Aging of imaging properties of a CMOS flat-panel detector for dental cone-beam computed tomography

被引:4
作者
Kim, D. W. [1 ]
Han, J. C. [1 ,3 ]
Yun, S. [1 ,3 ]
Kim, H. K. [1 ,2 ]
机构
[1] Pusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea
[2] Pusan Natl Univ, Ctr Adv Med Engn Res, Busan 46241, South Korea
[3] Samsung Elect Co, Suwon 16677, Gyeonggi Do, South Korea
来源
JOURNAL OF INSTRUMENTATION | 2017年 / 12卷
关键词
Computerized Tomography (CT) and Computed Radiography (CR); X-ray radiography and digital radiography (DR); X-RAY-DETECTOR; PHOTODIODE-ARRAY DETECTORS; ACTIVE PIXEL SENSORS; SCINTILLATING CRYSTALS; RADIATION-DAMAGE; GAIN; CT; TECHNOLOGIES; MAMMOGRAPHY; PERFORMANCE;
D O I
10.1088/1748-0221/12/01/P01005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have experimentally investigated the long-term stability of imaging properties of a flat-panel detector in conditions used for dental x-ray imaging. The detector consists of a CsI: Tl layer and CMOS photodiode pixel arrays. Aging simulations were carried out using an 80-kVp x-ray beam at an air-kerma rate of approximately 5 mGy s(-1) at the entrance surface of the detector with a total air kerma of up to 0.6 kGy. Dark and flood-field images were periodically obtained during irradiation, and the mean signal and noise levels were evaluated for each image. We also evaluated the modulation-transfer function (MTF), noise-power spectrum (NPS), and detective quantum efficiency (DQE). The aging simulation showed a decrease in both the signal and noise of the gain-offset-corrected images, but there was negligible change in the signal-to-noise performance as a function of the accumulated dose. The gain-offset correction for analyzing images resulted in negligible changes in MTF, NPS, and DQE results over the total dose. Continuous x-ray exposure to a detector can cause degradation in the physical performance factors such the detector sensitivity, but linear analysis of the gain-offset-corrected images can assure integrity of the imaging properties of a detector during its lifetime.
引用
收藏
页数:14
相关论文
共 39 条
[31]   Flat detectors and their clinical applications [J].
Spahn, M .
EUROPEAN RADIOLOGY, 2005, 15 (09) :1934-1947
[32]   Inorganic scintillators in medical imaging [J].
van Eijk, CWE .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :R85-R106
[33]   Three-dimensional C-arm cone-beam CT: Applications in the interventional suite [J].
Wallace, Michael J. ;
Kuo, Michael D. ;
Glaiberman, Craig ;
Binkert, Christoph A. ;
Orth, Robert C. ;
Soulez, Gilles .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (06) :799-813
[34]   Noise power spectra of images from digital mammography detectors [J].
Williams, MB ;
Mangiafico, PA ;
Simoni, PU .
MEDICAL PHYSICS, 1999, 26 (07) :1279-1293
[35]   Signal and Noise Characteristics Induced by Unattenuated X Rays from a Scintillator in Indirect-Conversion CMOS Photodiode Array Detectors [J].
Yun, Seungman ;
Kim, Ho Kyung ;
Lim, Chang Hwy ;
Cho, Min Kook ;
Achterkirchen, Thorsten ;
Cunningham, Ian .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :1121-1128
[36]   50 μm pixel pitch wafer-scale CMOS active pixel sensor x-ray detector for digital breast tomosynthesis [J].
Zhao, C. ;
Konstantinidis, A. C. ;
Zheng, Y. ;
Anaxagoras, T. ;
Speller, R. D. ;
Kanicki, J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (23) :8977-9001
[37]   Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization [J].
Zhao, Chumin ;
Kanicki, Jerzy ;
Konstantinidis, Anastasios C. ;
Patel, Tushita .
MEDICAL PHYSICS, 2015, 42 (11) :6294-6308
[38]   Scintillating crystals in a radiation environment [J].
Zhu, RY ;
Ma, DA ;
Newman, H .
NUCLEAR PHYSICS B, 1995, :547-556
[39]   Radiation damage in scintillating crystals [J].
Zhu, RY .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 413 (2-3) :297-311