Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance

被引:0
作者
Stathakis, Sotirios [1 ]
Mavroidis, Panayiotis [1 ]
Narayanasamy, Ganesh [1 ]
Markovic, Miljenko [1 ]
Myers, Pamela [2 ]
Papanikolaou, Niko [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, San Antonio, TX 78229 USA
[2] Baylor Coll Med, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF BUON | 2015年 / 20卷 / 04期
关键词
DQA; IMRT; measurements at extended SSD; SBRT; ION-CHAMBER ARRAY; IMRT DOSE VERIFICATION; SPATIAL-RESOLUTION; BEAMS; RADIOTHERAPY; PHOTON; DOSIMETRY; DETECTOR; SIZE; FILM;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Methods: Five patients with 38 fields have been analyzed in this study. The plans were optimized for the following clinical sites: one. liver, one lung, one brain, one prostate aiid one spine. The detector array used for the measurements was the PTW Seven29 array. All the plans were optimized and calculated using Eclipse v8.9. The center of the array was setup at 215 cm from the source and all the fields were measured and analyzed one by one. All the 30 measurements were performed on a NovalisTX linear accelerator equipped with a high definition multileaf collimator. The evaluation was based mainly on gamma index passing rates using 2 mm distance to agreement (DTA) and 2% dose difference. Results: The accuracy of the Eclipse Treatment Planning System (TPS) at extended Source to Surface Distances (SSDs) using an ionization chamber was measured to be within 1.0%. All the field measurements were performed and analyzed 35. individually. The percent of the points that had a gamma index of less than 1 using 3%/3 mm was >99% for all the measurements. In order to better evaluate our process and distinguish smaller differences a new set of results was obtained by applying gamma index tolerances of 2%/2mm. In this case, the gamma index passing rates ranged from 90.8 to 100% (95.5%+/- 3%). The profile comparison showed that the detector array measurements followed closely the calculated 40 profiles, even for fields optimized with multiple peaks and valleys. Conclusion: The choice of the IMRT QA device has an important role in the results of the patient specific QA of the delivered dose to the patient in the case of small targets as in the treatment of spinal targets. In this study, we demonstrated that an extended SSD measurement can improve the sampling resolution of a two-dimensional (2D) detector array, in our case the PTW 45 Seven29 array. This method was shown to be accurate and efficient for measuring highly modulated small fields for pre-treatment patient specific QA.
引用
收藏
页码:1154 / 1163
页数:10
相关论文
共 50 条
[41]   Improvement Using Planomics Features on Prediction and Classification of Patient-Specific Quality Assurance Using Head and Neck Volumetric Modulated Arc Therapy Plan [J].
Li, Bing ;
Chen, Junying ;
Guo, Wei ;
Mao, Ronghu ;
Zheng, Xiaoli ;
Cheng, Xiuyan ;
Cui, Tiantian ;
Lou, Zhaoyang ;
Wang, Ting ;
Li, Dingjie ;
Tao, Hongyan ;
Lei, Hongchang ;
Ge, Hong .
FRONTIERS IN NEUROSCIENCE, 2021, 15
[42]   Applications of artificial intelligence for machine- and patient-specific quality assurance in radiation therapy: current status and future directions [J].
Ono, Tomohiro ;
Iramina, Hiraku ;
Hirashima, Hideaki ;
Adachi, Takanori ;
Nakamura, Mitsuhiro ;
Mizowaki, Takashi .
JOURNAL OF RADIATION RESEARCH, 2024, 65 (04) :421-432
[43]   Clinical impact of using the deterministic patient dose calculation algorithm Acuros XB for lung stereotactic body radiation therapy [J].
Liu, Hong-Wei ;
Nugent, Zoann ;
Clayton, Ravinder ;
Dunscombe, Peter ;
Lau, Harold ;
Khan, Rao .
ACTA ONCOLOGICA, 2014, 53 (03) :324-329
[44]   Development of a three-dimensional scintillation detector for pencil beam verification in proton therapy patient-specific quality assurance [J].
Frelin, Anne-Marie ;
Daviau, Gautier ;
Bui, My Hoang Hoa ;
Fontbonne, Cathy ;
Fontbonne, Jean-Marc ;
Lebhertz, Dorothee ;
Mainguy, Erwan ;
Moignier, Cyril ;
Thariat, Juliette ;
Vela, Anthony .
MEDICAL PHYSICS, 2024, 51 (12) :9318-9329
[45]   Feasibility of using ring-mounted Halcyon Linac for single-isocenter/two-lesion lung stereotactic body radiation therapy [J].
Pokhrel, Damodar ;
Webster, Aaron ;
Mallory, Richard ;
Visak, Justin ;
Bernard, Mark E. ;
McGarry, Ronald C. ;
Kudrimoti, Mahesh .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05)
[46]   Evaluation of plan quality and treatment efficiency for single-isocenter/two-lesion lung stereotactic body radiation therapy [J].
Sanford, Lana ;
Molloy, Janette ;
Kumar, Sameera ;
Randall, Marcus ;
McGarry, Ronald ;
Pokhrel, Damodar .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01) :118-127
[47]   Dose-Escalated Stereotactic Body radiation Therapy for Prostate Cancer: Quality-of-life Comparison of Two Prospective Trials [J].
Quon, Harvey C. ;
Musunuru, Hima Bindu ;
Cheung, Patrick ;
Pang, Geordi ;
Mamedov, Alexandre ;
D'Alimonte, Laura ;
Deabreu, Andrea ;
Zhang, Liying ;
Loblaw, Andrew .
FRONTIERS IN ONCOLOGY, 2016, 6
[48]   Patient-Specific Quality Assurance for the Delivery of 60Co Intensity Modulated Radiation Therapy Subject to a 0.35-T Lateral Magnetic Field [J].
Li, H. Harold ;
Rodriguez, Vivian L. ;
Green, Olga L. ;
Hu, Yanle ;
Kashani, Rojano ;
Wooten, H. Omar ;
Yang, Deshan ;
Mutic, Sasa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 91 (01) :65-72
[49]   Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation [J].
Basavatia, Amar K. ;
Fiedler, Derek A. ;
Ritter, Johannes ;
Brodin, Patrik ;
Goddard, Lee C. ;
Jeong, Kyoungkeun ;
Hsu, Shu-Hui ;
Mynampati, Dinesh K. ;
Yaparpalvi, Ravindra ;
Tome, Wolfgang A. .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (05)
[50]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Kadoya, Noriyuki ;
Abe, Kota ;
Nemoto, Hikaru ;
Sato, Kiyokazu ;
Ieko, Yoshiro ;
Ito, Kengo ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) :351-356