共 30 条
Multilevel hybrid 2D strain imaging algorithm for ultrasound sector/phased arrays
被引:24
作者:
Chen, Hao
[1
,2
]
Varghese, Tomy
[1
,2
]
机构:
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Elect & Comp Engn, Madison, WI 53706 USA
关键词:
biological tissues;
biomechanics;
biomedical transducers;
biomedical ultrasonics;
cardiology;
liver;
medical image processing;
ultrasonic propagation;
ultrasonic transducer arrays;
IN-VIVO;
RADIOFREQUENCY ABLATION;
THEORETICAL FRAMEWORK;
ELASTOGRAPHY;
DISPLACEMENT;
TISSUE;
NOISE;
ELASTICITY;
REDUCTION;
TRACKING;
D O I:
10.1118/1.3121426
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Two-dimensional (2D) cross-correlation algorithms are necessary to estimate local displacement vector information for strain imaging. However, most of the current two-dimensional cross-correlation algorithms were developed for linear array transducers. Although sector and phased array transducers are routinely used for clinical imaging of abdominal and cardiac applications, strain imaging for these applications has been performed using one-dimensional (1D) cross-correlation analysis. However, one-dimensional cross-correlation algorithms are unable to provide accurate and precise strain estimation along all the angular insonification directions which can range from -45 degrees to 45 degrees with sector and phased array transducers. In addition, since sector and phased array based images have larger separations between beam lines as the pulse propagates deeper into tissue, signal decorrelation artifacts with deformation or tissue motion are more pronounced. In this article, the authors propose a multilevel two-dimensional hybrid algorithm for ultrasound sector and phased array data that demonstrate improved tracking and estimation performance when compared to the traditional 1D cross-correlation or 2D cross-correlation based methods. Experimental results demonstrate that the signal-to-noise and contrast-to-noise ratio estimates improve significantly for smaller window lengths for the hybrid method when compared to the currently used one-dimensional or two-dimensional cross-correlation algorithms. Strain imaging results on ex vivo thermal lesions created in liver tissue and in vivo on cardiac short-axis views demonstrate the improved image quality obtained with this method.
引用
收藏
页码:2098 / 2106
页数:9
相关论文