Evaluation of dead-time corrections for post-radionuclide-therapy 177Lu quantitative imaging with low-energy high-resolution collimators

被引:9
作者
Celler, Anna [1 ]
Piwowarska-Bilska, Hanna [3 ]
Shcherbinin, Sergey [1 ]
Uribe, Carlos [2 ]
Mikolajczak, Renata [4 ]
Birkenfeld, Bozena [3 ]
机构
[1] Univ British Columbia, Dept Radiol, Vancouver, BC, Canada
[2] Univ British Columbia, Dept Phys & Astron, Vancouver, BC V5Z 1M9, Canada
[3] Pomeranian Med Univ, Dept Nucl Med, Szczecin, Poland
[4] Radioisotope Ctr POLATOM, Natl Ctr Nucl Res, Otwock, Poland
基金
加拿大自然科学与工程研究理事会;
关键词
bremsstrahlung; dead time; internal dosimetry; Lu-177; quantitative imaging; radionuclide therapy; ANGER CAMERA DEADTIME; GAMMA-CAMERA; LOSSES; SPECT; SIMULATION; I-131;
D O I
10.1097/MNM.0000000000000011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesDead-time (DT) effects rarely cause problems in diagnostic single-photon emission computed tomography (SPECT) studies; however, in post-radionuclide-therapy imaging, DT can be substantial. Therefore, corrections may be necessary if quantitative images are used in image-based dosimetry or for evaluation of therapy outcomes. This task is particularly challenging if low-energy collimators are used. Our goal was to design a simple method to determine the dead-time correction factor (DTCF) without the need for phantom experiments and complex calculations.MethodsPlanar and SPECT/CT scans of a water phantom containing a 70 ml bottle filled with lutetium-177 (Lu-177) were acquired over 60 days. Two small Lu-177 markers were used in all scans. The DTCF based on the ratio of observed to true count rates measured over the entire spectrum and using photopeak primary photons only was estimated for phantom (DT present) and marker (no DT) scans. In addition, variations in counts in SPECT projections (potentially caused by varying bremsstrahlung and scatter) were investigated.ResultsFor count rates that were about two-fold higher than typically seen in post-therapy Lu-177 scans, the maximum DTCF reached a level of about 17%. The DTCF values determined directly from the phantom experiments using the total energy spectrum and photopeak counts only were equal to 13 and 16%, respectively. They were closely matched by those from the proposed marker-based method, which uses only two energy windows and measures photopeak primary photons (15-17%).ConclusionA simple, marker-based method allowing for determination of the DTCF in high-activity Lu-177 imaging studies has been proposed and validated using phantom experiments.
引用
收藏
页码:73 / 87
页数:15
相关论文
共 23 条
  • [1] ARNOLD JE, 1974, J NUCL MED, V15, P412
  • [2] Quantitative 177Lu SPECT (QSPECT) imaging using a commercially available SPECT/CT system
    Beauregard, Jean-Mathieu
    Hofman, Michael S.
    Pereira, Jucilene M.
    Eu, Peter
    Hicks, Rodney J.
    [J]. CANCER IMAGING, 2011, 11 (01) : 56 - 66
  • [3] Correction of count losses due to deadtime on a DST-XLi (SMVi-GE) camera during dosimetric studies in patients injected with iodine-131
    Delpon, G
    Ferrer, L
    Lisbona, A
    Bardiès, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07) : N79 - N90
  • [4] PERFORMANCE OF A SINGLE-CRYSTAL DIGITAL GAMMA-CAMERA FOR 1ST PASS CARDIAC STUDIES
    DOBBELEIR, A
    FRANKEN, PR
    HAM, HR
    BRIHAYE, C
    GUILLAUME, M
    KNAPP, FF
    VANDEVIVERE, J
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 1991, 12 (01) : 27 - 34
  • [5] CORRECTION METHOD FOR HIGH-COUNT-RATE QUANTITATIVE RADIONUCLIDE ANGIOGRAPHY
    FREEDMAN, GS
    DWYER, A
    KINSELLA, T
    [J]. RADIOLOGY, 1972, 104 (03) : 713 - &
  • [6] GUY MJ, 2000, IEEE NSS MIC P 15 20, P52
  • [7] Heard S, 2004, IEEE NUCL SCI CONF R, P3579
  • [8] Inoue Y, 1999, Radiat Med, V17, P201
  • [9] Inoue Y, 1998, Radiat Med, V16, P149
  • [10] Estimation of deadtime in imaging human subjects
    Inoue, Y
    Ohtake, T
    Yoshikawa, K
    Nishikawa, J
    Sasaki, Y
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (09) : 1232 - 1237