Retrieval analysis of PEEK rods for posterior fusion and motion preservation

被引:34
作者
Kurtz, Steven M. [1 ,2 ]
Lanman, Todd H. [3 ]
Higgs, Genymphas [1 ]
MacDonald, Daniel W. [1 ]
Berven, Sigurd H. [4 ]
Isaza, Jorge E. [5 ]
Phillips, Eual [1 ]
Steinbeck, Marla J. [1 ]
机构
[1] Drexel Univ, Philadelphia, CA 19104 USA
[2] Exponent Inc, Menlo Pk, CA 94025 USA
[3] Univ Calif Los Angeles, Dept Surg, Los Angeles, CA 90024 USA
[4] Univ Calif San Francisco, Dept Orthopaed Surg, San Francisco, CA USA
[5] Tulane Univ, Baton Rouge, LA USA
关键词
PEEK rods; Posterior fusion; Retrieval analysis; Tissue response; Revision; SPINAL IMPLANTS; ADJACENT;
D O I
10.1007/s00586-013-2920-4
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this study was to analyze explanted PEEK rod spinal systems in the context of their clinical indications. We evaluated damage to the implant and histological changes in explanted periprosthetic tissues. 12 patients implanted with 23 PEEK rods were revised between 2008 and 2012. PEEK rods were of the same design (CD Horizon Legacy, Medtronic, Memphis TN, USA). Retrieved components were assessed for surface damage mechanisms, including plastic deformation, scratching, burnishing, and fracture. Patient history and indications for PEEK rod implantation were obtained from analysis of the medical records. 11/12 PEEK rod systems were employed for fusion at one level, and motion preservation at the adjacent level. Surgical complications in the PEEK cohort included a small dural tear in one case that was immediately repaired. There were no cases of PEEK rod fracture or pedicle screw fracture. Retrieved PEEK rods exhibited scratching, as well as impressions from the set screws and pedicle screw saddles. PEEK debris was observed in two patient tissues, which were located adjacent to PEEK rods with evidence of scratching and burnishing. This study documents the surface changes and tissue reactions for retrieved PEEK rod stabilization systems. Permanent indentations by the set screws and pedicle screws were the most prevalent observations on the surface of explanted PEEK rods.
引用
收藏
页码:2752 / 2759
页数:8
相关论文
共 11 条
  • [1] Hybrid construct for two levels disc disease in lumbar spine
    Aunoble, Stephane
    Meyrat, Robert
    Al Sawad, Yasser
    Tournier, C.
    Leijssen, Philip
    Le Huec, Jean-Charles
    [J]. EUROPEAN SPINE JOURNAL, 2010, 19 (02) : 290 - 296
  • [2] Posterior lumbar fusion by peek rods in degenerative spine: preliminary report on 30 cases
    De Iure, F.
    Bosco, G.
    Cappuccio, M.
    Paderni, S.
    Amendola, L.
    [J]. EUROPEAN SPINE JOURNAL, 2012, 21 : S50 - S54
  • [3] Biomechanical Assessment of a PEEK Rod System for Semi-Rigid Fixation of Lumbar Fusion Constructs
    Gornet, Matthew F.
    Chan, Frank W.
    Coleman, John C.
    Murrell, Brian
    Nockels, Russ P.
    Taylor, Brett A.
    Lanman, Todd H.
    Ochoa, Jorge A.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2011, 133 (08):
  • [4] Highsmith Jason M, 2007, Neurosurg Focus, V22, pE11
  • [5] In Vivo Deformation, Surface Damage, and Biostability of Retrieved Dynesys Systems
    Ianuzzi, Allyson
    Kurtz, Steven M.
    Kane, William
    Shah, Priyanka
    Siskey, Ryan
    van Ooij, Andre
    Bindal, Rajesh
    Ross, Raymond
    Lanman, Todd
    Buettner-Janz, Karin
    Isaza, Jorge
    [J]. SPINE, 2010, 35 (23) : E1310 - E1316
  • [6] Kurtz SM, 2012, PDL HANDB SER, P201, DOI 10.1016/B978-1-4377-4463-7.10013-2
  • [7] Kurtz Steven M, 2009, SAS J, V3, P161, DOI 10.1016/j.esas.2009.11.003
  • [8] Hybrid dynamic stabilization: a biomechanical assessment of adjacent and supraadjacent levels of the lumbar spine Laboratory investigation
    Mageswaran, Prasath
    Techy, Fernando
    Colbrunn, Robb W.
    Bonner, Tara F.
    McLain, Robert F.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2012, 17 (03) : 232 - 242
  • [9] Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation
    Ponnappan, Ravi K.
    Serhan, Hassan
    Zarda, Brett
    Patel, Ravi
    Albert, Todd
    Vaccaro, Alexander R.
    [J]. SPINE JOURNAL, 2009, 9 (03) : 263 - 267
  • [10] Comparison of standard fusion with a "topping off" system in lumbar spine surgery: a protocol for a randomized controlled trial
    Siewe, Jan
    Otto, Christina
    Knoell, Peter
    Koriller, Marco
    Stein, Gregor
    Kaulhausen, Thomas
    Eysel, Peer
    Zarghooni, Kourosh
    Franklin, Jeremy
    Sobottke, Rolf
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2011, 12