Accuracy of Peripheral Thermometers for Estimating Temperature A Systematic Review and Meta-analysis

被引:161
作者
Niven, Daniel J. [1 ]
Gaudet, Jonathan E. [1 ]
Laupland, Kevin B. [2 ]
Mrklas, Kelly J. [3 ]
Roberts, Derek J. [4 ]
Stelfox, Henry Thomas [5 ]
机构
[1] Peter Lougheed Ctr, Dept Crit Care Med, Calgary, AB T1Y 6J4, Canada
[2] Royal Inland Hosp, Dept Med, Kamloops, BC V2C 2T1, Canada
[3] Foothills Med Ctr, Calgary, AB T2N 2T9, Canada
[4] Foothills Med Ctr, Dept Crit Care Med, Calgary, AB T2N 5A1, Canada
[5] Univ Calgary, Teaching Res & Wellness Bldg, Calgary, AB T2N 4Z6, Canada
基金
加拿大健康研究院;
关键词
INFRARED EAR THERMOMETRY; TYMPANIC MEMBRANE TEMPERATURES; PULMONARY-ARTERY CORE; BODY-TEMPERATURE; TEMPORAL-ARTERY; CRITICALLY-ILL; RECTAL TEMPERATURES; ANTIPYRETIC TREATMENT; BLADDER THERMOMETRY; SKIN TEMPERATURES;
D O I
10.7326/M15-1150
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Body temperature is commonly used to screen patients for infectious diseases, establish diagnoses, monitor therapy, and guide management decisions. Purpose: To determine the accuracy of peripheral thermometers for estimating core body temperature in adults and children. Data Sources: MEDLINE, EMBASE, Cochrane Central Register of Controlled Trials, and CINAHL Plus from inception to July 2015. Study Selection: Prospective studies comparing the accuracy of peripheral (tympanic membrane, temporal artery, axillary, or oral) thermometers with central (pulmonary artery catheter, urinary bladder, esophageal, or rectal) thermometers. Data Extraction: 2 reviewers extracted data on study characteristics, methods, and outcomes and assessed the quality of individual studies. Data Synthesis: 75 studies (8682 patients) were included. Most studies were at high or unclear risk of patient selection bias (74%) or index test bias (67%). Compared with central thermometers, peripheral thermometers had pooled 95% limits of agree-ment (random-effects meta-analysis) outside the predefined clinically acceptable range (+/- 0.5 degrees C), especially among patients with fever (-1.44 degrees C to 1.46 degrees C for adults; -1.49 degrees C to 0.43 degrees C for children) and hypothermia (-2.07 degrees C to 1.90 degrees C for adults; no data for children). For detection of fever (bivariate random-effects meta-analysis), sensitivity was low (64% [95% CI, 55% to 72%]; I-2 = 95.7%; P < 0.001) but specificity was high (96% [CI, 93% to 97%]; I-2 = 96.3%; P < 0.001). Only 1 study reported sensitivity and specificity for the detection of hypothermia. Limitations: High-quality data for some temperature measurement techniques are limited. Pooled data are associated with interstudy heterogeneity that is not fully explained by stratified and metaregression analyses. Conclusion: Peripheral thermometers do not have clinically acceptable accuracy and should not be used when accurate measurement of body temperature will influence clinical decisions. Primary Funding Source: None.
引用
收藏
页码:768 / +
页数:24
相关论文
共 136 条
[41]  
Fulbrook P, 1997, Intensive Crit Care Nurs, V13, P266, DOI 10.1016/S0964-3397(97)80425-9
[42]  
Giuffre M, 1990, Appl Nurs Res, V3, P52
[43]  
Giuliano K K, 2000, Am J Crit Care, V9, P254
[44]   Temperature measurement in critically ill orally intubated adults: A comparison of pulmonary artery core, tympanic, and oral methods [J].
Giuliano, KK ;
Scott, SS ;
Elliot, S ;
Giuliano, AJ .
CRITICAL CARE MEDICINE, 1999, 27 (10) :2188-2193
[45]  
GREEN M M, 1989, Journal of Emergency Medicine, V7, P437, DOI 10.1016/0736-4679(89)90136-4
[46]  
Haddadin R. B., 2007, Eastern Mediterranean Health Journal, V13, P1060
[47]   Tympanic temperature during therapeutic hypothermia [J].
Hasper, D. ;
Nee, J. ;
Schefold, J. C. ;
Krueger, A. ;
Storm, C. .
EMERGENCY MEDICINE JOURNAL, 2011, 28 (06) :483-485
[48]  
Hebbar Kiran, 2005, Pediatr Crit Care Med, V6, P557, DOI 10.1097/01.PCC.0000163671.69197.16
[49]   WHAT IS FEVER - NORMAL TEMPERATURE IN INFANTS LESS THAN 3 MONTHS OLD [J].
HERZOG, LW ;
COYNE, LJ .
CLINICAL PEDIATRICS, 1993, 32 (03) :142-146
[50]   Quantifying heterogeneity in a meta-analysis [J].
Higgins, JPT ;
Thompson, SG .
STATISTICS IN MEDICINE, 2002, 21 (11) :1539-1558