A liquid scintillation detector for radioassay of gadolinium-loaded liquid scintillator for the LZ Outer Detector

被引:12
作者
Haselschwardt, S. J. [1 ]
Shaw, S. [1 ]
Nelson, H. N. [1 ]
Withere, M. S. [1 ,2 ]
Yeh, M. [3 ]
Lesko, K. T. [2 ]
Cole, A. [2 ]
Kyre, S. [1 ]
White, D. T. [1 ]
机构
[1] Univ Calif Santa Barbara, Dept Phys, Broida Hall, Santa Barbara, CA 93106 USA
[2] LBNL, 1 Cyclotron Rd, Berkeley, CA 94720 USA
[3] BNL, POB 5000, Upton, NY 11973 USA
关键词
Low-background; Gadolinium; Liquid scintillator; EFFICIENCY; CRYSTALS; REMOVAL;
D O I
10.1016/j.nima.2019.05.055
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report on the design and performance of the LUX-ZEPLIN (LZ) "Screener", a small liquid scintillator detector consisting of approximate to 23 kg of LAB-based gadolinium-loaded liquid scintillator (GdLS) to be used in the LZ Outer Detector. The Outer Detector will be filled with 17.3 tonnes of GdLS and will surround the central liquid xenon time projection chamber of LZ. Its primary function will be to tag neutron events in the liquid xenon which could mimic a WIMP dark matter signal. To meet the deadtime requirements for the Outer Detector, the radioimpurity levels in the GdLS must be kept below less than or similar to 0.07 mBq/kg. This background level corresponds to a rate of approximate to 50 Hz above an energy threshold of 100 keV. The Screener was operated in the ultra-low-background environment of the former LUX water shield in the Davis Laboratory at the Sanford Underground Research Facility for radioassay of the GdLS. Careful selection of detector materials and use of ultra-low-background PMTs allows the measurement of a variety of radioimpurities. The C-14/C-12 ratio in the scintillator is measured to be (2.83 +/- 0.06(stat.)+/- 0.01(sys.)) x 10(-1)7. Use of pulse shape discrimination allows the concentration of isotopes throughout the U-238, U-235, and Th-232 chains to be measured by fitting the collected spectra from alpha and beta events. We find that equilibrium is broken in the U-238 and Th-232 chains and that a significant portion of the contamination in the GdLS results from decays in the Ac-227 subchain of the U-235 series. Predictions for the singles rate in the Outer Detector are presented. The rate from radioimpurities above 100 keV in the GdLS is estimated to be 97.9 +/- 6.4 Hz, with 65.5 +/- 1.9 Hz resulting from alpha-decays.
引用
收藏
页码:148 / 163
页数:16
相关论文
共 60 条
[51]   Scintillation decay time and pulse shape discrimination in oxygenated and deoxygenated solutions of linear alkylbenzene for the SNO plus experiment [J].
O'Keeffe, H. M. ;
O'Sullivan, E. ;
Chen, M. C. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 640 (01) :119-122
[52]   Development of a Gd loaded liquid scintillator for electron anti-neutrino spectroscopy [J].
Piepke, AG ;
Moser, SW ;
Novikov, VM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 432 (2-3) :392-398
[53]   RESPONSE OF SOME SCINTILLATION CRYSTALS TO CHARGED PARTICLES [J].
TAYLOR, CJ ;
JENTSCHKE, WK ;
REMLEY, ME ;
EBY, FS ;
KRUGER, PG .
PHYSICAL REVIEW, 1951, 84 (05) :1034-1043
[54]   Removal of cosmogenic Be-7 from scintillators [J].
Vogelaar, RB ;
Benziger, J ;
Calaprice, FP ;
Darnton, N .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 372 (1-2) :59-62
[55]  
von Krosigk B., 2015, THESIS
[56]   ABSOLUTE QUANTUM EFFICIENCY OF PHOTOFLUORESCENCE OF ANTHRACENE CRYSTALS [J].
WRIGHT, GT .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1955, 68 (04) :241-248
[57]   ABSOLUTE SCINTILLATION EFFICIENCY OF ANTHRACENCE CRYSTALS [J].
WRIGHT, GT .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1955, 68 (11) :929-937
[58]   Oxygen quenching in a LAB based liquid scintillator and the nitrogen bubbling model [J].
Xiao Hua-Lin ;
Deng Jing-Shan ;
Wang Nai-Yan .
CHINESE PHYSICS C, 2010, 34 (05) :571-575
[59]   Purification of lanthanides for large neutrino detectors: Thorium removal from gadolinium chloride [J].
Yeh, M. ;
Cumming, J. B. ;
Hans, S. ;
Hahn, R. L. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 618 (1-3) :124-130
[60]   Study of a prototype water Cherenkov detector for the Daya Bay neutrino experiment [J].
Yu, Zeyuan ;
Lu, Haoqi ;
Yang, Changgen ;
Wang, Lingyu ;
Guan, Mengyun ;
Liu, Jinchang ;
Wang, Ruiguang ;
Wang, Zhimin ;
Xu, Jilei ;
Cao, Jun ;
Wang, Yifang .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 682 :26-30