The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures

被引:10
作者
Delichas, Miltiadis G. [1 ]
Psarrakos, Kyriakos
Hatziioannou, Konstantinos
Giannoglou, Georgios
Molyvda-Athanasopoulou, Elisabeth
Papanastassiou, Emmanouil
Sioundas, Anastasios
机构
[1] Aristotle Univ Thessaloniki, Sch Med, Fac Med, Dept Med Phys, GR-54006 Thessaloniki, Greece
[2] Papageorgiou Hosp, Dept Med Phys, GR-56403 Thessaloniki, Greece
[3] Aristotle Univ Thessaloniki, Sch Med, Dept Cardiol, GR-54006 Thessaloniki, Greece
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2005年 / 21卷 / 04期
关键词
interventional cardiology; complexity; radiation dose;
D O I
10.1016/S1120-1797(05)80004-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the number of catheters used and with the number of vessels with stenosis. DAP values for patients with a prior bypass surgery, were higher compared to those without such a medical record to surgery. In PTCA, the use of coronary stems did not enhance the patient radiation dose significantly. Noticeable differences were found in the percentage contribution of each projection to the total DAP between the three types of single vessel PTCA. Finally low variations in DAP were found among the cardiologists performing both procedures.
引用
收藏
页码:153 / 157
页数:5
相关论文
共 18 条
  • [1] Bakalyar DM, 1997, CATHETER CARDIO DIAG, V42, P121, DOI 10.1002/(SICI)1097-0304(199710)42:2<121::AID-CCD4>3.3.CO
  • [2] 2-Y
  • [3] Bernardi G, 2000, CATHETER CARDIO INTE, V51, P1
  • [4] The impact of cardiology on the collective effective dose in the North of England
    Broadhead, DA
    Chapple, CL
    Faulkner, K
    Davies, ML
    McCallum, H
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (833) : 492 - 497
  • [5] Factors affecting patient radiation exposure during routine coronary angiography in a tertiary referral centre
    Clark, AL
    Brennan, AG
    Robertson, LJ
    McArthur, JD
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (866) : 184 - 189
  • [6] Delichas MG, 2003, RADIAT PROT DOSIM, V103, P149, DOI 10.1093/oxfordjournals.rpd.a006126
  • [7] RADIATION-DOSES AND SOMATIC RISK TO PATIENTS DURING CARDIAC RADIOLOGICAL PROCEDURES
    FAULKNER, K
    LOVE, HG
    SWEENEY, JK
    BARDSLEY, RA
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) : 359 - 363
  • [8] FAULKNER K, 1992, RADIAT PROT DOSIM, V43, P131
  • [9] Patient radiation exposure during coronary angiography and intervention
    Fransson, SG
    Persliden, J
    [J]. ACTA RADIOLOGICA, 2000, 41 (02) : 142 - 144
  • [10] HART D, 2002, NRPBW14