Aging of imaging properties of a CMOS flat-panel detector for dental cone-beam computed tomography

被引:4
|
作者
Kim, D. W. [1 ]
Han, J. C. [1 ,3 ]
Yun, S. [1 ,3 ]
Kim, H. K. [1 ,2 ]
机构
[1] Pusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea
[2] Pusan Natl Univ, Ctr Adv Med Engn Res, Busan 46241, South Korea
[3] Samsung Elect Co, Suwon 16677, Gyeonggi Do, South Korea
来源
关键词
Computerized Tomography (CT) and Computed Radiography (CR); X-ray radiography and digital radiography (DR); X-RAY-DETECTOR; PHOTODIODE-ARRAY DETECTORS; ACTIVE PIXEL SENSORS; SCINTILLATING CRYSTALS; RADIATION-DAMAGE; GAIN; CT; TECHNOLOGIES; MAMMOGRAPHY; PERFORMANCE;
D O I
10.1088/1748-0221/12/01/P01005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have experimentally investigated the long-term stability of imaging properties of a flat-panel detector in conditions used for dental x-ray imaging. The detector consists of a CsI: Tl layer and CMOS photodiode pixel arrays. Aging simulations were carried out using an 80-kVp x-ray beam at an air-kerma rate of approximately 5 mGy s(-1) at the entrance surface of the detector with a total air kerma of up to 0.6 kGy. Dark and flood-field images were periodically obtained during irradiation, and the mean signal and noise levels were evaluated for each image. We also evaluated the modulation-transfer function (MTF), noise-power spectrum (NPS), and detective quantum efficiency (DQE). The aging simulation showed a decrease in both the signal and noise of the gain-offset-corrected images, but there was negligible change in the signal-to-noise performance as a function of the accumulated dose. The gain-offset correction for analyzing images resulted in negligible changes in MTF, NPS, and DQE results over the total dose. Continuous x-ray exposure to a detector can cause degradation in the physical performance factors such the detector sensitivity, but linear analysis of the gain-offset-corrected images can assure integrity of the imaging properties of a detector during its lifetime.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Super Resolution Dual-Energy Cone-Beam CT Imaging With Dual-Layer Flat-Panel Detector
    Su, Ting
    Zhu, Jiongtao
    Zhang, Xin
    Tan, Yuhang
    Cui, Han
    Zeng, Dong
    Guo, Jinchuan
    Zheng, Hairong
    Ma, Jianhua
    Liang, Dong
    Ge, Yongshuai
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2024, 43 (02) : 734 - 744
  • [32] Coronary artery imaging with flat-panel computed tomography
    Knollmann, F
    Pfoh, A
    CIRCULATION, 2003, 107 (08) : 1209 - 1209
  • [33] Dose and image quality of flat-panel detector volume computed tomography for sinus imaging
    Alspaugh, J.
    Christodoulou, E.
    Goodsitt, M.
    Stayman, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2634 - 2634
  • [34] Flat panel cone beam computed tomography of the sinuses
    Zoumalan, Richard A.
    Lebowitz, Richard A.
    Wang, Edwin
    Yung, Kathryn
    Babb, James S.
    Jacobs, Joseph B.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2009, 140 (06) : 841 - 844
  • [35] Lag correction for flat-panel cone-beam CT in SPECT/CT
    Wang, Jiong
    Hansis, Eberhard
    van de Haar, Peter
    Sowards-Emmerd, David
    Bredno, Joerg
    Shao, Lingxiong
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [36] Flat-panel cone-beam CT: A novel imaging technology for image-guided procedures
    Siewerdsen, JH
    Jaffray, DA
    Edmundson, GK
    Sanders, WP
    Wong, J
    Martinez, AA
    MEDICAL IMAGING 2001: VISUALIZATION, DISPLAY, AND IMAGE-GUIDED PROCEDURES, 2001, 4319 : 435 - 444
  • [37] Comparison of cone-beam computed tomography with photon-counting detector computed tomography for dental implant surgery
    Husain, Adib Al-Haj
    Mergen, Victor
    Valdec, Silvio
    Husain, Nadin Al-Haj
    Stadlinger, Bernd
    Essig, Harald
    Frauenfelder, Thomas
    Kessler, Peter
    Lie, Suen An Nynke
    Alkadhi, Hatem
    Winklhofer, Sebastian
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2025, 11 (01)
  • [38] Simulation of Motion Artifacts in Offset Flat-Panel Cone-Beam CT
    Hansis, Eberhard
    Shao, Lingxiong
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2471 - 2475
  • [39] NEQ description of 3-D imaging performance in flat-panel cone-beam CT
    Siewerdsen, J
    Jaffray, D
    MEDICAL PHYSICS, 2002, 29 (06) : 1321 - 1321
  • [40] Iterative cone-beam CT reconstruction with an off-center flat-panel detector for SPECT/CT
    Hansis, Eberhard
    Sowards-Emmerd, David
    Hines, Horace
    Shao, Lingxiong
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51