Blood pressure measurement in the elderly with atrial fibrillation: an observational study comparing different noninvasive sphygmomanometers

被引:0
作者
Mai, Shaojun [1 ]
Zhu, Hailan [1 ]
Li, Meijun [1 ]
Zeng, Yanfang [1 ]
Zhang, Yang [1 ]
Huo, Yanchang [4 ]
Pan, Xiong-Fei [2 ]
Huang, Yuli [3 ,4 ]
机构
[1] Southern Med Univ, Shunde Hosp, Dept Cardiol, Guangdong, Peoples R China
[2] Sichuan Univ, West China Second Univ Hosp, Minist Educ, Key Lab Birth Defects & Related Dis Women & Childr, Chengdu, Peoples R China
[3] George Inst Global Hlth, Newtown, NSW, Australia
[4] Southern Med Univ, Shunde Hosp, Dept Cardiol, Jiazi Rd 1 Lunjiao Town, Foshan 523808, Guangdong, Peoples R China
关键词
atrial fibrillation; blood pressure measurement; noninvasive sphygmomanometer; oscillographic sphygmomanometer; pulse wave sphygmomanometer; HYPERTENSION; ASSOCIATION; PREVENTION; DEVICE;
D O I
10.1177/20406223221137040
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Background:Atrial fibrillation (AF) has affected millions of adults in the world. It is important to monitor and manage blood pressure (BP) in AF patients. The accuracy of BP monitoring in AF patients with noninvasive methods remains questionable, however. Objectives:To compare the accuracy of different noninvasive BP devices (oscillographic sphygmomanometer and pulse wave device) for BP measurement in elderly patients with AF, with a mercury sphygmomanometer as a reference. Design:This study was an observational study. Methods:Patients with AF from the inpatient department of cardiology were included from 1 January to 31 December 2020. BP measurements were performed by two trained nurses using a tee junction connection on the cuff to connect three sphygmomanometers. The Bland-Altman plot analysis was conducted to compare the agreement of BP measurements. We also compared the agreement of BP measurements through metrics such as accuracy, bias, and precision. Results:A total of 202 patients (54.5% female) were included. The Bland-Altman plot analysis showed that the lower and upper limits of agreement (LoAs) of pulse wave/reference were similar to the predefined acceptable clinical limits (10/5 mmHg). The bias and precision in both systolic and diastolic BP were significantly less in pulse wave/reference (a bias of 1.8 and 0.77 mmHg and a precision of 5.20 and 4.66 mmHg, respectively), with corresponding higher accuracy readings (98.51% for P10 in systolic BP and 85.64% for P5 in diastolic BP). Conclusion:A novel noninvasive sphygmomanometer - pulse wave device has a good concordance with a mercury sphygmomanometer in BP monitoring, and may be applicable to perform BP measurements in the elderly with AF.
引用
收藏
页数:8
相关论文
共 26 条
[11]   Sample Size for Assessing Agreement between Two Methods of Measurement by Bland-Altman Method [J].
Lu, Meng-Jie ;
Zhong, Wei-Hua ;
Liu, Yu-Xiu ;
Miao, Hua-Zhang ;
Li, Yong-Chang ;
Ji, Mu-Huo .
INTERNATIONAL JOURNAL OF BIOSTATISTICS, 2016, 12 (02)
[12]   Reliability of Oscillometric Blood Pressure Monitoring in Atrial Fibrillation Patients Admitted for Electric Cardioversion [J].
Maselli, Monica ;
Giantin, Valter ;
Corrado, Domenico ;
Franchin, Alessandro ;
Attanasio, Francesca ;
Pengo, Valentina ;
Tramontano, Alessandra ;
De Toni, Pietro ;
Perissinotto, Egle ;
Manzato, Enzo .
JOURNAL OF CLINICAL HYPERTENSION, 2015, 17 (07) :558-564
[13]   Revised Equations for Estimated GFR From Serum Creatinine in Japan [J].
Matsuo, Seiichi ;
Imai, Enyu ;
Horio, Masaru ;
Yasuda, Yoshinari ;
Tomita, Kimio ;
Nitta, Kosaku ;
Yamagata, Kunihiro ;
Tomino, Yasuhiko ;
Yokoyama, Hitoshi ;
Hishida, Akira .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2009, 53 (06) :982-992
[14]   Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring [J].
Pagonas, Nikolaos ;
Schmidt, Sven ;
Eysel, Joerg ;
Compton, Friederike ;
Hoffmann, Clemens ;
Seibert, Felix ;
Hilpert, Justus ;
Tschoepe, Carsten ;
Zidek, Walter ;
Westhoff, Timm H. .
HYPERTENSION, 2013, 62 (03) :579-584
[15]   Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis [J].
Park, Seong-Hi ;
Choi, Yun-Kyoung .
BLOOD PRESSURE MONITORING, 2020, 25 (01) :2-12
[16]   Measurement of arm blood pressure using different oscillometry manometers compared to auscultatory readings [J].
Shahriari, M ;
Rotenberg, DK ;
Nielsen, JK ;
Wiinberg, N ;
Nielsen, PE .
BLOOD PRESSURE, 2003, 12 (03) :155-159
[17]   Blood pressure measurement in atrial fibrillation: review and meta-analysis of evidence on accuracy and clinical relevance [J].
Stergiou, George S. ;
Kyriakoulis, Konstantinos G. ;
Stambolliu, Emelina ;
Destounis, Antonios ;
Karpettas, Nikos ;
Kalogeropoulos, Petros ;
Kollias, Anastasios .
JOURNAL OF HYPERTENSION, 2019, 37 (12) :2430-2441
[18]   Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis [J].
Stergiou, George S. ;
Kollias, Anastasios ;
Destounis, Antonios ;
Tzamouranis, Dimitrios .
JOURNAL OF HYPERTENSION, 2012, 30 (11) :2074-2082
[19]  
UKNGC, 2019, HYP AD DIAGN MAN NAT
[20]   The Japanese Society of Hypertension Guidelines for the Management of Hypertension (JS']JSH 2019) [J].
Umemura, Satoshi ;
Arima, Hisatomi ;
Arima, Shuji ;
Asayama, Kei ;
Dohi, Yasuaki ;
Hirooka, Yoshitaka ;
Horio, Takeshi ;
Hoshide, Satoshi ;
Ikeda, Shunya ;
Ishimitsu, Toshihiko ;
Ito, Masaaki ;
Ito, Sadayoshi ;
Iwashima, Yoshio ;
Kai, Hisashi ;
Kamide, Kei ;
Kanno, Yoshihiko ;
Kashihara, Naoki ;
Kawano, Yuhei ;
Kikuchi, Toru ;
Kitamura, Kazuo ;
Kitazono, Takanari ;
Kohara, Katsuhiko ;
Kudo, Masataka ;
Kumagai, Hiroo ;
Matsumura, Kiyoshi ;
Matsuura, Hideo ;
Miura, Katsuyuki ;
Mukoyama, Masashi ;
Nakamura, Satoko ;
Ohkubo, Takayoshi ;
Ohya, Yusuke ;
Okura, Takafumi ;
Rakugi, Hiromi ;
Saitoh, Shigeyuki ;
Shibata, Hirotaka ;
Shimosawa, Tatsuo ;
Suzuki, Hiromichi ;
Takahashi, Shori ;
Tamura, Kouichi ;
Tomiyama, Hirofumi ;
Tsuchihashi, Takuya ;
Ueda, Shinichiro ;
Uehara, Yoshinari ;
Urata, Hidenori ;
Hirawa, Nobuhito .
HYPERTENSION RESEARCH, 2019, 42 (09) :1235-1481