Evaluation of SRS MapCHECK with StereoPHAN phantom as a new pre-treatment system verification for SBRT plans

被引:1
作者
Sadowski, Bartlomiej Mieszko [1 ]
Fillmann, Marta [1 ]
Szalkowski, Dariusz [1 ]
Kukolowicz, Pawel [1 ]
机构
[1] Maria Sklodowska Curie Natl Res Inst Oncol, Dept Med Phys, Warsaw, Poland
关键词
dosimetry; quality assurance; stereotactic radiotherapy; STEREOTACTIC RADIOSURGERY; EBT3; FILM; CALIBRATION; QUALITY;
D O I
10.2478/pjmpe-2022-0010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: The aim of this study was to evaluate the new 2-Dimensional diode array SRS MapCHECK (SunNuclear, Melbourne, USA) with dedicated phantom StereoPHAN (SunNuclear, Melbourne, USA) for the pre-treatment verification of the stereotactic body radiotherapy (SBRT). Material and methods: For the system, the short and mid-long stability, dose linearity with MU, angular dependence, and field size dependence (ratio of relative output factor) were measured. The results of verification for 15 pre-treatment cancer patients (5 brains, 5 lungs, and 5 livers) performed with SRS MapCHECK and EBT3 Gafchromic films were compared. All the SBRT plans were optimized with the Eclipse (v. 15.6, Varian, Palo Alto, USA) treatment planning system (TPS) using the Acuros XB (Varian, Palo Alto, USA) dose calculation algorithm and were delivered to the Varian EDGE (R) (Varian, Palo Alto, USA) accelerator equipped with a high-definition multileaf collimator. The 6MV flattening-filter-free beam (FFF) was used. Results: Short and mid-long stability of SRS MapCHECK was very good (0.1%-0.2%), dose linearity with MU and dependence of the response of the detector on field size results were also acceptable (for dose linearity R-2 = 1 and 6% difference between microDiamond and SRS MapCHECK response for the smallest field of 1 x 1 cm(2)). The angular dependence was very good except for the angles close to 90 degrees and 270 degrees. For pre-treatment plan verification, the gamma method was used with the criteria of 3% dose difference and 3 mm distance to agreement (3%/3 mm), and 2%/2 mm, 1%/1 mm, 3%/1 mm, and 2%/1 mm. The highest passing rate for all criteria was observed on the SRS MapCHECK system. Conclusions: It is concluded that SRS MapCHECK with StereoPHAN has sufficient potential for pre-treatment verification of the SBRT plans, so that verification of stereotactic plans can be significantly accelerated.
引用
收藏
页码:84 / 89
页数:6
相关论文
共 14 条
[1]   Comprehensive evaluation of the high-resolution diode array for SRS dosimetry [J].
Ahmed, Saeed ;
Zhang, Geoffrey ;
Moros, Eduardo G. ;
Feygelman, Vladimir .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10) :13-23
[2]   Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions [J].
Han, Zhaohui ;
Ng, Sook Kien ;
Bhagwat, Mandar S. ;
Lyatskaya, Yulia ;
Zygmanski, Piotr .
MEDICAL PHYSICS, 2010, 37 (07) :3704-3714
[3]  
International Atomic Energy Agency (IAEA), 2017, TECHNICAL REPORTS SE, V483
[4]   Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance [J].
Jin, Hosang ;
Keeling, Vance P. ;
Johnson, Daniel A. ;
Ahmad, Salahuddin .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (03) :80-92
[5]   Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density [J].
Leon Marroquin, Elsa Y. ;
Herrera Gonzalez, Jose A. ;
Camacho Lopez, Miguel A. ;
Villarreal Barajas, Jose E. ;
Garcia-Garduno, Olivia A. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05) :466-481
[6]   An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan [J].
Lewis, David ;
Micke, Andre ;
Yu, Xiang ;
Chan, Maria F. .
MEDICAL PHYSICS, 2012, 39 (10) :6339-6350
[7]   Point dose verification of Cranial Stereotactic Radiosurgery using micro Ionization Chamber and EBT3 film for 6MV FF and FFF beams in Varian TrueBeam® LINAC [J].
Mamballikalam, Gopinath ;
Senthilkumar, S. ;
Basith, Ahamed P. M. ;
Inipully, Rohit ;
Jayadevan, P. M. ;
Clinto, C. O. ;
Bos, Jaon R. C. .
POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2020, 26 (03) :135-142
[8]   Quality and safety considerations in stereotactic radiosurgery and stereotactic body radiation therapy: Executive summary [J].
Solberg, Timothy D. ;
Balter, James M. ;
Benedict, Stanley H. ;
Fraass, Benedick A. ;
Kavanagh, Brian ;
Miyamoto, Curtis ;
Pawlicki, Todd ;
Potters, Louis ;
Yamada, Yoshiya .
PRACTICAL RADIATION ONCOLOGY, 2012, 2 (01) :2-9
[9]  
SRS MapCHECK TM User Guide, MODEL 1179
[10]   A contemporary review of stereotactic radiotherapy: Inherent dosimetric complexities and the potential for detriment [J].
Taylor, Michael L. ;
Kron, Tomas ;
Franich, Rick D. .
ACTA ONCOLOGICA, 2011, 50 (04) :483-508