Monte Carlo calculations for reference dosimetry of electron beams with the PTW Roos and NE2571 ion chambers

被引:27
作者
Muir, B. R. [1 ]
Rogers, D. W. O. [1 ]
机构
[1] Carleton Univ, Dept Phys, Carleton Lab Radiotherapy Phys, Ottawa, ON K1S 5B6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
beam quality conversion factor; electron beams; dosimetry protocols; ionization chamber; Monte Carlo; QUALITY CORRECTION FACTORS; REPLACEMENT CORRECTION FACTORS; STOPPING-POWER RATIOS; HIGH-ENERGY PHOTON; WALL CORRECTION FACTORS; IONIZATION CHAMBERS; EFFECTIVE POINT; P-WALL; PERTURBATION; SIMULATION;
D O I
10.1118/1.4829577
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate recommendations for reference dosimetry of electron beams and gradient effects for the NE2571 chamber and to provide beam quality conversion factors using Monte Carlo simulations of the PTW Roos and NE2571 ion chambers. Methods: The EGSnrc code system is used to calculate the absorbed dose-to-water and the dose to the gas in fully modeled ion chambers as a function of depth in water. Electron beams are modeled using realistic accelerator simulations as well as beams modeled as collimated point sources from realistic electron beam spectra or monoenergetic electrons. Beam quality conversion factors are calculated with ratios of the doses to water and to the air in the ion chamber in electron beams and a cobalt-60 reference field. The overall ion chamber correction factor is studied using calculations of water-to-air stopping power ratios. Results: The use of an effective point of measurement shift of 1.55 mm from the front face of the PTW Roos chamber, which places the point of measurement inside the chamber cavity, minimizes the difference between R-50, the beam quality specifier, calculated from chamber simulations compared to that obtained using depth-dose calculations in water. A similar shift minimizes the variation of the overall ion chamber correction factor with depth to the practical range and reduces the root-mean-square deviation of a fit to calculated beam quality conversion factors at the reference depth as a function of R-50. Similarly, an upstream shift of 0.34 r(cav) allows a more accurate determination of R-50 from NE2571 chamber calculations and reduces the variation of the overall ion chamber correction factor with depth. The determination of the gradient correction using a shift of 0.22 r(cav) optimizes the root-mean-square deviation of a fit to calculated beam quality conversion factors if all beams investigated are considered. However, if only clinical beams are considered, a good fit to results for beam quality conversion factors is obtained without explicitly correcting for gradient effects. The inadequacy of R-50 to uniquely specify beam quality for the accurate selection of k(Q) factors is discussed. Systematic uncertainties in beam quality conversion factors are analyzed for the NE2571 chamber and amount to between 0.4% and 1.2% depending on assumptions used. Conclusions: The calculated beam quality conversion factors for the PTW Roos chamber obtained here are in good agreement with literature data. These results characterize the use of an NE2571 ion chamber for reference dosimetry of electron beams even in low-energy beams. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:16
相关论文
共 57 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]  
[Anonymous], 2001, IAEA TECHN REP SER
[3]  
[Anonymous], 1984, 37 ICRU
[4]   Wall correction factors, Pwall, for thimble ionization chambers [J].
Buckley, LA ;
Rogers, DWO .
MEDICAL PHYSICS, 2006, 33 (02) :455-464
[5]   CSnrc: Correlated sampling Monte Carlo calculations using EGSnrc [J].
Buckley, LA ;
Kawrakow, I ;
Rogers, DWO .
MEDICAL PHYSICS, 2004, 31 (12) :3425-3435
[6]   Wall correction factors, Pwall, for parallel-plate ionization chambers [J].
Buckley, Lesley A. ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2006, 33 (06) :1788-1796
[7]   R(50) as a beam quality specifier for selecting stopping-power ratios and reference depths for electron dosimetry [J].
Burns, DT ;
Ding, GX ;
Rogers, DWO .
MEDICAL PHYSICS, 1996, 23 (03) :383-388
[8]  
Cojocaru C., 2010, IAEAE2CN182, Book of Extended Synopses for Symp. on Standards, P27
[9]   CALCULATION OF STOPPING POWER RATIOS USING REALISTIC CLINICAL ELECTRON-BEAMS [J].
DING, GX ;
ROGERS, DWO ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :489-501
[10]  
Ding GX, 1995, PIRS0439 NRC