Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system

被引:10
作者
Hasani, Mohsen [1 ]
Farhood, Bagher [2 ]
Ghorbani, Mahdi [3 ]
Naderi, Hamideh [4 ]
Saadatmand, Sepideh [4 ]
Zandi, Saeed Karimkhani [4 ]
Knaup, Courtney [5 ]
机构
[1] Univ Tehran Med Sci, Canc Inst, Canc Res Ctr, Dept Radiotherapy Phys, Tehran, Iran
[2] Kashan Univ Med Sci, Fac Paramed Sci, Dept Med Phys & Radiol, Kashan, Iran
[3] Shahid Beheshti Univ Med Sci, Canc Res Ctr, Tehran, Iran
[4] Qom Univ Med Sci, Canc Inst, Dept Radiotherapy Phys, Qom, Iran
[5] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
CT-RED curve; Radiotherapy; Treatment planning system; Audit; Monaco; RADIATION-THERAPY; QUALITY-ASSURANCE; CT-NUMBER; ACCURACY; PHOTON; PHANTOM; PERFORMANCE; CALIBRATION; PARAMETERS; IMPACT;
D O I
10.1007/s13246-019-00745-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This study aims to evaluate the effect of CT-RED curve on the dose calculation by the Monaco radiotherapy TPS. Four CT images of the CIRS phantom were obtained by different CT scanners. The accuracy of the dose calculation in the three algorithms of the Monaco TPS (Monte Carlo, collapse cone, and pencil beam) is also evaluated based on TECDOC 1583. The CT-RED curves from the CT scanners were transferred to the Monaco TPS to audit the different algorithms of the TPS. The dose values were measured with an ionization chamber in the CIRS phantom. Then, the dose values were calculated by the Monaco algorithms in the corresponding points. For the Monaco TPS and based on TECDOC 1583, the accuracy of the dose calculation in all the three algorithms is within the agreement criteria for most of the points evaluated. For low dose regions, the differences between the calculated and measured dose values are higher than the agreement criteria in a number of points. For the majority of the points, the algorithms underestimate the calculated dose values. It was also found that the use of different CT-RED curves can lead to minor discrepancies in the dose calculation by the Monaco TPS, especially in low dose regions. However, it appears that these differences are not clinically significant in most of the cases.
引用
收藏
页码:489 / 502
页数:14
相关论文
共 47 条
  • [1] Andreo P, 2000, 398 IAEA TRS
  • [2] Andreo P, 2004, TECHNICAL REPORT SER, V430
  • [3] Anjum MN, 2008, IRAN J RADIAT RES, V5, P169
  • [4] [Anonymous], 2008, 1583 TECDOC INT AT E
  • [5] [Anonymous], 2004, 85 AM ASS PHYS MED T
  • [6] [Anonymous], 2002, 74 AAPM
  • [7] [Anonymous], 2007, 1540 TECDOC INT AT E
  • [8] FACTORS AFFECTING THE MEASUREMENT OF SIZE AND CT-NUMBER IN COMPUTED-TOMOGRAPHY
    BAXTER, BS
    SORENSON, JA
    [J]. INVESTIGATIVE RADIOLOGY, 1981, 16 (04) : 337 - 341
  • [9] Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179
    Bissonnette, Jean-Pierre
    Balter, Peter A.
    Dong, Lei
    Langen, Katja M.
    Lovelock, D. Michael
    Miften, Moyed
    Moseley, Douglas J.
    Pouliot, Jean
    Sonke, Jan-Jakob
    Yoo, Sua
    [J]. MEDICAL PHYSICS, 2012, 39 (04) : 1946 - 1963
  • [10] PATIENT-SPECIFIC 3D PRETREATMENT AND POTENTIAL 3D ONLINE DOSE VERIFICATION OF MONTE CARLO-CALCULATED IMRT PROSTATE TREATMENT PLANS
    Boggula, Ramesh
    Jahnke, Lennart
    Wertz, Hansjoerg
    Lohr, Frank
    Wenz, Frederik
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (04): : 1168 - 1175