Repeatability and reproducibility of measurements obtained by dual-energy X-ray absorptiometry on pig carcasses

被引:11
作者
Kipper, M. [1 ,2 ]
Marcoux, M. [1 ]
Andretta, I. [3 ]
Pomar, C. [1 ]
机构
[1] Agr & Agri Food Canada, Sherbrooke Res & Dev Ctr, Sherbrooke, PQ J1M 0C8, Canada
[2] Univ Fed Santa Maria, Dept Zootecnia, BR-97105900 Santa Maria, RS, Brazil
[3] Univ Fed Rio Grande do Sul, Dept Zootecnia, BR-91540000 Porto Alegre, RS, Brazil
关键词
accuracy; body composition; dual-energy X-ray absorptiometry; precision; region of interest; pigs; SOFT-TISSUE COMPOSITION; BODY-COMPOSITION MEASUREMENTS; BONE; DXA; ACCURACY; WEIGHT;
D O I
10.1093/jas/skx046
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
The precision of a dual-energy X-ray absorptiometry (DXA) device in terms of repeatability and reproducibility was evaluated on nine left half-carcasses from pigs with large variability in body weight and fat content. Repeatability was assessed by scanning each carcass 10 times sequentially in the same position. Reproducibility was assessed by scanning each carcass in 10 different positions. Images were analyzed with DXA software using a custom region of interest (ROI) and the standard head, trunk, arm, and leg ROI. Predicted values from the DEXA for bone mineral content (BMC), bone area, bone mineral density (BMD), total weight, soft-tissue weight, fat-tissue weight, and lean-tissue weight were considered. Repeatability was associated with the variance between measurements on the same carcass in the same position (repeatability conditions). An average variance value was obtained with all the carcasses combined, and the SD was calculated as the square root of this combined variance. The CV was the ratio between the SD of the measurements and its average value. Reproducibility was calculated for each carcass as the difference between the variance obtained under the reproducibility conditions and that obtained under the repeatability conditions. The effects of the ROI and conditions were evaluated by ANOVA and Tukey's test. Means of BMC, bone area, BMD, fat tissue, and lean tissue differed among the ROI (P < 0.05) in both the repeatability and reproducibility conditions. The CV of DXA measurements under repeatability condition obtained in the head, arm, and leg ROI was lesser than 1%. Only the repeatability errors of fat tissue differed (P < 0.05) among the ROI, with the lowest precision found for the trunk ROI. The reproducibility errors of BMC, bone area, fat tissue, and lean tissue differed (P < 0.05) among the ROI. The custom ROI had reproducibility errors greater than 2% for fat tissue and greater than 3.5% for BMC and bone area. In addition, the trunk ROI had the highest reproducibility errors for fat tissue (20.7%) and lean tissue (6.2%) when compared to the other ROI. In conclusion, repeatability and reproducibility results obtained for most of the studied ROI indicate that DXA is a valuable tool for carcass evaluation. From a methodological viewpoint and considering the variations observed in this study, the ROI should be chosen based on the item to be evaluated or on the conditions in which the DXA measurements are to be taken.
引用
收藏
页码:2027 / 2037
页数:11
相关论文
共 29 条
  • [1] Burdick R. K., 2005, DESIGN ANAL GAUGE R, P201
  • [2] Manual segmentation of DXA scan images results in reliable upper and lower extremity soft and rigid tissue mass estimates
    Burkhart, Timothy A.
    Arthurs, Katherine L.
    Andrews, David M.
    [J]. JOURNAL OF BIOMECHANICS, 2009, 42 (08) : 1138 - 1142
  • [3] Canada Pork International, 1995, CAN PORK BUYERS MAN
  • [4] Dual-energy x-ray absorptiometry and body composition: Differences between devices and comparison with reference methods
    Genton, L
    Hans, D
    Kyle, UG
    Pichard, C
    [J]. NUTRITION, 2002, 18 (01) : 66 - 70
  • [5] ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES
    GLUER, CC
    BLAKE, G
    LU, Y
    BLUNT, BA
    JERGAS, M
    GENANT, HK
    [J]. OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) : 262 - 270
  • [6] Hass Z. J., 2014, C APPL STAT AGR 26 A, P115
  • [7] Hologic, 1996, QDR 4500 FAN BEAM XR, P203
  • [8] Dual energy X-ray absorptiometry (DXA) can be used to predict live animal and whole carcass composition of sheep
    Hunter, T. E.
    Suster, D.
    Dunshea, F. R.
    Cummins, L. J.
    Egan, A. R.
    Leury, B. J.
    [J]. SMALL RUMINANT RESEARCH, 2011, 100 (2-3) : 143 - 152
  • [9] ISO, 1993, 35341 ISO, P47
  • [10] Measurement of soft tissue composition by dual energy X-ray absorptiometry
    Jebb, SA
    [J]. BRITISH JOURNAL OF NUTRITION, 1997, 77 (02) : 151 - 163