Optimal slice thickness for object detection with longitudinal partial volume effects in computed tomography

被引:26
作者
Monnin, Pascal [1 ,2 ]
Sfameni, Nicolas [1 ]
Gianoli, Achille [1 ]
Ding, Sandrine [1 ]
机构
[1] Haute Ecole Sante Vaud HESAV, Filiere TRM, Lausanne, Switzerland
[2] Univ Hosp Lausanne CHUV, Inst Radiat Phys IRA, Lausanne, Switzerland
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2017年 / 18卷 / 01期
关键词
artifacts and distortion; computed tomography; contrast; noise; spatial resolution; SPIRAL CT; SENSITIVITY PROFILES; MULTIDETECTOR-CT; PULMONARY NODULES; RESOLUTION; PHANTOM; NOISE; ALGORITHM; OBSERVER; LESIONS;
D O I
10.1002/acm2.12005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Longitudinal partial volume effects (z-axial PVE), which occur when an object partly occupies a slice, degrade image resolution and contrast in computed tomography (CT). Z-axial PVE is unavoidable for subslice objects and reduces their contrast according to their fraction contained within the slice. This effect can be countered using a smaller slice thickness, but at the cost of an increased image noise or radiation dose. The aim of this study is to offer a tool for optimizing the reconstruction parameters (slice thickness and slice spacing) in CT protocols in the case of partial volume effects. This optimization is based on the tradeoff between axial resolution and noise. For that purpose, we developed a simplified analytical model investigating the average statistical effect of z-axial PVE on contrast and contrast-to-noise ratio (CNR). A Catphan 500 phantom was scanned with various pitches and CTDI and reconstructed with different slice thicknesses to assess the visibility of subslice targets that simulate low contrast anatomical features present in CT exams. The detectability score of human observers was used to rank the perceptual image quality against the CNR. Contrast and CNR reduction due to z-axial PVE measured on experimental data were first compared to numerical calculations and then to the analytical model. Compared to numerical calculations, the simplified algebraic model slightly overestimated the contrast but the differences remained below 5%. It could determine the optimal reconstruction parameters that maximize the objects visibility for a given dose in the case of z-axial PVE. An optimal slice thickness equal to three-fourth of the object width was correctly proposed by the model for nonoverlapping slices. The tradeoff between detectability and dose is maximized for a slice spacing of half the slice thickness associated with a slice width equal to the characteristic object width.
引用
收藏
页码:251 / 259
页数:9
相关论文
共 24 条
  • [1] Detection of liver metastases under 2 cm: comparison of different acquisition protocols in four row multidetector-CT (MDCT)
    Abdelmoumene, A
    Chevallier, P
    Chalaron, M
    Schneider, F
    Verdun, FR
    Frascarolo, P
    Meuli, R
    Schnyder, P
    Denys, A
    [J]. EUROPEAN RADIOLOGY, 2005, 15 (09) : 1881 - 1887
  • [2] BROOKS R A, 1976, Medical Physics (Woodbury), V3, P237, DOI 10.1118/1.594240
  • [3] DETERMINATION OF SPIRAL CT SLICE SENSITIVITY PROFILES USING A POINT RESPONSE PHANTOM
    DAVROS, WJ
    HERTS, BR
    WALMSLEY, JJ
    OBUCHOWSKI, NA
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1995, 19 (05) : 838 - 843
  • [4] Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: Phantom study
    Funama, Y
    Awai, K
    Nakayama, Y
    Kakei, K
    Nagasue, N
    Shimamura, M
    Sato, N
    Sultana, S
    Morishita, S
    Yamashita, Y
    [J]. RADIOLOGY, 2005, 237 (03) : 905 - 910
  • [5] Automated assessment of low contrast sensitivity for CT systems using a model observer
    Hernandez-Giron, I.
    Geleijns, J.
    Calzado, A.
    Veldkamp, W. J. H.
    [J]. MEDICAL PHYSICS, 2011, 38 : S25 - S35
  • [6] ICRP, 2007, ANN ICRP, V37, P32
  • [7] Presampling, algorithm factors, and noise: Considerations for CT in particular and for medical imaging in general
    Kachelriess, M
    Kalender, WA
    [J]. MEDICAL PHYSICS, 2005, 32 (05) : 1321 - 1334
  • [8] A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS
    KALENDER, WA
    POLACIN, A
    SUSS, C
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) : 167 - 176
  • [9] RECONSTRUCTED HELICAL CT SCANS - IMPROVEMENT IN Z-AXIS RESOLUTION COMPARED WITH OVERLAPPED AND NONOVERLAPPED CONVENTIONAL CT SCANS
    KASALES, CJ
    HOPPER, KD
    ARIOLA, DN
    TENHAVE, TR
    MEILSTRUP, JW
    MAHRAJ, RPM
    VANHOOK, D
    WESTACOTT, S
    SEFCZEK, RJ
    BARR, JD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (05) : 1281 - 1284
  • [10] MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA
    LANDIS, JR
    KOCH, GG
    [J]. BIOMETRICS, 1977, 33 (01) : 159 - 174