Microneedle array for transdermal biological fluid extraction and in situ analysis

被引:249
作者
Mukerjee, E
Collins, SD
Isseroff, RR
Smith, RL
机构
[1] Univ Calif Davis, MISL, Davis, CA 95616 USA
[2] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[3] Univ Calif Davis, Sch Med, Dept Dermatol, Davis, CA 95616 USA
关键词
silicon microneedle; transdermal fluid movement; ISF extraction;
D O I
10.1016/j.sna.2003.11.008
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents the design, fabrication and testing of a hollow microneedle array with integrated, fluidic microchannels. The microneedles are machined from single crystal silicon to a shank height of 250-350 mum with 300 mum center-to-center spacing. The needle size, density and shape are controlled by independent processing steps. Piercing of the protective stratum corenum layer of the skin by the microneedles, providing access to the living epithelial layers underneath, is successfully demonstrated. Filling of the integrated microchannels by capillary action with both non-biological fluids (glycerol, ethanol, surrogate interstitial fluid (ISF) and water) as well as biological fluids (ISF and whole blood) is demonstrated. The ability to extract interstitial fluid from human skin is successfully achieved and verified by the in situ measurement of glucose concentration. The microchip presented here represents the first step towards the realization of a transdermal, ISF extraction and analysis microsystem. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:267 / 275
页数:9
相关论文
共 20 条
[1]   A SILICON-BASED, 3-DIMENSIONAL NEURAL INTERFACE - MANUFACTURING PROCESSES FOR AN INTRACORTICAL ELECTRODE ARRAY [J].
CAMPBELL, PK ;
JONES, KE ;
HUBER, RJ ;
HORCH, KW ;
NORMANN, RA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (08) :758-768
[2]   Surface micromachined metallic microneedles [J].
Chandrasekaran, S ;
Brazzle, JD ;
Frazier, AB .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2003, 12 (03) :281-288
[3]  
Elias P M, 1989, Curr Probl Dermatol, V18, P10
[4]  
GARDENIERS JGE, 2002, MEMS 2002 IEEE INT C
[5]  
Ghandhi S.K., 1994, VLSI FABRICATION PRI
[6]  
GRISS P, 2002, MEMS 2002 IEEE INT C
[7]   Microfabricated microneedles: A novel approach to transdermal drug delivery [J].
Henry, S ;
McAllister, DV ;
Allen, MG ;
Prausnitz, MR .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1998, 87 (08) :922-925
[8]   ASSESSMENT OF TISSUE-REACTIONS TO BIOMATERIALS - A MODEL SYSTEM USING THE FOREIGN-BODY RESPONSE IN THE MOUSE LUNG TO INTRAVENOUSLY INJECTED DIVINYL COPOLYMER BEADS [J].
HOOD, CI ;
SCHOEN, FJ ;
COLEMAN, SE ;
MICKLEY, LD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (09) :1031-1041
[9]   Fully integrated PCR-capillary electrophoresis microsystem for DNA analysis [J].
Lagally, ET ;
Emrich, CA ;
Mathies, RA .
LAB ON A CHIP, 2001, 1 (02) :102-107
[10]   Electrophoresis separation in open microchannels. A method for coupling electrophoresis with MALDI-MS [J].
Liu, J ;
Tseng, K ;
Garcia, B ;
Lebrilla, CB ;
Mukerjee, E ;
Collins, S ;
Smith, R .
ANALYTICAL CHEMISTRY, 2001, 73 (09) :2147-2151