Retrospective analysis of mode survival, reliability of atrial sensing and incidence of atrial tachyarrhythmias in 307 single-lead VOID pacemaker patients

被引:8
作者
Zupan, Igor
Lipar, Luka
Zizek, David
Boute, Wim
Vidmar, Masa
Gabrijelcic, Tone
Rakovec, Peter
Brecelj, Ales
机构
[1] Univ Ljubljana, Ctr Med, Dept Cardiol, Ljubljana 1000, Slovenia
[2] Vitatron BV, NL-6825 MJ Arnhem, Netherlands
[3] Educ Res Inst, Ljubljana 1000, Slovenia
[4] Univ Ljubljana, Med Ctr, Dept Cardiovasc Surg, Ljubljana 61000, Slovenia
来源
EUROPACE | 2006年 / 8卷 / 10期
关键词
single-lead VDD pacemaker; long-term follow-up; atrial tachyarrhythmia; mode survival; retrospective analysis;
D O I
10.1093/europace/eul091
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The aim of this retrospective analysis was to investigate VDD mode survival, development of atrial. tachyarrhythmias' (AT), and long-term atrial sensing performance of VDD pacing systems. Methods and results We implanted single-lead VDD pacemakers in patients with isolated atrioventricular block and performed a retrospective analysis of 307 patients who had their devices implanted between May 1994 and September 2001. In 39 patients (12.7%), the pacing mode had to be reprogrammed to a single-chamber ventricular pacing mode, mostly due to permanent AT. In 16 of these patients, the atrial. sensing safety margin was less than 150%. The atrial. sensing safety margin was insufficient, i.e. less than 100% in only seven patients. Although only 12 (3.9%) of the patients had a history of paroxysmal AT at the time of pacemaker implantation, 200 (65%) patients presented with AT during follow-up. The mean AT burden at the last follow-up was 2.5%. Conclusion These data illustrate that single-lead VDD pacemakers can be applied without serious complications in a highly selected group of patients. Our main concern is the development of AT in a large part of our population. Over a 10-year period, two thirds of our patients presented with AT.
引用
收藏
页码:855 / 858
页数:4
相关论文
共 27 条
[1]  
CARLETON RA, 1966, CLIN SCI, V30, P151
[2]  
CHAMBERLAIN DA, 1973, BRIT HEART J, V35, P559
[3]   Long-term survival of VDD pacing [J].
Chamberlain-Webber, R ;
Barnes, E ;
Papouchado, M ;
Crick, JP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11) :2246-2248
[4]   CLINICAL-EVALUATION OF VDD PACING WITH A UNIPOLAR SINGLE-PASS LEAD [J].
CORNACCHIA, D ;
FABBRI, M ;
MARESTA, A ;
GRASSI, G ;
VAIANI, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :604-618
[5]   ATRIAL SYNCHRONIZED VENTRICULAR PACING - CONTRIBUTION OF THE CHRONOTROPIC RESPONSE TO IMPROVED EXERCISE PERFORMANCE [J].
FANANAPAZIR, L ;
BENNETT, DH ;
MONKS, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (03) :601-608
[6]   Incidence of atrial fibrillation in patients with different mode of pacing. Long-term follow-up [J].
Folino, AF ;
Buja, G ;
Dal Corso, L ;
Nava, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01) :260-263
[7]  
HOFFMANN E, 2000, PACING CLIN ELECT II, V23, pA415
[8]   Optimal pacing for symptomatic AV block: A comparison of VDD and DDD pacing [J].
Huang, M ;
Krahn, AD ;
Yee, R ;
Klein, GJ ;
Skanes, AC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (01) :19-23
[9]   Pacing to prevent atrial fibrillation [J].
Israel, CW ;
Hohnloser, SH .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (09) :S20-S26
[10]   Development of atrial fibrillation in patients with atrioventricular block after atrioventricular synchronized pacing [J].
Kong, CW ;
Tuan, TC ;
Yin, WH ;
Yu, WC ;
Chen, SA ;
Lin, YJ ;
Huang, CY ;
Chung, SL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (03) :352-357