Tibiofemoral contact mechanics with a femoral resurfacing prosthesis and a non-functional meniscus

被引:18
作者
Becher, Christoph [1 ]
Huber, Roland [2 ]
Thermann, Hajo [3 ]
Tibesku, Carsten Oliver [4 ]
von Skrbensky, Gobert [2 ]
机构
[1] Hannover Med Sch, Dept Orthopaed, D-30625 Hannover, Germany
[2] Med Univ Vienna, Bone & Joint Biomech Lab, A-1090 Vienna, Austria
[3] ATOS Clin Ctr, Ctr Knee & Foot Surg Sports Trauma, D-69115 Heidelberg, Germany
[4] Sporthopaedicum Straubing, D-94315 Straubing, Germany
关键词
Osteochondral defect; Tibiofemoral joint; Meniscus; Articular prosthetic device; Peak contact pressure; ARTICULAR-CARTILAGE DEFECTS; LONG-TERM PROGNOSIS; FOLLOW-UP; MEDIAL MENISCUS; CHONDROCYTE TRANSPLANTATION; PARTIAL MENISCECTOMY; KNEE ARTHROSCOPIES; POSTERIOR HORN; PRESSURE; STRESS;
D O I
10.1016/j.clinbiomech.2009.05.013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Increased contact stress with a femoral resurfacing prosthesis implanted in the medial femoral condyle and a non-functional meniscus is of concern for potential deleterious effects on tibiofemoral contact mechanics. Methods: Peak contact pressures were determined in seven fresh frozen human cadaveric specimens using a pressure sensitive sensor placed in the medial compartment above the menisci. A knee simulator was used to test each knee in static stance positions (5 degrees/15 degrees/30 degrees/45 degrees) and through 10 dynamic knee flexion cycles (5-45 degrees) with single body weight ground reaction force which was adjusted to the living body weight of the cadaver donor. All specimens were tested in three different conditions: untreated knee (A); flush implantation of a 20 mm resurfacing prosthesis (HemiCAP (R)) in the weight bearing area of the medial femoral condyle (B): complete radial tear at the posterior horn of the medial meniscus with the femoral resurfacing device in place (C). Findings: On average, flush device implantation resulted in no statistically significant differences when compared to the untreated normal knee. The meniscal tear resulted in a significant increase of the mean maximum peak contact pressures by 63%, 57%, and 57% (all P <= 0.05) at 15 degrees, 30 degrees and 45 degrees static stance positions and 78% (P <= 0.05) through the dynamic knee flexion cycle. No significant different maximum peak contact pressures were observed at 5 degrees stance position. Interpretation: Although the condition of a meniscal tear without the resurfacing device could not be compared, possible effects of reduced meniscal tissue and biomechanical integrity of the meniscus must be considered in an in vivo application. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:648 / 654
页数:7
相关论文
共 38 条
[11]  
HIGUCHI H, 2000, CLIN ORTHOP RELAT R, V16, P1
[12]   Articular cartilage defects in 1,000 knee arthroscopies [J].
Hjelle, K ;
Solheim, E ;
Strand, T ;
Muri, R ;
Brittberg, M .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2002, 18 (07) :730-734
[13]   Direct measurement of hoop strains in the intact and torn human medial meniscus [J].
Jones, RS ;
Keene, GCR ;
Learmonth, DJA ;
Bickerstaff, D ;
Nawana, NS ;
Costi, JJ ;
Pearcy, MJ .
CLINICAL BIOMECHANICS, 1996, 11 (05) :295-300
[14]   Safety of, and biological and functional response to, a novel metallic implant for the management of focal full-thickness cartilage defects: Preliminary assessment in an animal model out to 1 year [J].
Kirker-Head, CA ;
Van Sickle, DC ;
Ek, SW ;
McCool, JC .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2006, 24 (05) :1095-1108
[15]   Cadaveric analysis of contact stress restoration after osteochondral transplantation of a cylindrical cartilage defect [J].
Kock, Niels B. ;
Smolders, Jose M. H. ;
van Susante, Job L. C. ;
Buma, Pieter ;
van Kampen, Albert ;
Verdonschot, Nico .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2008, 16 (05) :461-468
[16]   The effect of angled osteochondral grafting on contact pressure - A biomechanical study [J].
Koh, JL ;
Kowalski, A ;
Lautenschlager, E .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (01) :116-119
[17]   MECHANICAL CHANGES IN KNEE AFTER MENISCECTOMY [J].
KRAUSE, WR ;
POPE, MH ;
JOHNSON, RJ ;
WILDER, DG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (05) :599-604
[18]   Microfracture of chondral defects in the knee associated with different results in patients aged 40 years or younger? [J].
Kreuz, Peter C. ;
Erggelet, Christoph ;
Steinwachs, Matthias R. ;
Krause, Stefanie J. ;
Lahm, Andreas ;
Niemeyer, Philipp ;
Ghanem, Nadir ;
Uhl, Markus ;
Suedkamp, Norbert .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (11) :1180-1186
[19]  
KUROSAWA H, 1980, CLIN ORTHOP RELAT R, P283
[20]   Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee [J].
Lee, Stephen J. ;
Aadalen, Kirk J. ;
Malaviya, Prasanna ;
Lorenz, Eric P. ;
Hayden, Jennifer K. ;
Farr, Jack ;
Kang, Richard W. ;
Cole, Brian J. .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (08) :1334-1344