Quantification and reduction of ghosting artifacts in interleaved echo-planar imaging

被引:45
作者
Reeder, SB [1 ]
Atalar, E [1 ]
Bolster, BD [1 ]
McVeigh, ER [1 ]
机构
[1] JOHNS HOPKINS UNIV,SCH MED,BALTIMORE,MD 21205
关键词
MRI; artifacts; reconstruction; EPI; functional imaging;
D O I
10.1002/mrm.1910380312
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A mathematical analysis of ghosting artifacts often seen in interleaved echo-planar images (EPI) is presented. These artifacts result from phase and amplitude discontinuities between lines of k-space in the phase-encoding direction, and timing misregistrations from system filter delays. Phase offsets and time delays are often measured using ''reference'' scans, to reduce ghosting through postprocessing. From the expressions describing ghosting artifacts, criteria were established for reducing ghosting to acceptable levels. Subsequently, the signal-to-noise ratio (SNR) requirements for estimation of time delays acid phase offsets, determined from reference scans, was evaluated to establish the effect of estimation error on artifact reduction for interleaved EPI, Artifacts resulting from these effects can be reduced to very low levels when appropriate reference scan estimation is used. This has important implications for functional MRI (fMRI) and applications involving small changes in signal intensity.
引用
收藏
页码:429 / 439
页数:11
相关论文
共 27 条
[1]   ANALYSIS OF EDDY CURRENTS IN NUCLEAR-MAGNETIC-RESONANCE IMAGING [J].
AHN, CB ;
CHO, ZH .
MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (01) :149-163
[2]  
[Anonymous], 1980, Gos. Izd-vo Fiz.-Mat. Literatury
[3]   OPTIMIZATION OF TAG THICKNESS FOR MEASURING POSITION WITH MAGNETIC-RESONANCE-IMAGING [J].
ATALAR, E ;
MCVEIGH, ER .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :152-160
[4]   CINEANGIOGRAPHY OF THE HEART IN A SINGLE BREATH HOLD WITH A SEGMENTED TURBOFLASH SEQUENCE [J].
ATKINSON, DJ ;
EDELMAN, RR .
RADIOLOGY, 1991, 178 (02) :357-360
[5]   IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING [J].
BRUDER, H ;
FISCHER, H ;
REINFELDER, HE ;
SCHMITT, F .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) :311-323
[6]   SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS [J].
CONTURO, TE ;
SMITH, GD .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :420-437
[7]   QUALITATIVE MAPPING OF CEREBRAL BLOOD-FLOW AND FUNCTIONAL LOCALIZATION WITH ECHO-PLANAR MR-IMAGING AND SIGNAL TARGETING WITH ALTERNATING RADIO-FREQUENCY [J].
EDELMAN, RR ;
SIEWERT, B ;
DARBY, DG ;
THANGARAJ, V ;
NOBRE, AC ;
MESULAM, MM ;
WARACH, S .
RADIOLOGY, 1994, 192 (02) :513-520
[8]   ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING [J].
FARZANEH, F ;
RIEDERER, SJ ;
PELC, NJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :123-139
[9]   PHASE ERRORS IN MULTISHOT ECHO-PLANAR IMAGING [J].
FEINBERG, DA ;
OSHIO, K .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :535-539
[10]   GRADIENT-ECHO SHIFTING IN FAST MRI TECHNIQUES (GRASE IMAGING) FOR CORRECTION OF FIELD INHOMOGENEITY ERRORS AND CHEMICAL-SHIFT [J].
FEINBERG, DA ;
OSHIO, K .
JOURNAL OF MAGNETIC RESONANCE, 1992, 97 (01) :177-183