Electromagnetic interference of implantable cardiac devices from a shoulder massage machine

被引:7
作者
Yoshida, Saeko [1 ]
Fujiwara, Kousaku [2 ]
Kohira, Satoshi [3 ]
Hirose, Minoru [1 ,2 ]
机构
[1] Kitasato Univ, Grad Sch Med Sci, Dept Med Safety Engn, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[2] Kitasato Univ, Sch Allied Hlth Sci, Clin Engn Course, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[3] Kitasato Univ Hosp, Dept Med Engn, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
关键词
Pacemaker; EMI; Massage machine; PACEMAKERS;
D O I
10.1007/s10047-014-0765-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter-defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a "don't use" warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
引用
收藏
页码:243 / 249
页数:7
相关论文
共 13 条
[1]  
[Anonymous], 2005, 147082005 ISO
[2]  
Electromagnetic Compatibility Conference (EMCC), 1999, EL COMP C EMCC GLOSS, P122
[3]  
Gimbel JR, 2007, MAYO CLIN PROC, V82, P318
[4]   Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven [J].
Hirose, M ;
Hida, M ;
Sato, E ;
Kokubo, K ;
Nie, M ;
Kobayashi, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (06) :540-548
[5]   Electromagnetic interference of pacemakers by mobile phones [J].
Irnich, W ;
Batz, L ;
Muller, R ;
Tobisch, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1431-1446
[6]   INTERFERENCE IN PACEMAKERS [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (06) :1021-1048
[7]  
Ministry of Internal Affairs and Communications (MIC), 2002, SURV STUD REP INFL M
[8]  
Ministry of Internal Affairs and Communications (MIC), 2011, SURV STUD REP INFL M
[9]  
Toyoshima T, 1996, CARD PACING, V12, P488
[10]  
Toyoshima T., 2007, ELECTROMAGNETIC INTE, P56