Effect of a laparoscopic instrument with rotatable handle piece on biomechanical stress during laparoscopic procedures

被引:15
作者
Steinhilber, Benjamin [1 ]
Seibt, Robert [1 ]
Reiff, Florian [1 ]
Rieger, Monika A. [1 ]
Kraemer, Bernhard [2 ]
Rothmund, Ralf [2 ]
机构
[1] Univ Tubingen Hosp, Inst Occupat & Social Med & Hlth Serv Res, Wilhelmstr 27, D-72074 Tubingen, Germany
[2] Univ Tubingen Hosp, Dept Gyneacol & Obstet, D-72074 Tubingen, Germany
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2016年 / 30卷 / 01期
关键词
Laparoscopic surgery; Biomechanical stress; Rotatable handle piece; Simulation; sEMG; MINIMALLY INVASIVE SURGERY; ERGONOMIC HANDLES; DESIGN; MUSCLE; TASKS; ELECTROMYOGRAPHY; NORMALIZATION; PERFORMANCE; GUIDELINES; WORKLOAD;
D O I
10.1007/s00464-015-4164-3
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective To investigate the effect of a pistol grip laparoscopic instrument with a rotatable handle piece (rot-HP) on biomechanical stress and precision as well as a possible interaction between the instrument and working height (WH). Background Biomechanical stress induced by laparoscopic surgery (LS) is associated with work-related upper limb musculoskeletal disorders in surgeons. Ergonomic handle designs of laparoscopic instruments may reduce the risk of musculoskeletal disorders. Methods Without LS experience, 57 healthy subjects (30 women; 27 men, median age: 26) were observed while performing a laparoscopic exercise. Subjects had to pick up coloured pins and place them into a colour-coded wooden set-up inside a pelvitrainer. The exercise was performed at two WHs using the rot-HP and a standard laparoscopic (fixed) handle piece in randomized sequence. Biomechanical stress was monitored via surface electromyography (sEMG) on fife muscles from the upper limb and shoulder region. Further, the wrist angle (palmar and dorsi flexion) and posture of the dominant upper arm were recorded. Precision was assessed using the number of validly placed pins and process time. Results sEMG parameters and upper arm postures indicated no differences in biomechanical stress related to either laparoscopic handle piece. The higher WH was associated with higher trapezius and deltoid activity and elevated arm postures (p < 0.05). Neutral wrist positions were more frequent using the rot-HP, and the lower WH resulted in slightly more neutral wrist positions. Precision was similar during all experimental conditions. Conclusions The rot-HP did not decrease biomechanical stress in the shoulder or lower arm muscles. However, wrist angle position may be optimized without affecting precision. Long-term effects of the rot-HP on preventing complaints of the upper extremity should be evaluated. Low WH is recommended to reduce biomechanical stress in the shoulder during laparoscopic surgery.
引用
收藏
页码:78 / 88
页数:11
相关论文
共 36 条
[21]   Ergonomic deficiencies in the operating room: Examples from minimally invasive surgery [J].
Matern, Ulrich .
WORK-A JOURNAL OF PREVENTION ASSESSMENT & REHABILITATION, 2009, 33 (02) :165-168
[22]   Normalization of surface EMG amplitude from the upper trapezius muscle in ergonomic studies - A review [J].
Mathiassen, SE ;
Winkel, J ;
Hagg, GM .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 1995, 5 (04) :197-226
[23]   Patients Benefit While Surgeons Suffer: An Impending Epidemic [J].
Park, Adrian ;
Lee, Gyusung ;
Seagull, F. Jacob ;
Meenaghan, Nora ;
Dexter, David .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2010, 210 (03) :306-313
[24]   The effect of using laparoscopic instruments on muscle activation patterns during minimally invasive surgical training procedures [J].
Quick, NE ;
Gillette, JC ;
Shapiro, R ;
Adrales, GL ;
Gerlach, D ;
Park, AE .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2003, 17 (03) :462-465
[25]   Design and evaluation of a new ergonomic handle for instruments in minimally invasive surgery [J].
Sancibrian, Ramon ;
Gutierrez-Diez, Maria C. ;
Torre-Ferrero, Carlos ;
Benito-Gonzalez, Maria A. ;
Redondo-Figuero, Carlos ;
Manuel-Palazuelos, Jose C. .
JOURNAL OF SURGICAL RESEARCH, 2014, 188 (01) :88-99
[26]   Survey on Surgical Instrument Handle Design: Ergonomics and Acceptance [J].
Santos-Carreras, Laura ;
Hagen, Monika ;
Gassert, Roger ;
Bleuler, Hannes .
SURGICAL INNOVATION, 2012, 19 (01) :50-59
[27]  
Steinhilber B, P ISEK 2014 20 C INT, P217
[28]   In-line vs pistol-grip handles in a laparoscopic simulator - A randomized controlled crossover trial [J].
Uchal, M ;
Brogger, J ;
Rukas, R ;
Karlsen, B ;
Bergamaschi, R .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2002, 16 (12) :1771-1773
[29]   Assessment of fatigue, monitor placement, and surgical experience during simulated laparoscopic surgery [J].
Uhrich, ML ;
Underwood, RA ;
Standeven, JW ;
Soper, NJ ;
Engsberg, JR .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2002, 16 (04) :635-639
[30]   Optimal ergonomics for laparoscopic surgery in minimally invasive surgery suites: a review and guidelines [J].
van Det, M. J. ;
Meijerink, W. J. H. J. ;
Hoff, C. ;
Totte, E. R. ;
Pierie, J. P. E. N. .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2009, 23 (06) :1279-1285