Minimizing Pedicle Screw Pullout Risks A Detailed Biomechanical Analysis of Screw Design and Placement

被引:80
作者
Bianco, Rohan-Jean [1 ,2 ,3 ]
Arnoux, Pierre-Jean [3 ]
Wagnac, Eric [1 ,3 ]
Mac-Thiong, Jean-Marc [2 ,4 ,5 ]
Aubin, Carl-Eric [1 ,2 ,4 ]
机构
[1] Polytech Montreal, Dept Mech Engn, Montreal, PQ, Canada
[2] Sainte Justine Univ Hosp Ctr, Montreal, PQ, Canada
[3] Univ Aix Marseille 1, Lab Biomcca Appl, UMRT24 IFSTTAR, Marseille, France
[4] Univ Montreal, Dept Surg, Montreal, PQ, Canada
[5] Hop Sacre Coeur, Dept Surg, Montreal, PQ, Canada
来源
CLINICAL SPINE SURGERY | 2017年 / 30卷 / 03期
基金
加拿大自然科学与工程研究理事会;
关键词
finite element analysis; pedicle screw; pullout test; spinal instrumentation; FINITE-ELEMENT MODEL; INSERTIONAL TORQUE; FIXATION; BONE; INSTRUMENTATION; CALIBRATION; SCOLIOSIS; STRENGTH; FAILURE; TESTS;
D O I
10.1097/BSD.0000000000000151
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: Detailed biomechanical analysis of the anchorage performance provided by different pedicle screw designs and placement strategies under pullout loading. Objective: To biomechanically characterize the specific effects of surgeon-specific pedicle screw design parameters on anchorage performance using a finite element model. Summary of Background Data: Pedicle screw fixation is commonly used in the treatment of spinal pathologies. However, there is little consensus on the selection of an optimal screw type, size, and insertion trajectory depending on vertebra dimension and shape. Methods: Different screw diameters and lengths, threads, and insertion trajectories were computationally tested using a design of experiment approach. A detailed finite element model of an L3 vertebra was created including elastoplastic bone properties and contact interactions with the screws. Loads and boundary conditions were applied to the screws to simulate axial pullout tests. Force-displacement responses and internal stresses were analyzed to determine the specific effects of each parameter. Results: The design of experiment analysis revealed significant effects (P<0.01) for all tested principal parameters along with the interactions between diameter and trajectory. Screw diameter had the greatest impact on anchorage performance. The best insertion trajectory to resist pullout involved placing the screw threads closer to the pedicle walls using the straightforward insertion technique, which showed the importance of the cortical layer grip. The simulated cylindrical single-lead thread screws presented better biomechanical anchorage than the conical dual-lead thread screws in axial loading conditions. Conclusions: The model made it possible to quantitatively measure the effects of both screw design characteristics and surgical choices, enabling to recommend strategies to improve single pedicle screw performance under axial loading.
引用
收藏
页码:E226 / E232
页数:7
相关论文
共 35 条
  • [21] Rationale behind the current state-of-the-art treatment of scoliosis (in the pedicle screw era)
    Lenke, Lawrence G.
    Kuklo, Timothy R.
    Ondra, Stephen
    Polly, David W., Jr.
    [J]. SPINE, 2008, 33 (10) : 1051 - 1054
  • [22] Biomechanical evaluation of a new anterior spinal implant
    Liu, CL
    Chen, HH
    Cheng, CK
    Kao, HC
    Lo, WH
    [J]. CLINICAL BIOMECHANICS, 1998, 13 : S40 - S45
  • [23] Biomechanical analysis of pedicle screw thread differential design in an osteoporotic cadaver model
    Mehta, H.
    Santos, E.
    Ledonio, C.
    Sembrano, J.
    Ellingson, A.
    Pare, P.
    Murrell, B.
    Nuckley, D. J.
    [J]. CLINICAL BIOMECHANICS, 2012, 27 (03) : 234 - 240
  • [24] Accuracy and safety of pedicle screw placement in neuromuscular scoliosis with free-hand technique
    Modi, Hitesh N.
    Suh, Seung Woo
    Fernandez, Harry
    Yang, Jae Hyuk
    Song, Hae-Ryong
    [J]. EUROPEAN SPINE JOURNAL, 2008, 17 (12) : 1686 - 1696
  • [25] Montgomery D.C., 1997, Design and Analysis of Experiments
  • [26] Placement of thoracolumbar pedicle screws using three-dimensional image guidance: experience in a large patient cohort Clinical article
    Nottmeier, Eric W.
    Seemer, Will
    Young, Phillip M.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2009, 10 (01) : 33 - 39
  • [27] Can insertional torque predict screw loosening and related failures?: An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion
    Okuyama, K
    Abe, E
    Suzuki, T
    Tamura, Y
    Chiba, M
    Sato, K
    [J]. SPINE, 2000, 25 (07) : 858 - 864
  • [28] Effect of specimen fixation method on pullout tests of pedicle screws
    Pfeiffer, M
    Gilbertson, LG
    Goel, VK
    Griss, P
    Keller, JC
    Ryken, TC
    Hoffman, HE
    [J]. SPINE, 1996, 21 (09) : 1037 - 1044
  • [29] Cortical bone trajectory for lumbar pedicle screws
    Santoni, B. G.
    Hynes, R. A.
    McGilvray, K. C.
    Rodriguez-Canessa, G.
    Lyons, A. S.
    Henson, M. A. W.
    Womack, W. J.
    Puttlitz, C. M.
    [J]. SPINE JOURNAL, 2009, 9 (05) : 366 - 373
  • [30] Application of a calibration method provides more realistic results for a finite element model of a lumbar spinal segment
    Schmidt, Hendrik
    Heuer, Frank
    Drumm, Joerg
    Klezl, Zdenek
    Claes, Lutz
    Wilke, Hans-Joachim
    [J]. CLINICAL BIOMECHANICS, 2007, 22 (04) : 377 - 384