Displacement of the double-lumen endobronchial tube can be detected by bronchial cuff pressure change

被引:16
作者
Araki, K [1 ]
Nomura, R [1 ]
Urushibara, R [1 ]
Yoshikawa, Y [1 ]
Hatano, Y [1 ]
机构
[1] WAKAYAMA MED COLL, DEPT ANESTHESIOL, WAKAYAMA 640, JAPAN
关键词
D O I
10.1097/00000539-199706000-00032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We measured the bronchial cuff pressure of left-sided double-lumen endobronchial tubes (DLTs) in 54 patients to confirm the effect of DLT displacement on cuff pressure. After positioning the cephalad surface of the bronchial cuff of the DLT 2.5 cm distal to the carina (23 patients in the first part of the study) or just below the carina (23 patients in the second part), the cuff was withdrawn in 0.5-cm steps during right-sided, one-lung ventilation. The bronchial cuff pressure was measured, and the capnogram and pressure-volume loop, displayed by a side-stream spirometer, was evaluated. After positioning the cephalad surface of the bronchial cuff just below the carina (eight patients in the third part), bronchial cuff pressure was measured while confirming the position of the bronchial cuff during the 30 min of the operative procedure. The bronchial cuff pressure decreased significantly by 28.4 cm H2O (P < 0.01) and 21.3 cm H2O (P < 0.01) in the first and second parts, respectively, before the pressure-volume loop or the capnogram changed. The bronchial cuff pressure in the third part showed no significant change. We conclude that bronchial cuff pressure monitoring was very helpful in detecting displacement of the DLT during right-sided, one-lung ventilation.
引用
收藏
页码:1349 / 1353
页数:5
相关论文
共 22 条
  • [1] Bardoczky G I, 1992, J Cardiothorac Vasc Anesth, V6, P731, DOI 10.1016/1053-0770(92)90062-C
  • [2] CONTINUOUS SPIROMETRY FOR DETECTION OF DOUBLE-LUMEN ENDOBRONCHIAL TUBE DISPLACEMENT
    BARDOCZKY, GI
    LEVARLET, M
    ENGELMAN, E
    DEFRANCQUEN, P
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (05) : 499 - 502
  • [3] MARGIN OF SAFETY IN POSITIONING MODERN DOUBLE-LUMEN ENDOTRACHEAL-TUBES
    BENUMOF, JL
    PARTRIDGE, BL
    SALVATIERRA, C
    KEATING, J
    [J]. ANESTHESIOLOGY, 1987, 67 (05) : 729 - 738
  • [4] MALPOSITION OF LEFT-SIDED DOUBLE-LUMEN ENDOBRONCHIAL TUBES
    BRODSKY, JB
    SHULMAN, MS
    MARK, JBD
    [J]. ANESTHESIOLOGY, 1985, 62 (05) : 667 - 669
  • [5] PRESSURE MEASUREMENTS IN ENDOTRACHEAL CUFFS - COMMON ERROR
    COX, PM
    SCHATZ, ME
    [J]. CHEST, 1974, 65 (01) : 84 - 87
  • [6] DEVRIES JW, 1993, J CLIN MONITOR, V9, P207
  • [7] DEVRIES JW, 1992, J CLIN MONITOR, V8, P236
  • [8] CHANGE IN TRACHEAL BLOOD-FLOW DURING ENDOTRACHEAL INTUBATION
    JOH, S
    MATSUURA, H
    KOTANI, Y
    SUGIYAMA, K
    HIROTA, Y
    KIYOMITSU, Y
    KUBOTA, Y
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1987, 31 (04) : 300 - 304
  • [9] PREVENTION OF COMPLICATIONS FROM PROLONGED TRACHEAL INTUBATION
    LEWIS, FR
    SCHLOBOHM, RM
    THOMAS, AN
    [J]. AMERICAN JOURNAL OF SURGERY, 1978, 135 (03) : 452 - 457
  • [10] EFFECTS OF NITROUS-OXIDE AND OXYGEN ON TRACHEAL TUBE CUFF GAS VOLUMES
    MEHTA, S