Continuous Monitoring of Intraocular Pressure Rationale and Progress Toward A Clinical Device

被引:59
作者
Sit, Arthur J. [1 ]
机构
[1] Mayo Clin, Coll Med, Dept Ophthalmol, Rochester, MN 55905 USA
关键词
intraocular pressure; telemetry; tonometry; circadian rhythms; OPEN-ANGLE GLAUCOMA; INTRA-OCULAR PRESSURE; APPLANATING SUCTION CUP; SELF-TONOMETRY; BLOOD-PRESSURE; CIRCADIAN-RHYTHM; NONCONTACT TONOMETER; GOLDMANN TONOMETRY; CONTACT-LENS; MONKEY EYES;
D O I
10.1097/IJG.0b013e3181862490
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Intraocular pressure (IOP) is a dynamic physiologic parameter with regular circadian variations and unpredictable short-term and long-term fluctuations. Current methods of measuring IOP are suboptimal with a typical clinical practice only performing periodic IOP measurements during regular office hours. Diurnal and 24-hour IOP measurements obtained on an in-patient basis can increase measurements but are inconvenient and expensive, and do not allow ambulatory monitoring of IOP. The goal of continuous IOP monitoring is to provide automated 24-hour recording of ambulatory IOP. Continuous IOP monitoring involves 2 complementary paradigms. Temporary noninvasive monitoring. possibly involving a contact lens-based pressure sensor. would be used to measure 24-hour IOP on a periodic basis. Permanent monitoring would be more invasive, using an implantable pressure sensor. Despite numerous previous attempts at continuous IOP monitoring, a device suitable for clinical use is not yet available. However, devices currently in development for permanent IOP monitoring seem to be nearly ready for human testing. The technologic issues for temporary monitoring may be greater than for permanent monitoring.
引用
收藏
页码:272 / 279
页数:8
相关论文
共 76 条
[1]   Continuous monitoring of circadian variations in intraocular pressure by telemetry system throughout a 12-week treatment with timolol maleate in rabbits [J].
Akaishi, T ;
Ishida, N ;
Shimazaki, A ;
Hara, H ;
Kuwayama, Y .
JOURNAL OF OCULAR PHARMACOLOGY AND THERAPEUTICS, 2005, 21 (06) :436-444
[2]  
ANJOU CI, 1961, ACTA OPHTHALMOL, V39, P852
[3]  
Asrani S, 2000, J GLAUCOMA, V9, P134
[4]   Clinical utility of Intraocular pressure monitoring outside of normal office hours in patients with glaucoma [J].
Barkana, Yaniv ;
Anis, Sarah ;
Liebmann, Jeffrey ;
Tello, Celso ;
Ritch, Robert .
ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (06) :793-797
[5]   Fluctuation of intraocular pressure and glaucoma progression in the early manifest glaucoma trial [J].
Bengtsson, Boel ;
Leske, M. Cristina ;
Hyman, Leslie ;
Heijl, Anders .
OPHTHALMOLOGY, 2007, 114 (02) :205-209
[6]   Reproducibility of self-measured intraocular pressure with the phosphene tonometer in patients with ocular hypertension and early to advanced glaucoma [J].
Brigatti, L ;
Maguluri, S .
JOURNAL OF GLAUCOMA, 2005, 14 (01) :36-39
[7]   STRESS-STRAIN BEHAVIOR OF CORNEO-SCLERAL ENVELOPE OF EYE .2. INVIVO MEASUREMENTS IN RHESUS-MONKEY EYES [J].
BRUBAKER, RF ;
JOHNSON, SA ;
BEELER, GW .
EXPERIMENTAL EYE RESEARCH, 1977, 24 (05) :425-435
[8]   THE PULSAIR-KEELER NONCONTACT TONOMETER IN SELF-TONOMETRY - PRELIMINARY-RESULTS [J].
CARENINI, BB ;
BROGLIATTI, B ;
TONETTO, C ;
RENIS, E .
INTERNATIONAL OPHTHALMOLOGY, 1992, 16 (4-5) :295-297
[9]   Circadian fluctuation of mean ocular perfusion pressure is a consistent risk factor for normal-tension glaucoma [J].
Choi, Jaewan ;
Kim, Kyung Hoon ;
Jeong, Jinho ;
Cho, Hyun-soo ;
Lee, Chang Hwan ;
Kook, Michael S. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (01) :104-111