Three sets of measurements were obtained using the color-LED topographer, the Placido topographer, and the reflectometer. Vector analysis was used in the astigmatism analysis. The repeatability was evaluated using the within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was verified using Bland-Altman plots. The paired Student t test was used to assess statistical significance. RESULTS: Thirty-two eyes (32 patients) were evaluated. All devices provided highly repeatable corneal power and astigmatism measurements (ICC > 0.9) except for the Placido topographer with regard to J45 (ICC = 0.721). The color-LED topographer and the reflectometer obtained similar mean values of corneal power, astigmatism magnitude, J0, and J45 (P > .05), which was also true when comparing the color-LED topographer and the Placido topographer, except for the mean corneal power (P = .0007). The Bland-Altman plots showed a wide data spread for all analyzed variables. CONCLUSIONS: The color-LED topographer provided highly repeatable corneal power and astigmatism measurements. Even though it obtained values similar to those of the reflectometer and the Placido topographer, the wide data spread discourages their interchangeable use to assess corneal power and astigmatism. (C) 2015 ASCRS and ESCRS
引用
收藏
页码:2242 / 2250
页数:9
相关论文
共 23 条
[1]   Practical astigmatism analysis for refractive outcomes in cataract and refractive surgery [J].
Alpins, NA ;
Goggin, M .
SURVEY OF OPHTHALMOLOGY, 2004, 49 (01) :109-122
[2]   Dual versus single Scheimpflug camera for anterior segment analysis: Precision and agreement [J].
Aramberri, Jaime ;
Araiz, Luis ;
Garcia, Ane ;
Illarramendi, Igor ;
Olmos, Jaione ;
Oyanarte, Izaskun ;
Romay, Amaya ;
Vigara, Itxaso .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (11) :1934-1949
[3]  
Bland JM, 1996, BRIT MED J, V313, P744
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting [J].
Cho, P ;
Lam, AKC ;
Mountford, J ;
Ng, L .
OPTOMETRY AND VISION SCIENCE, 2002, 79 (03) :175-183
[6]   Comparison and Repeatability of Keratometric and Corneal Power Measurements Obtained by Orbscan II, Pentacam, and Galilei Corneal Tomography Systems [J].
Crawford, Alexandra Z. ;
Patel, Dipika V. ;
McGhee, Charles N. J. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 156 (01) :53-60
[7]   Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism [J].
Holladay, JT ;
Moran, JR ;
Kezirian, GM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (01) :61-79
[8]   Evaluating and reporting astigmatism for individual and aggregate data [J].
Holladay, JT ;
Dudeja, DR ;
Koch, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (01) :57-65
[9]   Comparison of corneal power, corneal astigmatism, and axis location in normal eyes obtained from an autokeratometer and a corneal topographer [J].
Kobashi, Hidenaga ;
Kamiya, Kazutaka ;
Igarashi, Akihito ;
Ishii, Rie ;
Sato, Nobuyuki ;
Wang, Guoqin ;
Shimizu, Kimiya .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04) :648-654
[10]   Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer [J].
Mao, Xinjie ;
Savini, Giacomo ;
Zhuo, Zuopao ;
Feng, Yifan ;
Zhang, Jian ;
Wang, Qinmei ;
Chen, Hao ;
Huang, Jinhai .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10) :1561-1569