Controlled clinical comparison of VersaTREK and BacT/ALERT blood culture systems

被引:21
作者
Mirrett, Stanley
Hanson, Kimberly E.
Reller, L. Barth
机构
[1] Duke Univ, Med Ctr, Clin Microbiol Lab, Durham, NC 27710 USA
[2] Duke Univ, Sch Med, Dept Pathol, Durham, NC 27710 USA
[3] Duke Univ, Sch Med, Dept Med, Durham, NC 27710 USA
关键词
D O I
10.1128/JCM.01697-06
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
To assess the relative yields in automated microbial detection systems of bacteria and yeasts isolated from the blood of adult patients with suspected sepsis, we compared the new VersaTREK system (VTI) (TREK Diagnostic Systems, Cleveland, 011) to the BacT/ALERT 3D system (3D) (bioMerieux, Inc., Durham, NC). Identical protocols were followed for the two systems. Paired aerobic (REDOX 1) and anaerobic (REDOX 2) VTI media were compared with standard aerobic (SA) and anaerobic (SN) 3D media; each of the four culture bottles was filled with 6 to 9 ml of blood. All bottles flagged positive by the instruments were subcultured to determine both true-positive (growth) and false-positive (no growth) cultures. Additionally, to assess false-negative bottles, terminal subcultures were done on all negative companion bottles to true-positive bottles. All isolates were identified by standard methods. All 4 bottles were adequately filled and yielded 413 clinically significant isolates in 5,389 (79%) of the 6,786 4-bottle sets obtained. Although no overall difference in yield or in time to detection was detected between the two systems, significantly more streptococci and enterococci as a group were detected by VTI. Moreover, significantly more microorganisms were detected by VTI for patients receiving antimicrobial therapy. The two systems were comparable (P, not significant) at detecting the 179 unimicrobial episodes of bacteremia seen. False-positive rates for aerobic and anaerobic bottles, respectively, were 1.6% and 0.9% for VTI and 0.7% and 0.8% for 3D. We conclude that the VTI and 3D systems are comparable for detection of bloodstream infections with bacteria or yeasts.
引用
收藏
页码:299 / 302
页数:4
相关论文
共 7 条
[1]   Controlled comparative evaluation of BacT/Alert FAN and ESP 80A aerobic media as means for detecting bacteremia and fungemia [J].
Doern, GV ;
Barton, A ;
Rao, S .
JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (09) :2686-2689
[2]   CONTROLLED CLINICAL-EVALUATION OF ISOLATOR AND ESP AEROBIC BLOOD CULTURE SYSTEMS FOR DETECTION OF BLOOD-STREAM INFECTIONS [J].
KIRKLEY, BA ;
EASLEY, KA ;
WASHINGTON, JA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (06) :1547-1549
[3]  
McNemar Q., 1969, Psychological statistics, V4th ed
[4]  
Murray P.R., 1999, MANUAL CLIN MICROBIO, V7th
[5]  
Weinstein MP, 1997, CLIN INFECT DIS, V24, P584, DOI 10.1093/clind/24.4.584
[6]   Current blood culture methods and systems: Clinical concepts, technology, and interpretation of results [J].
Weinstein, MP .
CLINICAL INFECTIOUS DISEASES, 1996, 23 (01) :40-46
[7]   COMPARISON OF DIFCO ESP AND ORGANON-TEKNIKA BACT/ALERT CONTINUOUS-MONITORING BLOOD CULTURE SYSTEMS [J].
ZWADYK, P ;
PIERSON, CL ;