An ex vivo evaluation of an inflatable bone tamp used to reduce fractures within vertebral bodies under load

被引:42
作者
Belkoff, SM
Jasper, LE
Stevens, SS
机构
[1] Johns Hopkins Univ, Dept Orthopaed Surg, Johns Hopkins Bayview Med Ctr, Orthopaed Instrumentat Lab, Baltimore, MD 21224 USA
[2] Kyphon Inc, Sunnyvale, CA USA
关键词
inflatable bone tamp; vertebroplasty; biomechanical evaluation; compression fractures; osteoporosis; kyphoplasty;
D O I
10.1097/00007632-200208010-00009
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Ex vivo biomechanical study using osteoporotic cadaveric vertebral bodies. Objective. To determine if fracture reduction could be achieved by the inflatable bone tamp (tamp) in vertebral bodies under simulated physiologic loads. Summary of Background Data. Previous ex vivo biomechanical studies showed that kyphoplasty restored vertebral body height with vertebral body endplates under stress-free conditions. Methods. Simulated compression fractures were experimentally created in 18 osteoporotic vertebral bodies alternatingly assigned to one of two treatment (tamp inflation) groups: low axial load (111 N) or high axial load (222 N). Each vertebral body was then placed between two platens in a special radiolucent loading fixture and subjected to the preassigned load to simulate in vivo physiologic loading. The tamps were inflated and post-reduction heights were measured fluoroscopically. The effect of applied load and condition on vertebral body height was checked for significance (P<0.05). Results. Comparing the experimental conditions (initial, postcompression, postinflation), there were no significant:vertebral body height differences between the load groups (low load Vs. high load). However, vertebral bodyheight differences between conditions within each load group were all significant. For the low-load and high-load groups, mean postinflation heights (24.4 and 24.4 mm) were significantly greater than mean,postcompression heights (21.6 and 22:5 mm) but significantly less than initial vertebral body heights (26.6 and 26.3 mm), respectively. Initial heights were fully restored in 22% (two of nine) of vertebral bodies in both groups. Conclusion. The inflatable bone tamp restored some of the height lost to compression fractures in vertebral bodies under simulated physiologic loads.
引用
收藏
页码:1640 / 1643
页数:4
相关论文
共 23 条
  • [1] Percutaneous vertebroplasty for pain relief and spinal stabilization
    Barr, JD
    Barr, MS
    Lemley, TJ
    McCann, RM
    [J]. SPINE, 2000, 25 (08) : 923 - 928
  • [2] An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of compression fracture
    Belkoff, SM
    Mathis, JM
    Fenton, DC
    Scribner, RM
    Reiley, ME
    Talmadge, K
    [J]. SPINE, 2001, 26 (02) : 151 - 156
  • [3] Belkoff SM, 2001, AM J NEURORADIOL, V22, P1212
  • [4] Biomechanical evaluation of a new bone cement for use in vertebroplasty
    Belkoff, SM
    Mathis, JM
    Erbe, EM
    Fenton, DC
    [J]. SPINE, 2000, 25 (09) : 1061 - 1064
  • [5] Bostrom MP, 1997, SPINE S, V22, P38, DOI DOI 10.1097/00007632-199712151-00007
  • [6] Debussche-Depriester C., 1991, NEURORADIOLOGY, V33, P149
  • [8] DICK JC, 1947, J ANAT, V81, P201
  • [9] PREDICTION OF VERTEBRAL STRENGTH BY DUAL PHOTON-ABSORPTIOMETRY AND QUANTITATIVE COMPUTED-TOMOGRAPHY
    ERIKSSON, SAV
    ISBERG, BO
    LINDGREN, JU
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1989, 44 (04) : 243 - 250
  • [10] GANGI A, 1994, AM J NEURORADIOL, V15, P83