The usefulness and limitations of hand-held refractometers in veterinary laboratory medicine: An historical and technical review

被引:112
作者
George, JW [1 ]
机构
[1] Univ Calif Davis, Sch Vet Med, Dept Pathol Microbiol & Immunol, Davis, CA 95616 USA
关键词
refractometry; serum protein; urine specific gravity;
D O I
10.1111/j.1939-165X.2001.tb00432.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Medical hand-held refractometers have been used in veterinary practice since their development in the 1960s. They have become ubiquitous for the measurement of protein and urine solute concentrations because of their rapidity of analysis, ease of use, and relatively low cost. Refraction of light offers advantages for the determination of solute concentrations because the measurement requires no chemical alteration of the specimen. Numerous authors have reported that the results of protein estimation by refractometry for domestic mammals correlate well with those obtained by the biuret method, although others have reported both higher and lower refractometric results compared with biuret results. Major discrepancies between biuret and refractometric results have been reported for avian samples. Some of the variation in reported results may be due to differences in design by refractometer manufacturers. Another possible source may be variation in the biuret reagent mixture and assay conditions. Refractometers also can be used to calculate serum water concentration. A table that converts index of refraction to serum water concentration can be used to convert electrolyte concentration from mmol/L of serum to mmol/L of serum water, a more accurate indicator of effective electrolyte concentration. Refractometers are especially useful for determining urine specific gravity on veterinary samples because they require relatively small sample volumes. Specific gravity continues to be the most common unit for reporting total solids concentration. Some solutes, such as acetone, may cause false increases in specific gravity by refractometry, as they increase refraction but are less dense than water. (C) 2001 American Society for Veterinary Clinical Pathology.
引用
收藏
页码:201 / 210
页数:10
相关论文
共 53 条
  • [1] *AM OPT CORP, 1964, INSTR US CAR AO TS M
  • [2] *AM OPT CORP, 1976, INSTR US CAR AO TS M
  • [3] *AM OPT CORP, 1966, CONV TABL GOLDB REFR
  • [4] DETERMINATION OF CHICKEN AND TURKEY PLASMA AND SERUM-PROTEIN CONCENTRATIONS BY REFRACTOMETRY AND THE BIURET METHOD
    ANDREASEN, CB
    LATIMER, KS
    KIRCHER, IM
    BROWN, J
    [J]. AVIAN DISEASES, 1989, 33 (01) : 93 - 96
  • [5] *AT CO LTD, 1995, SER PROT REFR SPR NE
  • [6] BARRY KG, 1960, J LAB CLIN MED, V55, P803
  • [7] INACCURACY IN NEONATAL MEASUREMENT OF URINE CONCENTRATION WITH A REFRACTOMETER
    BENITEZ, OA
    BENITEZ, M
    STIJNEN, T
    BOOT, W
    BERGER, HM
    [J]. JOURNAL OF PEDIATRICS, 1986, 108 (04) : 613 - 616
  • [8] BOVEE KC, 1969, J AM VET MED ASSOC, V155, P30
  • [9] Comparison of four methods for determination of total protein concentrations in pleural and peritoneal fluid from dogs
    Braun, JP
    Guelfi, JF
    Pagès, JP
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2001, 62 (03) : 294 - 296
  • [10] Bumpus S. E., 1998, Journal of Veterinary Emergency & Critical Care, V8, P21, DOI 10.1111/j.1476-4431.1998.tb00131.x