A simple method to prolong the service life of radioactive sources for external radiotherapy

被引:2
作者
Xu, Yingjie [1 ]
Tian, Yuan [1 ]
Dai, Jianrong [1 ]
机构
[1] Chinese Acad Med Sci, Canc Inst & Hosp, Dept Radiat Oncol, Beijing 100021, Peoples R China
基金
中国国家自然科学基金;
关键词
radioactive source; service life; MODEL;
D O I
10.1120/jacmp.v15i4.4789
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A radioactive source is usually replaced and disposed after being used for a certain amount of time (usually a half-life). In this study, a simple method is proposed to prolong its service life. Instead of replacing the used source with a new source of full activity, a new source of less activity is added in the source holder in front of the used one, so that the total activity of two sources is equal to the initial activity of the used source or even higher. Similarly, more sources can be added to the previous ones. Attenuation of front source(s) to the back source(s) was evaluated with exponential attenuation equation, and variation of source-focus distance (SFD) with inverse square law for Leksell 4C Gamma Knife, which served as an example of external radiotherapy units. When the number of front sources increased from 1 to 3, the relative air kerma decreased from 36.5% to 5.0%. Both the attenuation effect and SFD variation contributed to the decrease in air kerma, with the former being the major factor. If the height of the source can be decreased in some way, such as increasing the specific activity of sources, the sources can be used more efficiently. The method prolongs the service life of sources by several factors, and reduces the expense of source exchange and reclamation.
引用
收藏
页码:161 / 167
页数:7
相关论文
共 14 条
[1]   A simplified model of the source channel of the Leksell GammaKnife® tested with PENELOPE [J].
Al-Dweri, FM ;
Lallena, AM ;
Vilches, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2687-2703
[2]   Dosimetric Consistency of Co-60 Teletherapy Unit-a ten years Study [J].
Baba, Misba H. ;
Mohib-ul-Haq, M. ;
Khan, Aijaz A. .
INTERNATIONAL JOURNAL OF HEALTH SCIENCES-IJHS, 2013, 7 (01) :15-21
[3]  
Boswell JM, 2000, P S 50 YEARS EXC SCI, P99
[4]  
CUNNINGHAM JR, 1964, AMER J ROENTGENOL RA, V92, P202
[5]  
Dai J., 1999, MED PHYS RADIOTHERAP, V2
[6]   Comparative analysis of 60Co intensity-modulated radiation therapy [J].
Fox, Christopher ;
Romeijn, H. Edwin ;
Lynch, Bart ;
Men, Chunhua ;
Aleman, Dionne M. ;
Dempsey, James F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (12) :3175-3188
[7]   Teletherapy sources with imported and indigenous Co-60 activity [J].
George, Jain Reji ;
Kushwah, Raksha ;
Sastry, K. V. S. .
JOURNAL OF MEDICAL PHYSICS, 2009, 34 (03) :180-181
[8]   Investigation of an efficient source design for Cobalt-60-based tomotherapy using EGSnrc Monte Carlo simulations [J].
Joshi, Chandra P. ;
Darko, Johnson ;
Vidyasagar, P. B. ;
Schreiner, L. John .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (03) :575-592
[9]  
Podgorsak E.B., 2005, RAD ONCOLOGY PHYS, P123
[10]   Has the time come for doing away with Cobalt-60 teletherapy for cancer treatments [J].
Ravichandran, R. .