Exploiting monosegmented flow analysis to perform in-line standard additions using a single stock standard solution in spectrophotornetric sequential injection procedures

被引:25
作者
Silva, MSP [1 ]
Masini, JC [1 ]
机构
[1] Univ Sao Paulo, Inst Quim, BR-05513970 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
flow analysis; automation; spectrophotometry; iron; standard addition;
D O I
10.1016/S0003-2670(02)00594-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The robustness of sequential injection analysis (SIA) was combined with the monosegmented flow analysis (MSFA) approach, in which there is no dispersion of the reaction zone with carrier, to develop a methodology to perform in-line dilution. This approach allows one to know accurately the dilution of sample and reagent inside the monosegment, without the need for determination of dispersion coefficients. As a consequence, the methodology allowed the mechanization of procedures to perform standard additions and to construct analytical curves using a single stock standard solution, with very simple and conventional computation of the sample concentration. The method was illustrated with experiments using the bromothymol blue (BTB) dye, in which no reactions are involved, as well as with the spectrophotometric methodology for determination of Fe(II) using o-1,10-phenanthroline as chromogenic reagent. The resulting method presented a sampling frequency of 30 analyses per hour and a detection limit of 25 mug 1(-1). (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:345 / 352
页数:8
相关论文
共 28 条
[1]  
[Anonymous], 1989, FLOW INJECTION ANAL
[2]   A FAST PROCEDURE FOR STANDARD ADDITIONS IN FLOW-INJECTION ANALYSIS [J].
ARAUJO, MCU ;
PASQUINI, C ;
BRUNS, RE ;
ZAGATTO, EAG .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :337-343
[3]   Simultaneous multiple injection to perform titration and standard addition in monosegmented flow analysis [J].
Assali, M ;
Raimundo, IM ;
Facchin, I .
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY, 2001, 23 (03) :83-89
[4]   NOVEL SINGLE STANDARD CALIBRATION AND DILUTION METHOD PERFORMED BY THE SEQUENTIAL INJECTION TECHNIQUE [J].
BARON, A ;
GUZMAN, M ;
RUZICKA, J ;
CHRISTIAN, GD .
ANALYST, 1992, 117 (12) :1839-1844
[5]   Simultaneous multiple injection in monosegmented flow analysis [J].
Brito, VO ;
Raimundo, IM .
ANALYTICA CHIMICA ACTA, 1998, 371 (2-3) :317-+
[6]  
CARBONELL V, 1991, J ANAL ATOM SPECTR, V6, P223
[7]   Sequential injection standard addition for on-line measurement of mercury in the river Elbe [J].
Elsholz, O ;
Frank, C ;
Stachel, B ;
Reincke, H ;
Ebinghaus, R .
ANALYTICA CHIMICA ACTA, 2001, 438 (1-2) :251-258
[8]   EXPANSION OF DYNAMIC-RANGE OF FLAME ATOMIC-ABSORPTION SPECTROMETRY BY AN EFFICIENT FLOW-INJECTION DILUTION SYSTEM BASED ON DISPERSION OF MICROLITER-VOLUME SAMPLES [J].
FANG, ZL ;
WELZ, B ;
SPERLING, M .
ANALYTICAL CHEMISTRY, 1993, 65 (13) :1682-1688
[9]   Sequential injection analysis as a tool for in situ monitoring of Fe(II), Fe(III), NO3- and NO2- in natural and waste waters [J].
Galhardo, CX ;
Masini, JC .
ANALYTICA CHIMICA ACTA, 2001, 438 (1-2) :39-48
[10]   A SIMPLE PROCEDURE FOR STANDARD ADDITIONS IN FLOW-INJECTION ANALYSIS - SPECTROPHOTOMETRIC DETERMINATION OF NITRATE IN PLANT-EXTRACTS [J].
GINE, MF ;
REIS, BF ;
ZAGATTO, EAG ;
KRUG, FJ ;
JACINTHO, AO .
ANALYTICA CHIMICA ACTA, 1983, 155 (DEC) :131-138