A Double-Blind, Prospective, Randomized Controlled Trial Comparing Highly Cross-Linked and Conventional Polyethylene in Primary Total Hip Arthroplasty

被引:55
|
作者
Calvert, George T. [1 ]
Devane, Peter A. [1 ,2 ]
Fielden, Jan [1 ,3 ]
Adams, Kathryn [1 ,2 ]
Horne, J. Geoffrey [1 ,2 ]
机构
[1] Wellington Sch Med, Dept Surg, Div Orthoped Surg, Wellington, New Zealand
[2] Capital Coast DHB Ltd, Dept Orthoped, Wellington, New Zealand
[3] Massey Univ, Sch Hlth Sci, Wellington, New Zealand
关键词
total hip arthroplasty; cross-linked polyethylene; polyethylene wear; osteolysis; prospective randomized trial; MOLECULAR-WEIGHT POLYETHYLENE; ACETABULAR CUPS; WEAR; OSTEOLYSIS; REPLACEMENT; CRACKING; LINERS;
D O I
10.1016/j.arth.2008.02.011
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Highly cross-linked polyethylene was developed to reduce articular bearing wear in total hip arthroplasty. In vitro studies have shown reduced wear in comparison with conventional polyethylene. A double-blind, prospective, randomized trial was performed comparing highly cross-linked and conventional polyethylene in 1 19 patients. The primary outcome variables were linear, 3-dimensional, and volumetric wear as determined by analysis of digitized radiographs using validated wear measurement software. Linear, 3-dimensional, and volumetric wear rates were significantly less in the highly cross-linked polyethylene group between 6 months and 4 years postoperatively (P < .05). The data presented here show that highly crosslinked polyethylene reduces short-term polyethylene wear. The intermediate and long-term clinical results of highly cross-linked polyethylene remain unknown pending further follow-up.
引用
收藏
页码:505 / 510
页数:6
相关论文
共 50 条
  • [1] Clinical Results of the Wear Performance of Cross-Linked Polyethylene in Total Hip Arthroplasty Prospective Randomized Trial
    Ise, Kentaro
    Kawanabe, Keiichi
    Tamura, Jiro
    Akiyama, Haruhiko
    Goto, Koji
    Nakamura, Takashi
    JOURNAL OF ARTHROPLASTY, 2009, 24 (08) : 1216 - 1220
  • [2] Wear Rate of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Randomized Controlled Trial
    McCalden, Richard W.
    MacDonald, Steven J.
    Rorabeck, Cecil H.
    Bourne, Robert B.
    Chess, David G.
    Charron, Kory D.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (04) : 773 - 782
  • [3] The Seven-Year Wear of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Double-Blind, Randomized Controlled Trial Using Radiostereometric Analysis
    Thomas, Geraint E. R.
    Simpson, David J.
    Mehmood, Shahid
    Taylor, Adrian
    McLardy-Smith, Peter
    Gill, Harinderjit Singh
    Murray, David W.
    Glyn-Jones, Sion
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (08) : 716 - 722
  • [4] Durability of Highly Cross-Linked Polyethylene in Total Hip and Total Knee Arthroplasty
    Dion, Neil T.
    Bragdon, Charles
    Muratoglu, Orhun
    Freiberg, Andrew A.
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2015, 46 (03) : 321 - +
  • [5] A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty
    Mu, Zongyou
    Tian, Jialiang
    Wu, Taixiang
    Yang, Jing
    Pei, Fuxing
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (03) : 599 - 604
  • [6] Wear Kinetics of Highly Cross-Linked and Conventional Polyethylene Are Similar at Medium-term Follow-Up After Primary Total Hip Arthroplasty
    Mayer, Constantin
    Rommelmann, Moritz
    Behringer, Michael
    Jaeger, Marcus
    Krauspe, Ruediger
    Zilkens, Christoph
    JOURNAL OF ARTHROPLASTY, 2018, 33 (08) : 2671 - 2676
  • [7] A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up
    Engh, Charles A., Jr.
    Hopper, Robert H., Jr.
    Huynh, Cathy
    Ho, Henry
    Sritulanondha, Supatra
    Engh, Charles A., Sr.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (08) : 2 - 7
  • [8] Midterm Results of Primary Total Hip Arthroplasty Using Highly Cross-Linked Polyethylene
    Lee, Ji-Ho
    Lee, Byung Woo
    Lee, Byoung-Joo
    Kim, Shin-Yoon
    JOURNAL OF ARTHROPLASTY, 2011, 26 (07) : 1014 - 1019
  • [9] Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty?
    Glyn-Jones, Sion
    Saac, Sherif
    Hauptfleisch, Jennifer
    McLardy-Smith, Peter
    Murray, David W.
    Singh, Harinderjit
    JOURNAL OF ARTHROPLASTY, 2008, 23 (03) : 337 - 343
  • [10] Conventional Versus Cross-Linked Polyethylene for Total Hip Arthroplasty
    Surace, Michele F.
    Monestier, Luca
    Vulcano, Ettore
    Harwin, Steven F.
    Cherubino, Paolo
    ORTHOPEDICS, 2015, 38 (09) : 556 - 561