Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating

被引:170
作者
Baker, KB
Tkach, JA
Nyenhuis, JA
Phillips, M
Shellock, FG
Gonzalez-Martinez, J
Rezai, AR
机构
[1] Cleveland Clin Fdn, Dept Neurol, Cleveland, OH 44195 USA
[2] Cleveland Clin Fdn, Dept Radiol, Cleveland, OH 44195 USA
[3] Purdue Univ, Dept Comp & Elect Engn, W Lafayette, IN 47907 USA
[4] Univ So Calif, Keck Sch Med, Los Angeles, CA 90089 USA
[5] Cleveland Clin Fdn, Dept Neurosurg, Sect Stereotact & Funct Neurosurg, Cleveland, OH 44195 USA
关键词
magnetic resonance imaging; bioeffects; safety; neurostimulation systems; implants; deep brain stimulation; specific absorption rate;
D O I
10.1002/jmri.20103
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the magnetic resonance imaging (MRI)-related heating per unit of whole body averaged specific absorption rate (SAR) of a conductive implant exposed to two different 1.5-Tesla/64 MHz MR systems. Materials and Methods: Temperature changes at the electrode contacts of a deep brain stimulation lead were measured using fluoroptic thermometry. The leads were placed in a typical surgical implant configuration within a gel-filled phantom of the human head and torso. MRI was performed using two different transmit/receive body coils on two different generation 1.5-Tesla MR systems from the same manufacturer. Temperature changes were normalized to whole body averaged SAR values and compared between the two scanners. Results: Depending on the landmark location, the normalized temperature change for the implant was significantly higher on one MR system compared to the other (P<0.001). Conclusion: The findings revealed marked differences across two MR systems in the level of radiofrequency (RF)induced temperature changes per unit of whole body SAR for a conductive implant. Thus, these data suggest that using SAR to guide MR safety recommendations for neurostimulation systems or other similar implants across different MR systems is unreliable and, therefore, potentially dangerous. Better, more universal, measures are required in order to ensure patient safety.
引用
收藏
页码:315 / 320
页数:6
相关论文
共 16 条
  • [1] Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes
    Achenbach, S
    Moshage, W
    Diem, B
    Bieberle, T
    Schibgilla, V
    Bachmann, K
    [J]. AMERICAN HEART JOURNAL, 1997, 134 (03) : 467 - 473
  • [2] *ASTM INT, 2002, STAND TEST METH MEAS
  • [3] Finelli DA, 2002, AM J NEURORADIOL, V23, P1795
  • [4] Magnetic resonance safety testing of a newly-developed fiber-optic cardiac pacing lead
    Greatbatch, W
    Miller, V
    Shellock, FG
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (01) : 97 - 103
  • [5] International Electrotechnical Commission (IEC), 2002, 60601233 IEC
  • [6] *MEDTR, 2001, SOL NEUR DEEP BRAIN
  • [7] Gelled versus nongelled phantom material for measurement of MRI-induced temperature increases with bioimplants
    Park, SM
    Nyenhuis, JA
    Smith, CD
    Lim, EJ
    Foster, KS
    Baker, KB
    Hrdlicka, G
    Rezai, AR
    Ruggieri, P
    Sharan, A
    Shellock, FG
    Stypulkowski, PH
    Tkach, J
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 2003, 39 (05) : 3367 - 3371
  • [8] Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging-related heating at 1.5 tesla
    Rezai, AR
    Finelli, D
    Nyenhuis, JA
    Hrdlicka, G
    Tkach, J
    Sharan, A
    Rugieri, P
    Stypulkowski, PH
    Shellock, FG
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (03) : 241 - 250
  • [9] Schaefer D J, 1998, Magn Reson Imaging Clin N Am, V6, P775
  • [10] Shellock F G, 2000, J Magn Reson Imaging, V12, P214, DOI 10.1002/1522-2586(200008)12:2<214::AID-JMRI2>3.0.CO