A simple chemiluminescence method for determination of chemical oxygen demand values in water

被引:48
作者
Hu, YG [1 ]
Yang, ZY [1 ]
机构
[1] Huazhong Univ Sci & Technol, Inst Environm Sci & Engn, Wuhan 430074, Peoples R China
基金
中国国家自然科学基金;
关键词
chemiluminescence; photodiode; chemical oxygen demand;
D O I
10.1016/j.talanta.2003.11.037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, a low cost chemiluminescence detector with a photodiode instead of photo-multiplier tube (PMT) was developed for environmental monitoring of water quality. Based on the chemiluminescent reaction of luminol-H2O2-Cr3+ system, light emission caused by luminol-H2O2-Cr3+ system was detected by the photodiode, and its intensity caused by the appearance of Cr3+ after samples digestion was proportional to the chemical oxygen demand (COD). Effects for COD determining such as pH, concentrations, interference, and digestion procedures were investigated. The experimental results show that this suggested method uses an instrument that is simpler and cheaper than the previous ones used for the same purpose. The data obtained by the present method were fairly in good agreement with those obtained by the standard reflux titrimetric method. It has been applied to determine real samples with satisfactory results. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:521 / 526
页数:6
相关论文
共 24 条
[1]  
American Public Health Association, 1989, STAND METH EX WAT WA
[2]  
[Anonymous], 1992, STAND METH EX WAT WA
[3]  
ARNOLD CMS, 1987, ANAL CHEM, V59, P1789
[4]   HALIDE ION ENHANCEMENT OF CHROMIUM(III), IRON(II), AND COBALT(II) CATALYSIS OF LUMINOL CHEMI-LUMINESCENCE [J].
CHANG, CA ;
PATTERSON, HH .
ANALYTICAL CHEMISTRY, 1980, 52 (04) :653-656
[5]  
CUESTA A, 1996, SPECTROCHIM ACTA B, V51, P179
[6]  
DAN D, 2002, ANAL CHIM ACTA, V420, P39
[7]   FIA determination of CrIII in tap water with chemiluminescence detection [J].
Economou, A ;
Clark, AK ;
Fielden, PR .
ANALYTICAL COMMUNICATIONS, 1998, 35 (12) :389-390
[8]   RAPID AND PRECISE METHOD FOR THE DETERMINATION OF THE CHEMICAL OXYGEN-DEMAND OF FACTORY EFFLUENT SAMPLES [J].
EDWARDS, SJ ;
ALLEN, M .
ANALYST, 1984, 109 (05) :671-672
[9]   FLOW-INJECTION CHEMILUMINESCENCE METHOD FOR THE SELECTIVE DETERMINATION OF CHROMIUM(III) [J].
ESCOBAR, R ;
LIN, QX ;
GUIRAUM, A ;
DELAROSA, FF .
ANALYST, 1993, 118 (06) :643-647
[10]   SIMULTANEOUS DETERMINATION OF CHROMIUM(III) AND CHROMIUM(VI) IN AQUEOUS-SOLUTIONS BY ION CHROMATOGRAPHY AND CHEMILUMINESCENCE DETECTION [J].
GAMMELGAARD, B ;
JONS, O ;
NIELSEN, B .
ANALYST, 1992, 117 (03) :637-640