Dosimetric Characteristics of a PIN Diode for Radiotherapy Application

被引:18
作者
Kumar, Rajesh [1 ]
Sharma, S. D. [1 ]
Philomina, A. [1 ]
Topkar, Anita [2 ]
机构
[1] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Bombay 400094, Maharashtra, India
[2] Bhabha Atom Res Ctr, Div Elect, Bombay 400085, Maharashtra, India
关键词
PIN diode; In vivo dosimetry; Diode characterisation; Radiotherapy; INCIDENT;
D O I
10.7785/tcrt.2012.500388
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The PIN diode developed by Bhabha Atomic Research Centre (BARC) was modified for its use as a dosimeter in radiation therapy. For this purpose the diode was mounted on a printed circuit board (PCB) and provided with necessary connections so that its response against irradiation can be recorded by a standard radiotherapy electrometer. The dosimetric characteristics of the diode were studied in Co-60 gamma rays as well as high energy X-rays. The measured sensitivity of this PIN diode is 4 nC/cGy which is about ten times higher than some commercial diode dosimeters. The leakage current from the diode is 0.04 nA. The response of the PIN diode is linear in the range of 20-1000 cGy which covers the full range of radiation dose encountered in radiotherapy treatments. The non-linearity of the diode response is 3.5% at 20 cGy and it is less than 1.5% at higher dose values. Its repeatability is within 0.5%. The angular response variation is about 5.6% within +/- 60 degrees with respect to normal beam incidence. The response of the PIN diode at 6 and 18 MV X-rays varies within 2% with respect to its response at Co-60 gamma rays. The source to surface distance (SSD) dependence of the PIN diode was studied for Co-60 beam. It was found that the response of the diode decreases almost linearly relative to given dose for beams with constant collimator setting but increasing SSD (decreasing dose-rate). Within this study the diode response varied by about 2.5% between the maximum and minimum SSD. The dose-rate dependence of the PIN diode for 6 and 15 MV-rays was studied. The variation in response of diode for both energies in the studied dose range is less than 1%. The field size dependence of the PIN diode response is within 1% with respect to the response of ionisation chamber. These studies indicate that the characteristics of the PIN diode are suitable for use in radiotherapy dosimetry.
引用
收藏
页码:361 / 367
页数:7
相关论文
共 26 条
[1]  
*AAPM, 2005, 87 AAPM
[2]   An audit of radiotherapy patient doses measured with in vivo semiconductor detectors [J].
Adeyemi, A ;
Lord, J .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832) :399-408
[3]  
Alecu R, 1999, Med Dosim, V24, P5, DOI 10.1016/S0958-3947(98)00045-4
[4]  
[Anonymous], MONITOR
[5]   Electronic dosimeters based on solid state detectors [J].
Barthe, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 184 (1-2) :158-189
[6]  
Barthe J., 1999, RADIAT PROT DOSIM, V84, P349
[7]   Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Watts, Ronald J. ;
Ahnesjo, Anders ;
Gibbons, John ;
Li, X. Allen ;
Lowenstein, Jessica ;
Mitra, Raj K. ;
Simon, William E. ;
Zhu, Timothy C. .
MEDICAL PHYSICS, 2008, 35 (09) :4186-4215
[8]   DOSIMETRIC CONTROL OF CONFORMAL TREATMENT OF PAROTID-GLAND TUMORS [J].
ESSERS, M ;
KEUS, R ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (02) :154-162
[9]   IN-VIVO DOSIMETRY DURING CONFORMAL THERAPY OF PROSTATIC-CANCER [J].
ESSERS, M ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :271-279
[10]  
European Society of Radiation Oncology, 2001, ESTRO BOOKLET, V5