One-Year Angiographic Follow-Up after WEB-SL Endovascular Treatment of Wide-Neck Bifurcation Intracranial Aneurysms

被引:40
作者
Sivan-Hoffmann, R. [1 ]
Gory, B. [1 ]
Riva, R. [1 ]
Labeyrie, P. -E [1 ]
Signorelli, F. [2 ]
Eldesouky, I. [1 ]
Gonike-Sadeh, U. [1 ]
Armoiry, X. [3 ]
Turjman, F. [1 ]
机构
[1] Hosp Civils Lyon, Hop Neurol Pierre Wertheimer, Dept Intervent Neuroradiol, F-69677 Bron, France
[2] Hosp Civils Lyon, Hop Neurol Pierre Wertheimer, Dept Neurosurg, F-69677 Bron, France
[3] Hosp Civils Lyon, Delegat Rech Clin & Innovat Cellule Innovat, UMR CNRS 5510, MATEIS, Lyon, France
关键词
STENT-ASSISTED COILING; CEREBRAL-ARTERY ANEURYSMS; FLOW-DISRUPTION; DEVICE; FEASIBILITY; RECURRENCES; EXPERIENCE; BALLOON; DL;
D O I
10.3174/ajnr.A4457
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Endovascular coiling of wide-neck intracranial aneurysms is associated with low rates of initial angiographic occlusion and high rates of recurrence. The WEB intrasaccular device has been developed specifically for this indication. To date, there has been no report of the long-term follow-up of a series of patients with aneurysms treated with this type of device, to our knowledge. Our aim was to evaluate a 1-year follow-up of angiographic results in a prospective single-center series of patients treated with the WEB-Single-Layer (SL) device. MATERIALS AND METHODS: All patients treated with the WEB-SL device in our center between August 2013 and May 2014 were prospectively included. One-year angiographic outcomes were assessed. Results at follow-up were graded as complete occlusion, neck remnant, or residual aneurysm. RESULTS: Eight patients with 8 unruptured wide-neck aneurysms were enrolled in this study. Average dome width was 7.5 mm (range, 5.4-10.7 mm), and average neck size was 4.9 mm (range, 2.6-6.5 mm). One-year angiographic follow-up obtained in all aneurysms included 1 complete aneurysm occlusion (12.5%), 6 neck remnants (75%), and 1 aneurysm remnant (12.5%). Of 8 aneurysms, worsening of aneurysm occlusion was observed in 2 (25%) by compression of the WEB device. There was no angiographic recurrence of initially totally occluded aneurysms. No bleeding was observed during the follow-up period. CONCLUSIONS: Endovascular therapy of intracranial aneurysms with the WEB-SL device allows treatment of wide-neck aneurysms with a high rate of neck remnant at 1 year, at least partially explained by WEB compression. Initial size selection and technologic improvements could be an option for optimization of aneurysm occlusion in WEB-SL treatment.
引用
收藏
页码:2320 / 2324
页数:5
相关论文
共 17 条
  • [1] Coil Occlusion of Wide-Neck Bifurcation Aneurysms Assisted by a Novel Intra- to Extra-Aneurysmatic Neck-Bridging Device (pCONus): Initial Experience
    Aguilar-Perez, M.
    Kurre, W.
    Fischer, S.
    Baezner, H.
    Henkes, H.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (05) : 965 - 971
  • [2] Endovascular treatment of bifurcation intracranial aneurysms with the WEB SL/SLS: 6-month clinical and angiographic results
    Ambrosi, Patricia Bozzetto
    Gory, Benjamin
    Sivan-Hoffmann, Rotem
    Riva, Roberto
    Signorelli, Francesco
    Labeyrie, Paul-Emile
    Eldesouky, Islam
    Sadeh-Gonike, Udi
    Armoiry, Xavier
    Turjman, Francis
    [J]. INTERVENTIONAL NEURORADIOLOGY, 2015, 21 (04) : 462 - 469
  • [3] "Y" and "X" Stent-Assisted Coiling of Complex and Wide-Neck Intracranial Bifurcation Aneurysms
    Bartolini, B.
    Blanc, R.
    Pistocchi, S.
    Redjem, H.
    Piotin, M.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (11) : 2153 - 2158
  • [4] Stent-Assisted Coiling Versus Balloon Remodeling of Wide-Neck Aneurysms: Comparison of Angiographic Outcomes
    Chalouhi, N.
    Starke, R. M.
    Koltz, M. T.
    Jabbour, P. M.
    Tjoumakaris, S. I.
    Dumont, A. S.
    Rosenwasser, R. H.
    Singhal, S.
    Gonzalez, L. F.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (10) : 1987 - 1992
  • [5] Remnants and Recurrences After the Use of the WEB Intrasaccular Device in Large-Neck Bifurcation Aneurysms
    Cognard, Christophe
    Januel, Anne Christine
    [J]. NEUROSURGERY, 2015, 76 (05) : 522 - 530
  • [6] Endovascular Treatment of Wide-Neck Anterior Communicating Artery Aneurysms Using WEB-DL and WEB-SL: Short-Term Results in a Multicenter Study
    Gherasim, D. N.
    Gory, B.
    Sivan-Hoffmann, R.
    Pierot, L.
    Raoult, H.
    Gauvrit, J. -Y.
    Desal, H.
    Barreau, X.
    Herbreteau, D.
    Riva, R.
    Impiombato, F. Ambesi
    Armoiry, X.
    Turjman, F.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (06) : 1150 - 1154
  • [7] Endovascular Treatment of Middle Cerebral Artery Aneurysms for 120 Nonselected Patients: A Prospective Cohort Study
    Gory, B.
    Rouchaud, A.
    Saleme, S.
    Dalmay, F.
    Riva, R.
    Caire, F.
    Mounayer, C.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (04) : 715 - 720
  • [8] Initial Experience of Intracranial Aneurysm Embolization Using the Balloon Remodeling Technique with Scepter C, a New Double-Lumen Balloon
    Gory, B.
    Kessler, I.
    Nakiri, G. Seizem
    Riva, R.
    Al-Khawaldeh, M.
    Mounayer, C.
    [J]. INTERVENTIONAL NEURORADIOLOGY, 2012, 18 (03) : 284 - 287
  • [9] Gory B, 2015, AJNR AM J NEURORADIO
  • [10] Solitaire AB Stent-Assisted Coiling of Wide-Necked Intracranial Aneurysms: Mid-term Results From the SOLARE Study
    Gory, Benjamin
    Klisch, Joachim
    Bonafe, Alain
    Mounayer, Charbel
    Beaujeux, Remy
    Moret, Jacques
    Lubicz, Boris
    Riva, Roberto
    Turjman, Francis
    [J]. NEUROSURGERY, 2014, 75 (03) : 215 - 219