Evaluation of a Low-Cost Bubble CPAP System Designed for Resource-Limited Settings

被引:11
作者
Bennett, Desmond J. [1 ]
Carroll, Ryan W. [4 ]
Kacmarek, Robert M. [1 ,2 ,3 ]
机构
[1] Massachusetts Gen Hosp, Dept Resp Care, 55 Fruit St, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Dept Anesthesia Crit Care & Pain Med, Boston, MA 02114 USA
[3] Harvard Med Sch, Boston, MA USA
[4] Massachusetts Gen Hosp, Dept Pediat Crit Care Med, Boston, MA 02114 USA
关键词
CPAP; bubbleCPAP; respiratory distress; low-resource settings; Uganda; neonate; POSITIVE AIRWAY PRESSURE; NONINVASIVE VENTILATION; RESPIRATORY-DISTRESS; LEAK COMPENSATION; LUNG MODEL; PREMATURE-INFANTS; PRETERM INFANTS; COUNTRIES; SUPPORT; CHILDREN;
D O I
10.4187/respcare.05762
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Respiratory compromise is a leading contributor to global neonatal death. CPAP is a method of treatment that helps maintain lung volume during expiration, promotes comfortable breathing, and improves oxygenation. Bubble CPAP is an effective alternative to standard CPAP. We sought to determine the reliability and functionality of a low-cost bubble CPAP device designed for low-resource settings. METHODS: The low-cost bubble CPAP device was compared to a commercially available bubble CPAP system. The devices were connected to a lung simulator that simulated neonates of 4 different weights with compromised respiratory mechanics (similar to 1, similar to 3, similar to 5, and similar to 10 kg). The devices' abilities to establish and maintain pressure and flow under normal conditions as well as under conditions of leak were compared. Multiple combinations of pressure levels (5, 8, and 10 cm H2O) and flow levels (3, 6, and 10 L/min) were tested. The endurance of both devices was also tested by running the systems continuously for 8 h and measuring the changes in pressure and flow. RESULTS: Both devices performed equivalently during the no-leak and leak trials. While our testing revealed individual differences that were statistically significant and clinically important (>10% difference) within specific CPAP and flow-level settings, no overall comparisons of CPAP or flow were both statistically significant and clinically important. Each device delivered pressures similar to the desired pressures, although the flows delivered by both machines were lower than the set flows in most trials. During the endurance trials, the low-cost device was marginally better at maintaining pressure, while the commercially available device was better at maintaining flow. CONCLUSIONS: The low-cost bubble CPAP device evaluated in this study is comparable to a bubble CPAP system used in developed settings. Extensive clinical trials, however, are necessary to confirm its effectiveness. (c) 2018 Daedalus Enterprises
引用
收藏
页码:395 / 403
页数:9
相关论文
共 30 条
  • [1] [Anonymous], 2015, 3 CAUS DOP ATHL
  • [2] A High-Value, Low-Cost Bubble Continuous Positive Airway Pressure System for Low-Resource Settings: Technical Assessment and Initial Case Reports
    Brown, Jocelyn
    Machen, Heather
    Kawaza, Kondwani
    Mwanza, Zondiwe
    Iniguez, Suzanne
    Lang, Hans
    Gest, Alfred
    Kennedy, Neil
    Miros, Robert
    Richards-Kortum, Rebecca
    Molyneux, Elizabeth
    Oden, Maria
    [J]. PLOS ONE, 2013, 8 (01):
  • [3] Cost-effectiveness analysis of a low-cost bubble CPAP device in providing ventilatory support for neonates in Malawi - a preliminary report
    Chen, Ariel
    Deshmukh, Ashish A.
    Richards-Kortum, Rebecca
    Molyneux, Elizabeth
    Kawaza, Kondwani
    Cantor, Scott B.
    [J]. BMC PEDIATRICS, 2014, 14
  • [4] Nasal continuous positive airway pressure and outcomes of preterm infants
    De Klerk, AM
    De Klerk, RK
    [J]. JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2001, 37 (02) : 161 - 167
  • [6] CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) USING GREGORY BOX
    DUNN, PM
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY OF MEDICINE-LONDON, 1974, 67 (04): : 245 - 247
  • [7] PULMONARY MECHANICS IN NORMAL INFANTS AND YOUNG-CHILDREN DURING 1ST 5 YEARS OF LIFE
    GERHARDT, T
    HEHRE, D
    FELLER, R
    REIFENBERG, L
    BANCALARI, E
    [J]. PEDIATRIC PULMONOLOGY, 1987, 3 (05) : 309 - 316
  • [8] DIMENSIONS OF THE GROWING TRACHEA RELATED TO AGE AND GENDER
    GRISCOM, NT
    WOHL, MEB
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 146 (02) : 233 - 237
  • [9] Development and validation of a simple algorithm for initiation of CPAP in neonates with respiratory distress in Malawi
    Hundalani, Shilpa G.
    Richards-Kortum, Rebecca
    Oden, Maria
    Kawaza, Kondwani
    Gest, Alfred
    Molyneux, Elizabeth
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2015, 100 (04): : F332 - F336
  • [10] Performance of Leak Compensation in All-Age ICU Ventilators During Volume-Targeted Neonatal Ventilation: A Lung Model Study
    Itagaki, Taiga
    Bennett, Desmond J.
    Chenelle, Christopher T.
    Fisher, Daniel F.
    Kacmarek, Robert M.
    [J]. RESPIRATORY CARE, 2017, 62 (01) : 10 - 21