Cerebral state index during propofol anesthesia

被引:107
作者
Jensen, Erik W.
Litvan, Hector
Revuelta, Miren
Rodriguez, Bernardo E.
Caminal, Pere
Martinez, Pablo
Vereecke, Hugo
Struys, Michel M. R. F.
机构
[1] Tech Univ Catalonia, Dept Cardiac Anesthesia, Barcelona, Spain
[2] Tech Univ Catalonia, Postoperat Intens Care Unit, Barcelona, Spain
[3] Tech Univ Catalonia, Biomed Engn Res Ctr, Barcelona, Spain
[4] State Univ Ghent Hosp, Dept Anesthesia, B-9000 Ghent, Belgium
[5] State Univ Ghent Hosp, Dept Anesthesia, B-9000 Ghent, Belgium
[6] Univ Ghent, Heymans Inst Pharmacol, Dept Anesthesia, B-9000 Ghent, Belgium
关键词
D O I
10.1097/00000542-200607000-00009
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: The objective of this study was to prospectively test the Cerebral State Index designed for measuring the depth of anesthesia. The Cerebral State Index is calculated using a fuzzy logic combination of four subparameters of the electroencephalographic signal. The performance of the Cerebral State index was compared with that of the Bispectral Index and the A-Line ARX Index. Methods: This study applied raw data from two previously published clinical protocols. The patients in protocol 1 were given a continuous propofol infusion, 300 ml/h, until 80% of burst suppression occurred. In protocol 2, a stepwise increased target-controlled infusion of propofol was administered to patients until loss of response to noxious stimuli while the Observer's Assessment of Alertness and Sedation was registered every 4 min. The Cerebral State index was calculated off-line from the recorded electroencephalographic data. The Spearman rank correlation coefficient between electronic indices and the effect site concentration of propofol was calculated along with the prediction probability of each index to predict the Observer's Assessment of Alertness and Sedation level. Results: The Spearman rank correlation coefficients between the Cerebral State Index, Bispectral Index, and A-Line ARX Index and the propofol effect site concentration were -0.94, -0.89, and -0.82, respectively, in protocol 1, whereas the prediction probability values between the Cerebral State Index, Bispectral Index, and A-Line ARX Index and the Observer's Assessment of Alertness and Sedation score in protocol 2 were 0.92, 0.93, and 0.91, respectively. Conclusion: The Cerebral State Index detects well the graduated levels of propofol anesthesia when compared with the propofol effect site concentration and the Observer's Assessment of Alertness and Sedation score.
引用
收藏
页码:28 / 36
页数:9
相关论文
共 29 条
  • [1] CHERNIK DA, 1990, J CLIN PSYCHOPHARM, V10, P244
  • [2] Recovery of consciousness after thiopental or propofol - Bispectral index and the isolated forearm technique
    Flaishon, R
    Windsor, A
    Sigl, J
    Sebel, PS
    [J]. ANESTHESIOLOGY, 1997, 86 (03) : 613 - 619
  • [3] Bispectral analysis measures sedation and memory effects of propofol, midazolam, isoflurane, and alfentanil in healthy volunteers
    Glass, PS
    Bloom, M
    Kearse, L
    Rosow, C
    Sebel, P
    Manberg, P
    [J]. ANESTHESIOLOGY, 1997, 86 (04) : 836 - 847
  • [4] Fuzzy logic and decision-making in anaesthetics
    Grant, P
    Naesh, O
    [J]. JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 2005, 98 (01) : 7 - 9
  • [5] ANFIS - ADAPTIVE-NETWORK-BASED FUZZY INFERENCE SYSTEM
    JANG, JSR
    [J]. IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1993, 23 (03): : 665 - 685
  • [6] Jensen EW, 1996, METHOD INFORM MED, V35, P256
  • [7] Pitfalls and challenges when assessing the depth of hypnosis during general anaesthesia by clinical signs and electronic indices
    Jensen, EW
    Litvan, H
    Struys, M
    Vazquez, PM
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2004, 48 (10) : 1260 - 1267
  • [8] Identification of causal relations between haemodynamic variables, auditory evoked potentials and isoflurane by means of fuzzy logic
    Jensen, EW
    Nebot, A
    Caminal, P
    Henneberg, SW
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1999, 82 (01) : 25 - 32
  • [9] JENSEN EW, 1998, EUR C INT TECHN SOFT, V3, P1889
  • [10] Application of Bispectral Index® and Narcotrend® index to the measurement of the electroencephalographic effects of isoflurane with and without burst suppression
    Kreuer, S
    Bruhn, J
    Larsen, R
    Grundmann, U
    Shafer, SL
    Wilhelm, W
    [J]. ANESTHESIOLOGY, 2004, 101 (04) : 847 - 854